The Noiles hinged knee prosthesis has been reported to diminish sheer
and tensile loading on the cement-bone interface. Theoretically, this
should eliminate the complication of loosening, commonly reported in hinged
total knee prostheses. We reviewed the cases of all of the patients who
have been treated with the Noiles knee prosthesis at our institution.
Eighteen knees in fourteen patients were available, seventeen of which were
followed for more than five years. Although all of the prostheses did well
initially, ten knees in eight patients (56 per cent) had failed by an
average of thirty-two months postoperatively. These eight patients included
four of the five who weighed more than ninety kilograms and all of those
who had had a prior arthroplasty. The failures were clearly defined along
the lines of sex distribution. Measurements on the radiographs showed a
significant difference between sexes in the metaphyseal diameter and thus
in the percentage of the femoral canal that was filled by the prosthesis
and cement. We concluded that prosthetic design must incorporate cortical
cement support to reduce stress at the cement-bone interface, especially in
a medullary canal with a larger diameter.