Fracture of an Acetabular Component Inserted without Cement: A Case Report
R. T. TROUSDALE, M.D.†; D. J. BERRY, M.D.†; J. JACOBS, M.D.‡; J. L. GILBERT, PH.D.§, ROCHESTER, MINNESOTA
View Disclosures and Other Information
Investigation performed at the Mayo Clinic, Rochester
The Journal of Bone & Joint Surgery.  1997; 79:901-5 
5 Recommendations (Recommend) | 3 Comments | Saved by 3 Users Save Case
The use of acetabular components without cement for total hip arthroplasty has increased dramatically in the last decade, for several reasons. With increased follow-up, fixation of sockets with cement has proved to be problematic, with as many as 48 per cent of such sockets reported to be radiographically loose at twenty years1,9,10,15,19,20. In a study of 330 total hip replacements performed with cement, Schulte et al. noted that 23 per cent of the ninety-eight hips in the patients who were alive at a minimum of twenty years had a loose or revised acetabular component. Kavanagh et al.10 reviewed the results of the first 333 Charnley total hip replacements performed with cement at the Mayo Clinic and found possible loosening of thirty-three (48 per cent) of sixty-nine cups in patients who were followed for twenty years.
Sockets designed to be inserted without cement are so-called user-friendly. They are technically easy to insert, and the position of the cup can be readily changed intraoperatively. Growth of bone into the sockets is reliable, as has been shown in many studies of retrieved specimens5,6,8,12,17. Elevated modular polyethylene liners may improve positioning of the socket (by modifying the angle of abduction or anteversion) and can accommodate multiple sizes of the femoral head, which is useful in revision operations.
Clinical and radiographic results of total hip arthroplasty without cementing of the acetabular component have been quite satisfactory11,13,14. However, when the results of such arthroplasties are evaluated, it should be kept in mind that loosening of cemented sockets historically has not been a major problem until at least five to ten years after the operation16. Although the fixation of acetabular components inserted without cement often is stable at the time of intermediate follow-up, problems not seen with cemented implants have been reported less than ten years after insertion of sockets without cement. These problems have included dissociation of the liner from the shell, catastrophic failure of the polyethylene liner, impingement on the iliopsoas, possibly accelerated rates of wear of the polyethylene, and acetabular osteolysis2-4,7,18.
We report a case in which the metal shell of an acetabular component that had been fixed to a large segment of autogenous graft without cement fractured after less than ten years.

*One or more of the authors has received or will receive benefits for personal or professional use from a commercial party related directly or indirectly to the subject of this article. In addition, benefits have been or will be directed to a research fund or foundation, educational institution, or other non-profit organization with which one or more of the authors is associated. No funds were received in support of this study.

‡Rush-Presbyterian-St. Luke's Medical Center, 1725 West Harrison Street, Suite 1063, Chicago, Illinois 60612-3824.

§Department of Biological Materials, Northwestern University, 311 East Chicago Avenue, Chicago, Illinois 60611.

*One or more of the authors has received or will receive benefits for personal or professional use from a commercial party related directly or indirectly to the subject of this article. In addition, benefits have been or will be directed to a research fund or foundation, educational institution, or other non-profit organization with which one or more of the authors is associated. No funds were received in support of this study.
‡Rush-Presbyterian-St. Luke's Medical Center, 1725 West Harrison Street, Suite 1063, Chicago, Illinois 60612-3824.
§Department of Biological Materials, Northwestern University, 311 East Chicago Avenue, Chicago, Illinois 60611.
 
Anchor for JumpAnchor for Jump
+Fig. 1 Radiograph of the pelvis, showing a dysplastic hip with end-stage osteoarthrosis.
 
Anchor for JumpAnchor for Jump
+Fig. 2 Immediate postoperative radiograph showing the acetabular component, which was placed superior and lateral to the normal hip center without cement. A large portion of the socket is covered by autogenous graft from the femoral head.
 
Anchor for JumpAnchor for Jump
+Fig. 3 Radiograph showing the fractured acetabular shell. The femoral component has subsided and is radiographically loose.
 
Anchor for JumpAnchor for Jump
+Fig. 4 Photograph showing the retrieved components. The acetabular component is fractured along its multiple screw-holes.
 
Anchor for JumpAnchor for Jump
+Fig. 5 Scanning electron micrograph of the site of the initiation of the fatigue crack. The large arrow indicates the free surface of the countersink of the screw-hole, and the curved arrow shows the intersection of the countersink with the fracture surface. On the fracture surface, there is an approximately 500-micrometer flat, fracture facet with the so-called river pattern (small arrow). These features are characteristic of a fatigue failure (magnification, x 90).
 
Anchor for JumpAnchor for Jump
+Fig. 6 Radiograph made after the revision, which included use of a reinforcement ring. The center of the hip has been restored to a more anatomical location.
A fifty-three-year-old woman who had a history of congenital dysplasia of the left hip was first seen in our clinic in 1987. The dysplasia had been diagnosed when the patient was a child, and an open reduction had been performed at the age of four years. One year later, she had a second open reduction. She did relatively well until 1987, when she had increasing pain in the hip that limited her daily activities (Fig. 1). She was able to walk no more than one block, and she needed a cane. In 1987, a total arthroplasty of the left hip was performed with insertion of a Harris-Galante-I porous-coated acetabular component (Zimmer, Warsaw, Indiana) and a proximally porous-coated femoral stem (Osteonics, Allendale, New Jersey) without cement (Fig. 2). A forty-eight-millimeter acetabular shell was placed superiorly, and autogenous graft from the femoral head was secured to the ilium to improve superolateral coverage. The graft was secured with four screws, and the acetabular component was secured with three screws. The patient initially did well, with a substantial decrease in pain and increase in the level of activity.
Seven years after the operation, mild so-called start-up pain (pain that occurs with the initiation of walking) developed in the thigh. At that time, it was noticed that the femoral component was loose. The acetabular component appeared to be intact. The pain in the thigh was not severe enough to warrant operative intervention. Over the next ten months, the pain in the thigh increased and discomfort developed in the groin and buttock. There was crepitation with motion of the hip, and the patient required crutches to walk. Approximately eight years after the primary total hip replacement, radiographs revealed that, although there was no history of trauma, the acetabular component had fractured (Fig. 3).
At revision, the polyethylene was noted to be intact but dissociated from the fractured shell. There was extensive abrasion, delamination, and pitting on the non-articulating surface of the liner as well as eccentric wear and foreign-body debris on the bearing surface with cracking on the superior rim. The acetabular shell was fractured along its outer third between the multiple screw-holes (Fig. 4). The fractured component was analyzed visually, and specimens were examined with a scanning electron microscope (model S120; Cambridge Instruments, Deerfield, Illinois). The morphological appearance of the fracture surface was consistent with a fatigue failure emanating from the countersink of a screw-hole. This was evident on visual examination in that an area adjacent to a posterosuperior screw-hole had a glossy, faceted appearance. The fracture surface exhibited characteristic patterns of fatigue on examination with the scanning electron microscope. The site of the initiation of the fatigue crack (Fig. 5) consisted of a 500-micrometer flat, fracture facet with the so-called river pattern emanating from the surface of the countersink. The countersink adjacent to the site of the initiation of the crack had persistent slip bands and a secondary fatigue crack indicating extensive fatigue damage in the vicinity of the fracture. The inferior portion of the acetabular component was well fixed but easily removed. The femoral component was grossly loose. Because of the large amount of graft used in the primary total hip replacement, reconstruction was performed with use of a fifty-six-millimeter acetabular reinforcement ring and a femoral head allograft to fill multiple cavitary defects. Approximately 70 per cent of the autogenous graft that had been placed during the primary total hip replacement had been incorporated. A standard femoral stem fixed with antibiotic-impregnated cement was used in the revision (Fig. 6).
At the time of the most recent follow-up examination, the recovery had been uneventful and the patient was pain-free and able to walk without aids.
The long-term survival of acetabular sockets inserted without cement is unknown. It is hoped that the rate of radiographic loosening will be lower than the high rates seen with cemented sockets with increased follow-up. The intermediate clinical and radiographic results of most sockets designed to be inserted without cement have been excellent, but there have been problems, even after periods of implantation of less than a decade. These complications include dissociation of the liner from the shell, catastrophic failure of the polyethylene liner, impingement on the iliopsoas, and possibly increased rates of polyethylene wear2-4,7,19. A fractured acetabular shell appears to be extremely rare.
The cause of the fracture in our patient may have been multifactorial. The thickness of the substrate of the Harris-Galante-I porous-coated cup was 2.3 millimeters with an additional 1.4 millimeters of titanium mesh. One of the changes in the design of the second generation of this component was an increase in the thickness of the shell by 1.5 millimeters. This patient also had dysplasia of the hip, and the initial acetabular shell was placed on a large segment of autogenous graft. The acetabular component became fixed by inferior ingrowth of bone. Subsequently, the graft appears to have been reabsorbed superiorly and laterally. As the shell was continually loaded during the eight years of use, the unsupported superolateral portion of the socket eventually failed through the screw-holes. The initial acetabular component was placed superiorly and laterally to the normal center of rotation of the hip. This position of the socket leads to a short limb with a marked increase in the joint-reaction contact forces. The valgus position of the femoral stem also decreases the offset of the femoral head and increases the abductor forces on the hip.
This case report demonstrates that an acetabular component inserted without cement can fracture when support by intact host bone is diminished. Maximizing contact between the cup and the host bone may decrease the risk of fracture of the component.
Andersson, G. B. J.; Freeman, M. A. R.; and Swanson, S. A. V.: Loosening of the cemented acetabular cup in total hip replacement. J. Bone and Joint Surg.,54-B(4): 590-599, 1972.54-B(4)590  1972 
 
Berry, D. J.; Barnes, C. L.; Scott, R. D.; Cabanela, M. E.; and Poss, R.: Catastrophic failure of the polyethylene liner of uncemented acetabular components. J. Bone and Joint Surg.,76-B(4): 575-578, 1994.76-B(4)575  1994 
 
Cameron, H. U.: Disassociation of a polyethylene liner from an acetabular cup. Orthop. Rev.,22: 1160-1161, 1993.221160  1993  [PubMed]
 
Cook, S. D., and Thomas, K. A.: Fatigue failure of noncemented porous-coated implants. A retrieval study. J. Bone and Joint Surg.,73-B(1): 20-24, 1991.73-B(1)20  1991 
 
Cook, S. D.; Thomas, K. A.; and Haddad, R. J., Jr.: Histologic analysis of retrieved human porous-coated total joint components. Clin. Orthop.,234: 90-101, 1988.23490  1988  [PubMed]
 
Cook, S. D.; Barrack, R. L.; Thomas, K. A.; and Haddad, R. J., Jr.: Quantitative analysis of tissue growth into human porous total hip components. J. Arthroplasty,3: 249-262, 1988.3249  1988  [PubMed]
 
Ferenz, C. C.: Polyethylene insert dislocation in a screw-in acetabular cup. A case report. J. Arthroplasty,3: 201-204, 1988.3201  1988  [PubMed]
 
Ferro, T. D.; Zettl-Schaffer, K. F.; Engh, C. A.; McGovern, T. F.; and Bragdon, C.: Quantification of bone ingrowth into porous coated acetabular components retrieved postmortem using backscattered SEM techniques. Trans. Orthop. Res. Soc.,17: 387, 1992.17387  1992 
 
Kavanagh, B. F.; Dewitz, M. A.; Ilstrup, D. M.; Stauffer, R. N.; and Coventry, M. B.: Charnley total hip arthroplasty with cement. Fifteen-year results. J. Bone and Joint Surg.,71-A: 1496-1503, Dec. 1989.71-A1496  1989 
 
Kavanagh, B. F.; Wallrichs, S.; Dewitz, M.; Berry, D.; Currier, B.; Ilstrup, D.; and Coventry, M. B.: Charnley low-friction arthroplasty of the hip. Twenty-year results with cement. J. Arthroplasty,9: 229-234, 1994.9229  1994  [PubMed]
 
Martell, J. M.; Pierson, R. H., III; Jacobs, J. J.; Rosenberg, A. G.; Maley, M.; and Galante, J. O.: Primary total hip reconstruction with a titanium fiber-coated prosthesis inserted without cement. J. Bone and Joint Surg.,75-A: 554-571, April 1993.75-A554  1993 
 
Pidhorz, L. E.; Urban, R. M.; Jacobs, J. J.; Sumner, D. R.; and Galante, J. O.: A quantitative study of bone and soft tissues in cementless porous-coated acetabular components retrieved at autopsy. J. Arthroplasty,8: 213-225, 1993.8213  1993  [PubMed]
 
Schmalzried, T. P., and Harris, W. H.: The Harris-Galante porous-coated acetabular component with screw fixation. Radiographic analysis of eighty-three primary hip replacements at a minimum of five years. J. Bone and Joint Surg.,74-A: 1130-1139, Sept. 1992.74-A1130  1992 
 
Schmalzried, T. P.; Wessinger, S. J.; Hill, G. E.; and Harris, W. H.: The Harris-Galante porous acetabular component press-fit without screw fixation. Five-year radiographic analysis of primary cases. J. Arthroplasty,9: 235-242, 1994.9235  1994  [PubMed]
 
Schulte, K. R.; Callaghan, J. J.; Kelley, S. S.; and Johnston, R. C.: The outcome of Charnley total hip arthroplasty with cement after a minimum twenty-year follow-up. The results of one surgeon. J. Bone and Joint Surg.,75-A: 961-975, July 1993.75-A961  1993 
 
Stauffer, R. N.: Ten-year follow-up study of total hip replacement. With particular reference to roentgenographic loosening of the components. J. Bone and Joint Surg.,64-A: 983-990, Sept. 1982.64-A983  1982 
 
Sumner, D. R.; Jasty, M.; Jacobs, J. J.; Urban, R. M.; Bragdon, C. R.; Harris, W. H.; and Galante, J. O.: Histology of porous-coated acetabular components. 25 cementless cups retrieved after arthroplasty. Acta Orthop. Scandinavica,64: 619-626, 1993.64619  1993 
 
Trousdale, R. T.; Cabanela, M. E.; and Berry, D. J.: Anterior iliopsoas impingement after total hip arthroplasty. J. Arthroplasty,10: 546-549, 1995.10546  1995  [PubMed]
 
Wixson, R. L.; Stulberg, S. D.; and Mehlhoff, M.: Total hip replacement with cemented, uncemented, and hybrid prostheses. A comparison of clinical and radiographic results at two to four years. J. Bone and Joint Surg.,73-A: 257-270, Feb. 1991.73-A257  1991 
 
Wroblewski, B. M.: 15-21-year results of the Charnley low-friction arthroplasty. Clin. Orthop.,211: 30-35, 1986.21130  1986  [PubMed]
 

Submit a comment

Topics

Anchor for JumpAnchor for Jump
+Fig. 1 Radiograph of the pelvis, showing a dysplastic hip with end-stage osteoarthrosis.
Anchor for JumpAnchor for Jump
+Fig. 2 Immediate postoperative radiograph showing the acetabular component, which was placed superior and lateral to the normal hip center without cement. A large portion of the socket is covered by autogenous graft from the femoral head.
Anchor for JumpAnchor for Jump
+Fig. 3 Radiograph showing the fractured acetabular shell. The femoral component has subsided and is radiographically loose.
Anchor for JumpAnchor for Jump
+Fig. 4 Photograph showing the retrieved components. The acetabular component is fractured along its multiple screw-holes.
Anchor for JumpAnchor for Jump
+Fig. 5 Scanning electron micrograph of the site of the initiation of the fatigue crack. The large arrow indicates the free surface of the countersink of the screw-hole, and the curved arrow shows the intersection of the countersink with the fracture surface. On the fracture surface, there is an approximately 500-micrometer flat, fracture facet with the so-called river pattern (small arrow). These features are characteristic of a fatigue failure (magnification, x 90).
Anchor for JumpAnchor for Jump
+Fig. 6 Radiograph made after the revision, which included use of a reinforcement ring. The center of the hip has been restored to a more anatomical location.
Andersson, G. B. J.; Freeman, M. A. R.; and Swanson, S. A. V.: Loosening of the cemented acetabular cup in total hip replacement. J. Bone and Joint Surg.,54-B(4): 590-599, 1972.54-B(4)590  1972 
 
Berry, D. J.; Barnes, C. L.; Scott, R. D.; Cabanela, M. E.; and Poss, R.: Catastrophic failure of the polyethylene liner of uncemented acetabular components. J. Bone and Joint Surg.,76-B(4): 575-578, 1994.76-B(4)575  1994 
 
Cameron, H. U.: Disassociation of a polyethylene liner from an acetabular cup. Orthop. Rev.,22: 1160-1161, 1993.221160  1993  [PubMed]
 
Cook, S. D., and Thomas, K. A.: Fatigue failure of noncemented porous-coated implants. A retrieval study. J. Bone and Joint Surg.,73-B(1): 20-24, 1991.73-B(1)20  1991 
 
Cook, S. D.; Thomas, K. A.; and Haddad, R. J., Jr.: Histologic analysis of retrieved human porous-coated total joint components. Clin. Orthop.,234: 90-101, 1988.23490  1988  [PubMed]
 
Cook, S. D.; Barrack, R. L.; Thomas, K. A.; and Haddad, R. J., Jr.: Quantitative analysis of tissue growth into human porous total hip components. J. Arthroplasty,3: 249-262, 1988.3249  1988  [PubMed]
 
Ferenz, C. C.: Polyethylene insert dislocation in a screw-in acetabular cup. A case report. J. Arthroplasty,3: 201-204, 1988.3201  1988  [PubMed]
 
Ferro, T. D.; Zettl-Schaffer, K. F.; Engh, C. A.; McGovern, T. F.; and Bragdon, C.: Quantification of bone ingrowth into porous coated acetabular components retrieved postmortem using backscattered SEM techniques. Trans. Orthop. Res. Soc.,17: 387, 1992.17387  1992 
 
Kavanagh, B. F.; Dewitz, M. A.; Ilstrup, D. M.; Stauffer, R. N.; and Coventry, M. B.: Charnley total hip arthroplasty with cement. Fifteen-year results. J. Bone and Joint Surg.,71-A: 1496-1503, Dec. 1989.71-A1496  1989 
 
Kavanagh, B. F.; Wallrichs, S.; Dewitz, M.; Berry, D.; Currier, B.; Ilstrup, D.; and Coventry, M. B.: Charnley low-friction arthroplasty of the hip. Twenty-year results with cement. J. Arthroplasty,9: 229-234, 1994.9229  1994  [PubMed]
 
Martell, J. M.; Pierson, R. H., III; Jacobs, J. J.; Rosenberg, A. G.; Maley, M.; and Galante, J. O.: Primary total hip reconstruction with a titanium fiber-coated prosthesis inserted without cement. J. Bone and Joint Surg.,75-A: 554-571, April 1993.75-A554  1993 
 
Pidhorz, L. E.; Urban, R. M.; Jacobs, J. J.; Sumner, D. R.; and Galante, J. O.: A quantitative study of bone and soft tissues in cementless porous-coated acetabular components retrieved at autopsy. J. Arthroplasty,8: 213-225, 1993.8213  1993  [PubMed]
 
Schmalzried, T. P., and Harris, W. H.: The Harris-Galante porous-coated acetabular component with screw fixation. Radiographic analysis of eighty-three primary hip replacements at a minimum of five years. J. Bone and Joint Surg.,74-A: 1130-1139, Sept. 1992.74-A1130  1992 
 
Schmalzried, T. P.; Wessinger, S. J.; Hill, G. E.; and Harris, W. H.: The Harris-Galante porous acetabular component press-fit without screw fixation. Five-year radiographic analysis of primary cases. J. Arthroplasty,9: 235-242, 1994.9235  1994  [PubMed]
 
Schulte, K. R.; Callaghan, J. J.; Kelley, S. S.; and Johnston, R. C.: The outcome of Charnley total hip arthroplasty with cement after a minimum twenty-year follow-up. The results of one surgeon. J. Bone and Joint Surg.,75-A: 961-975, July 1993.75-A961  1993 
 
Stauffer, R. N.: Ten-year follow-up study of total hip replacement. With particular reference to roentgenographic loosening of the components. J. Bone and Joint Surg.,64-A: 983-990, Sept. 1982.64-A983  1982 
 
Sumner, D. R.; Jasty, M.; Jacobs, J. J.; Urban, R. M.; Bragdon, C. R.; Harris, W. H.; and Galante, J. O.: Histology of porous-coated acetabular components. 25 cementless cups retrieved after arthroplasty. Acta Orthop. Scandinavica,64: 619-626, 1993.64619  1993 
 
Trousdale, R. T.; Cabanela, M. E.; and Berry, D. J.: Anterior iliopsoas impingement after total hip arthroplasty. J. Arthroplasty,10: 546-549, 1995.10546  1995  [PubMed]
 
Wixson, R. L.; Stulberg, S. D.; and Mehlhoff, M.: Total hip replacement with cemented, uncemented, and hybrid prostheses. A comparison of clinical and radiographic results at two to four years. J. Bone and Joint Surg.,73-A: 257-270, Feb. 1991.73-A257  1991 
 
Wroblewski, B. M.: 15-21-year results of the Charnley low-friction arthroplasty. Clin. Orthop.,211: 30-35, 1986.21130  1986  [PubMed]
 
Accreditation Statement
These activities have been planned and implemented in accordance with the Essential Areas and policies of the Accreditation Council for Continuing Medical Education (ACCME) through the joint sponsorship of the American Academy of Orthopaedic Surgeons and The Journal of Bone and Joint Surgery, Inc. The American Academy of Orthopaedic Surgeons is accredited by the ACCME to provide continuing medical education for physicians.
CME Activities Associated with This Article
Submit a Comment
Please read the other comments before you post yours. Contributors must reveal any conflict of interest.
Comments are moderated and will appear on the site at the discretion of JBJS editorial staff.

* = Required Field
(if multiple authors, separate names by comma)
Example: John Doe




Related Articles
Related Cases
Related Content
Topic Collections
Hip
Related Audio and Videos
PubMed Articles
Clinical Trials
Readers of This Also Read...
jbjs jobs
12/22/2011
VA - Charleston Area Medical Center
12/22/2011
ME - Central Maine Medical Center