Failure by Corrosion of a Steinmann Pin Used for Intramedullary Fixation
Jonathan Cohen; W. Sanford Foultz
View Disclosures and Other Information
Departments of Orthopaedic Surgery and Pathology and the Division of Laboratories and Research, The Children's Hospital Medical Center, Boston
1960 by The Journal of Bone and Joint Surgery, Incorporated
The Journal of Bone & Joint Surgery.  1960; 42:1201-1206 
5 Recommendations (Recommend) | 3 Comments | Saved by 3 Users Save Case
text A A A

Abstract

A case report is presented where a Steinmann pin used as an intramedullary rod in a fibula showed marked pitting corrosion and ultimately had to be removed because of tissue reaction. The composition of the steel was the principal factor in promoting corrosion, although cyclic stress and non-metallic inclusions probably influenced the localization of the corrosion process.

Figures in this Article
    This article is only available in the PDF format. Download the PDF to view the article, as well as its associated figures and tables.

    Topics

    Accreditation Statement
    These activities have been planned and implemented in accordance with the Essential Areas and policies of the Accreditation Council for Continuing Medical Education (ACCME) through the joint sponsorship of the American Academy of Orthopaedic Surgeons and The Journal of Bone and Joint Surgery, Inc. The American Academy of Orthopaedic Surgeons is accredited by the ACCME to provide continuing medical education for physicians.
    CME Activities Associated with This Article
    Submit a Comment
    Please read the other comments before you post yours. Contributors must reveal any conflict of interest.
    Comments are moderated and will appear on the site at the discretion of JBJS editorial staff.

    * = Required Field
    (if multiple authors, separate names by comma)
    Example: John Doe




    Related Articles
    Related Cases
    Related Content
    Related Audio and Videos
    PubMed Articles
    Mixing and matching causing taper wear: Corrosion associated with pseudotumour formation.
    The Journal of bone and joint surgery. British volume: Issue date- 2012 Feb
    Femoral Stem Fracture and In Vivo Corrosion of Retrieved Modular Femoral Hips.
    The Journal of arthroplasty: Issue date- 2011 Dec 28
    Clinical Trials
    Readers of This Also Read...
    jbjs jobs
    12/22/2011
    VA - Charleston Area Medical Center
    12/22/2011
    ME - Central Maine Medical Center