The effect of femoral stem geometry on interface motion in uncemented porous-coated total hip prostheses. Comparison of straight-stem and curved-stem designs

The Journal of Bone & Joint Surgery.  1992; 74:839-848 
5 Recommendations (Recommend) | 3 Comments | Saved by 3 Users Save Case
text A A A

Abstract

We compared the magnitudes of motion between the prosthesis and bone during axial and torsional loading in seven matched pairs of fresh-frozen femora of cadavera in which an uncemented, collarless, isthmus-filling, straight-stem (Harris-Galante) prosthesis had been placed in one femur and an uncemented, collarless, proximal-filling, curved-stem (anatomic) prosthesis had been placed in the other femur. The comparison was performed in order to determine the effect of the geometry of the stem on the magnitude of motion. Single-limb-stance loads and combined axial and torsional loads were applied to the implanted femoral prostheses with the use of a jig that simulated acetabular and trochanteric loading. Extensometers were used to measure motion at the prosthesis-bone interface. The prostheses were then removed and were reinserted, with cement applied to the proximal porous coating to simulate ingrowth of bone. The single-limb-stance and combined axial and torsional loads were reapplied and the magnitude of motion was recorded again. No significant differences in the magnitudes of the motion were found between the femora in which the straight stem had been implanted and the femora in which the curved stem had been implanted, during either simulated single-limb-stance or low-intensity torsional loading. When large torsional moments (twenty-two newton-meters) were applied, significantly less motion occurred at the bone-prosthesis interface, both proximally (p = 0.019) and distally (p = 0.0013), in the femora with the curved-stem implant than in the femora with the straight-stem implant. When cement had been applied proximally, proximal and distal motion between the prosthesis and the femur was decreased during simulated single-limb-stance and during torsional loading in the femora with the straight stem and the femora with the curved stem.

Figures in this Article
    This article is only available in the PDF format. Download the PDF to view the article, as well as its associated figures and tables.

    Topics

    Accreditation Statement
    These activities have been planned and implemented in accordance with the Essential Areas and policies of the Accreditation Council for Continuing Medical Education (ACCME) through the joint sponsorship of the American Academy of Orthopaedic Surgeons and The Journal of Bone and Joint Surgery, Inc. The American Academy of Orthopaedic Surgeons is accredited by the ACCME to provide continuing medical education for physicians.
    CME Activities Associated with This Article
    Submit a Comment
    Please read the other comments before you post yours. Contributors must reveal any conflict of interest.
    Comments are moderated and will appear on the site at the discretion of JBJS editorial staff.

    * = Required Field
    (if multiple authors, separate names by comma)
    Example: John Doe




    Related Articles
    Related Cases
    Related Content
    Topic Collections
    Hip
    Related Audio and Videos
    Clinical Trials
    Readers of This Also Read...
    jbjs jobs
    12/22/2011
    ME - Central Maine Medical Center
    12/22/2011
    VA - Charleston Area Medical Center
    12/22/2011
    Virginia - Charleston Area Medical Center