Fracture of the acetabulum during insertion of an oversized hemispherical component

The Journal of Bone & Joint Surgery.  1995; 77:111-117 
5 Recommendations (Recommend) | 3 Comments | Saved by 3 Users Save Case
text A A A

Abstract

A hemispherical metal-backed acetabular component that was oversized by two or four millimeters was introduced without cement into thirty fresh or embalmed cadaveric acetabula that were fifty, fifty-two, fifty-four, fifty-six, or fifty-eight millimeters in diameter. Manual tapping was used to insert the component into ten fresh and ten embalmed specimens, and a materials testing machine was used for the remaining five fresh and five embalmed specimens. The magnitude of force and the number of pulses necessary to seat the component were determined for the insertions performed with the materials testing machine; 2000 and 3000 newtons of force were required to insert the cups that were oversized by two and four millimeters, respectively. Visual and radiographic assessments were used to identify any resulting acetabular fractures. Over-all, eighteen (60 per cent) of the thirty acetabular specimens fractured. Of the eighteen fractures, only fifteen were detectable radiographically: ten on en face or lateral radiographs of the cup and five only on oblique radiographs of the cup. The remaining three fractures were apparent only after the removal of the acetabular component.(ABSTRACT TRUNCATED AT 250 WORDS)

Figures in this Article
    This article is only available in the PDF format. Download the PDF to view the article, as well as its associated figures and tables.

    Topics

    Accreditation Statement
    These activities have been planned and implemented in accordance with the Essential Areas and policies of the Accreditation Council for Continuing Medical Education (ACCME) through the joint sponsorship of the American Academy of Orthopaedic Surgeons and The Journal of Bone and Joint Surgery, Inc. The American Academy of Orthopaedic Surgeons is accredited by the ACCME to provide continuing medical education for physicians.
    CME Activities Associated with This Article
    Submit a Comment
    Please read the other comments before you post yours. Contributors must reveal any conflict of interest.
    Comments are moderated and will appear on the site at the discretion of JBJS editorial staff.

    * = Required Field
    (if multiple authors, separate names by comma)
    Example: John Doe




    Related Articles
    Related Cases
    Related Content
    Topic Collections
    Hip
    Related Audio and Videos
    PubMed Articles
    Multiple associated injuries are common with spine fractures during war.
    The spine journal : official journal of the North American Spine Society: Issue date- 2011 Nov 2
    Causes of Failure of Ceramic-on-Ceramic and Metal-on-Metal Hip Arthroplasties.
    Clinical orthopaedics and related research: Issue date- 2012 Feb
    Clinical Trials
    Readers of This Also Read...
    jbjs jobs
    12/22/2011
    VA - Charleston Area Medical Center
    12/22/2011
    ME - Central Maine Medical Center
    12/22/2011
    Virginia - Charleston Area Medical Center