0
Commentary   |    
The Multifactorial Nature of Polyethylene Wear in Vivo*
THOMAS P. SCHMALZRIED, M.D.; FREDERICK J. DOREY, PH.D.; HARRY MCKELLOP, PH.D.
The Journal of Bone & Joint Surgery.  1998; 80:1234-42 
5 Recommendations (Recommend) | 3 Comments | Saved by 3 Users Save Case
In the 1980s, the focus of the joint-replacement industry was on fixation of the implants, and progress has been made both with devices that are inserted with cement and with those that are inserted without cement. As has always been the case in the history of joint replacement, each advance has unmasked another limitation. More durable fixation allows for more frequent, more intense, and more variable use of the joint, and the indications for the procedure have gradually been expanded to include younger and more active patients. With greater anticipated longevity for both patients and devices, there are justified concerns about long-term skeletal remodeling, but this has not yet been demonstrated to be a widespread clinical problem. The current problem is osteolysis, which most commonly occurs in association with polyethylene wear particles.
The clinical assessment of the performance of a bearing couple has traditionally been based on radiographic studies. With this approach, which presumably measures a change in the thickness of the polyethylene, there is a tendency to attribute the change in thickness to variables inherent to the polyethylene bearing. This is especially true when issues related to the manufacturing and sterilization of polyethylene have been put in the spotlight by both science and industry. We must be cautious about this tendency. As polyethylene wear is a focal issue in joint replacement, it is essential that we appreciate the complexities of studying such wear in vivo. The October 1997 issue of The Journal contained an article by Livingston et al. entitled, "Complications of Total Hip Arthroplasty Associated with the Use of an Acetabular Component with a Hylamer Liner."18 While we share the authors' concern about the in vivo performance of Hylamer, we wish to point out that there were differences in their comparison groups, in addition to the difference in the type of polyethylene, that also affect wear.
Livingston et al.18, referring to a previous report, compared the average rate of wear of six of 143 Hylamer liners, which had been specifically identified because of a high rate of wear, with the average rate of wear in several series rather than with the average rate of wear for the high-wear components in those series. There are components with high rates of wear in all series (Table I). For example, in one of the series referred to by Livingston et al., there were rates of wear of as high as 1.41 millimeters per year13. This rate is substantially higher than the average for that series and higher than the rates for the Hylamer liners reported on by Livingston et al. If one looks not only at average rates of linear wear but also at the range of rates of wear, one realizes that all studies that included a range demonstrated substantial variability (Table I). Regardless of the duration of follow-up, there are hip components that have no or only slight radiographically measurable wear and there are components that demonstrate wear that is several times greater than the average for that study. Such large case-to-case variations in rates of wear have not previously been explained simply by differences in the wear resistance of the polyethylene28.
Polyethylene wear in vivo is multifactorial with a complex interaction of variables; it is therefore not surprising that rates of polyethylene wear are highly variable. There are patient-related variables, such as age and gender, that are associated with the activity of the patient and the use of the prosthesis. There are variables related to the hip prostheses, which include all aspects of the acetabular and femoral implants (not just the polyethylene). There are also variables related to the operative procedure for implantation, which include operative techniques and the initial, as well as the long-term, fixation of the implants. Loosening of the implant can adversely affect wear, and vice versa. These variables are important as they can affect the loads on and the motions of the bearing and the degree of three-body-wear mechanisms. There is also variability due to differences in the method with which the wear is assessed and to limitations of the techniques for measurement.
In vivo rates of wear are higher in the short term and decrease with time for several reasons. Penetration of the femoral head into the acetabular polyethylene is due to a combination of creep (plastic deformation) and wear. Creep decreases exponentially with time, such that most of the penetration that occurs after the first twelve to eighteen months is due to wear. There is also an initial period of higher wear from so-called running-in of the bearing: with use, the contacting surface of the polyethylene wears into a higher degree of conformity with the specific femoral head with which it articulates, and this results in a larger contact surface, lower contact stresses, and a lower rate of wear26,27,33. Furthermore, with modular components, a change in the position of the polyethylene liner relative to the metal shell can also cause a change in the relative position of the femoral head27. These issues should be considered when rates of wear are compared among hips that have been followed for different amounts of time.
A fundamental limitation of all radiographic studies of wear is that clinical rates of wear have traditionally been expressed with the use of time as the denominator. This has been done for reasons of convenience, not accuracy. More appropriately, the number of cycles should be used as the denominator in in vitro laboratory studies of simulated wear. Similar to the wear of a set of automobile tires, the wear of a prosthetic hip is a function of use or the number of cycles; it is not simply a function of time. The assumption made in clinical studies is that the activity of patients who have a joint replacement—that is, the actual use of the joint or the number of cycles to which the bearing is subjected—is similar from patient to patient or, if it is not, any differences will tend to "average out" with a large sample size. In view of the broad range of patients who now have total joint implants, the limitations of this assumption must be recognized.
We reported on the walking activity, as measured with an electronic digital pedometer, of 111 patients who had a total joint replacement24. These patients averaged just a little more than 0.9 million cycles per year. The most important result, however, was that there was a forty-fivefold difference in the number of gait cycles between the least active and the most active patient. The most active patient averaged 3.2 million cycles per year, about 3.6 times higher than the average. Age was associated with daily walking activity (p = 0.048) but with a high degree of variability (standard deviation, 3040 steps per day). Patients who were less than sixty years old walked about 30 per cent more on the average than those who were sixty years old or more (p = 0.023). Men walked about 30 per cent more on the average than women (p = 0.037). Men who were less than sixty years old walked about 40 per cent more on the average than the other patients (p = 0.011). These data demonstrate that the activity of the patient can contribute substantially to the variability in rates of wear seen in in vivo studies.
Many clinical studies of wear have involved retrospective comparisons of wear on the basis of a specific variable, such as the type of polyethylene, the type of femoral head, the presence of a metal backing, or the type of acetabular or femoral fixation. The strength of the conclusions of these studies is limited because of the tremendous number of potentially confounding variables. Furthermore, caution should be exercised when the results are extrapolated to other reconstructions with the same generic variable. An example is the issue of metal backing. One study4 indicated an increase in the rate of polyethylene wear with a specific type of metal-backed acetabular component designed to be inserted with cement compared with that associated with an all-polyethylene component designed to be inserted with cement. In contrast, in other reports, a different type of metal-backed acetabular component was associated with lower rates of polyethylene wear3,14. The issue cannot be as simple as the presence or absence of metal backing. The apparently contradictory results in these series could be due to specific differences in the design and manufacture of the different types of metal-backed acetabular components or to the fact that the femoral heads were composed of titanium alloy in one study and of cobalt-chromium alloy in the others, or a combination of these or other factors.
Livingston et al. stated: "As with the initial six failed arthroplasties, there were no identifiable risk factors associated with the latter five."18 Youth has generally been considered a risk factor for wear. The average age of the eleven patients who had a revision because of a high rate of wear in the study by Livingston et al. was only forty-four years. Wear is a function of use, not time. With the same amount of time in situ, a specific bearing combination in a younger patient will, in general, show a greater linear penetration than the bearing combination in an older patient because of the greater amount of use, in general, by the younger patient24,27. Even though Livingston et al. acknowledged differences in the ages of their patients in Groups 1 (DePuy stem, cobalt-chromium femoral head, and Hylamer liner), 1a (DePuy stem, alumina femoral head, and Hylamer liner), 2 (Osteonics stem, cobalt-chromium femoral head, and Hylamer liner), and 3 (Osteonics stem, cobalt-chromium femoral head, and conventional polyethylene liner), they stated that age did not correlate with the rate of wear. However, Table I from their article indicates that a younger age is associated with a higher rate of wear, as does our plot of the linear wear rates versus the patient ages given in the study by Livingston et al. (Fig. 1). Group 2, the 138 hips in which the components of the implant had been made by different manufacturers, is an outlier. When the rate of wear is plotted against age for the groups in which both bearing components had been made by the same manufacturer, the correlation coefficient (r2) is greater than 0.99 (Figs. 2 and 3).
If there were an inherent difference between the wear resistances of the polymers, then one would expect a difference in the rates of wear between groups that were matched for other covariables, such as age. This is not the case in the study by Livingston et al.18. In the groups with similar ages and femoral components inserted with cement, Hylamer liners that articulated with a DePuy femoral head had virtually the same low rate of wear (average, 0.13 millimeter per year) as conventional polyethylene liners that articulated with an Osteonics femoral head (average, 0.12 millimeter per year). The patients in whom a Hylamer liner articulated with a DePuy femoral head were actually slightly younger than the patients who had a conventional polyethylene liner and an Osteonics femoral head. The low rate of wear of Hylamer in these patients suggests that there were factors other than the inherent wear resistance of the polymer that contributed to higher rates of wear in the other groups of patients.
The countersurface (the femoral head) is a factor that, in previous studies, has been demonstrated to affect polyethylene wear22,23,33. Despite having a slightly higher average age (62.8 compared with 58.3 years; average difference, 4.5 years), Group 2 (patients with an Osteonics cobalt-chromium femoral head) had a higher average rate of wear (0.29 compared with 0.20 millimeter per year) than Group 1 (patients with a DePuy cobalt-chromium femoral head)18. In this short-term study, the higher rate of linear penetration in the Group-2 hips may, at least in part, have been the result of a mismatch of manufacturing tolerances resulting in a higher degree of creep and running-in of the bearing combination in that group.
Another factor that can increase the rate of wear is suggested by Figures 3-A through 3-D in the article by Livingston et al.18. Although the published images are somewhat blurred, the radiograph made two years postoperatively suggests that there is radiolucency at the metal-cement interface, indicative of so-called debonding of the femoral component. This radiolucency was followed by progressive femoral endosteal osteolysis, but there was no apparent pelvic osteolysis. These features raise the possibility that loosening with local fragmentation of the cement mantle caused the femoral osteolysis. The position of Livingston et al. was that a high rate of wear of the bearing surface caused the femoral osteolysis. If the osteolysis were due to Hylamer wear particles, this would imply that the particles migrated to the femoral endosteum through the space between the stem and the cement mantle. Another explanation for the outcome in this patient, suggested by these radiographs, is that loosening of the femoral component at the stem-cement interface generated metal and cement particles, which could account for both the localized osteolysis and an increased rate of wear from three-body mechanisms.
Livingston et al.18 did not indicate any criteria for the selection of the different implants used by the five surgeons in their study, and they did not discuss any differences in outcome among the surgeons. The data indicate that ceramic femoral heads were used in the youngest patients; implants inserted without cement were preferentially used in younger patients; and, regardless of the femoral implant, Hylamer was used in younger patients. In retrospective studies such as the one by Livingston et al., there are limitations related to demographic variables such as age and gender. With surgeon selection bias, a particular implant or particular implants may be selected on the basis of the anticipated activity level of a particular patient, regardless of age, gender, or other demographic data. For example, a forty-year-old man who had symptomatic coronary artery disease would be expected to be less active than a sixty-five-year-old female aerobics instructor. It is probable that the surgeons in the study by Livingston et al. selected bearing implants that they believed to be more durable, such as Hylamer liners and ceramic femoral heads, for use in more active patients, regardless of age and gender. This would explain the apparently paradoxical result of higher wear associated with alumina ceramic heads.
Livingston et al.18 apparently presumed that failure to find a significant difference between two groups at the 95 per cent confidence level means that there is no difference between them11. This error affects the presentation of the data, as in Table I where the authors reported "NS," meaning that "no significant difference could be detected," for certain subgroups rather than reporting the average rates of wear for those subgroups. This suggests that the actual values were, in fact, not equal. Given the small size of some of the subgroups, the values could have differed substantially without having been significantly different.
To evaluate the relative wear resistance of different bearing materials, in vitro studies are conducted under carefully controlled and monitored conditions. The hip-simulator method of McKellop et al., which has been demonstrated to produce wear similar to that found in vivo20, demonstrated a 9 per cent reduction in the rate of wear of Hylamer compared with that of conventional GUR 415 polyethylene19. This difference had low significance (p = 0.5). One set of controlled conditions is used in hip-simulator studies. As we discussed, the conditions and the rates of wear in vivo are highly variable. From a materials-testing perspective, the report by Livingston et al.18 does not diminish the value of hip-simulator testing. It has been suggested that, because of greater stiffness, Hylamer may demonstrate higher rates of wear in a more abrasive environment, such as in the presence of hard third bodies. Furthermore, all of the Hylamer liners in the study by Livingston et al. were sterilized with gamma irradiation in air. Shelf-life is another clinical variable. For given oxidation levels, the values for ultimate tensile strength, elongation to breaking, and toughness of Hylamer have been shown to be lower than those for conventional polyethylene8. Such time-dependent effects of oxidation may result in higher in vivo rates of wear of these components. This time-dependent effect would not have been detected by the wear-simulator studies that have been performed on unaged Hylamer.
Over the next decade, we will see a number of studies that compare the clinical performance of one bearing with that of another. The issues that we outlined involve general principles and illustrate the complexities in the clinical evaluation of wear.
Thomas P. Schmalzried, M.D.
Frederick J. Dorey, Ph.D.
Harry McKellop, Ph.D.
Joint Replacement Institute
at Orthopaedic Hospital
2400 South Flower Street
Los Angeles, California 90007

*One or more of the authors has received or will receive benefits for personal or professional use from a commercial party related directly or indirectly to the subject of this commentary. In addition, benefits have been or will be directed to a research fund, foundation, educational institution, or other non-profit organization with which one or more of the authors is associated. Funds were received in total or partial support of the research or clinical study presented in this commentary. The funding source was DePuy-DuPont Orthopaedics.

*One or more of the authors has received or will receive benefits for personal or professional use from a commercial party related directly or indirectly to the subject of this commentary. In addition, benefits have been or will be directed to a research fund, foundation, educational institution, or other non-profit organization with which one or more of the authors is associated. Funds were received in total or partial support of the research or clinical study presented in this commentary. The funding source was DePuy-DuPont Orthopaedics.
 
Anchor for JumpAnchor for Jump  TABLE I STUDIES OF WEAR IN VIVO
*When not provided by authors, the volumetric wear was estimated from linear wear with the equation: v = (pr2h.
StudyType of Acetabular BearingType of Femoral HeadDiameter of Femoral Head (mm)No. of HipsWear RateComments
Linear (mm/yr.)Avg. Volumetric* (mm3/yr.)
Avg.Range
Charnley et al.7 (1969)Polytetrafluoro-ethyleneStainless steel22392.26859
Charnley and Cupic5 (1973)PolyethyleneStainless steel22720.1246
Charnley and Halley6 (1975)PolyethyleneStainless steel22720.150—0.657
Griffith et al.12 (1978)PolyethyleneStainless steel224930.070—0.2427
Wroblewski30 (1985)PolyethyleneStainless steel22210.210—0.4180Radiographic measurement
210.190—0.5272Direct measurement
Wroblewski31 (1986)PolyethyleneStainless steel221030.100—0.433615—21-yr. follow-up
Wroblewski et al.32 (1992)PolyethyleneStainless steel22570.070.01—0.22719—25-yr. follow-up
Wroblewski et al.33 (1996)Cross-linked polyethyleneAlumina22190.060.024—0.3222Avg. 77-mo. follow-up
140.2387First 18 mos.
140.0413Avg. 91-mo. follow-up
90.03138-yr. follow-up
Livermore et al.17 (1990)PolyethyleneStainless steel222270.130—0.3949
Stainless steel28980.080—0.3049
Cobalt-chromium32600.100—0.3280
Schmalzried et al.25 (1992)Polyethylene120.120.04—0.3036Autopsy
Kabo et al.16 (1993)PolyethyleneStainless steel2250.1348Direct measurement
Cobalt-chromium2630.23122
28230.23144
Cobalt-chromium3290.21172
36—54200.38314
Cates et al.4 (1993)PolyethyleneTitanium alloy28990.080—0.3749All polyethylene cup
Titanium alloy281340.110—0.3168Metal-backed cup
Hernandez et al.13 (1994)PolyethyleneTitanium alloy28970.140—0.9286Hybrid fixation
Titanium alloy281340.220—1.41135Fixation without cement
Bankston et al.1 (1995)PolyethyleneStainless steel28770.0637Matched patient groups
Cobalt-chromium28770.0531Matched patient groups
Titanium alloy28770.0849Matched patient groups
Callaghan et al.3 (1995)PolyethyleneStainless steel22230.12465-yr. follow-up, machined cup
Stainless steel22610.11425-yr. follow-up, molded cup
Cobalt-chromium28200.14865-yr. follow-up, molded cup
Cobalt-chromium28430.11685-yr. follow-up; molded, metal-backed cup
Cobalt-chromium28630.07435-yr. follow-up, machined cup, hybrid fixation
Cobalt-chromium28430.11687—8-yr. follow-up; molded, metal-backed cup
Stainless steel22230.124610-yr. follow-up, machined cup
Stainless steel22610.083010-yr. follow-up, molded cup
Cobalt-chromium28200.127410-yr. follow-up, molded cup
Stainless steel22230.114215-yr. follow-up, machined cup
Stainless steel22610.093415-yr. follow-up, molded cup
Stainless steel22230.103820—22-yr. follow-up, machined cup
Nashed et al.21 (1995)PolyethyleneTitanium alloy240.10Fixation with cement, polyethylene cup
Titanium alloy620.13Fixation with cement, metal-backed cup
Titanium alloy150.25Fixation without cement
Cobalt-chromium740.17Fixation without cement
Bankston et al.2 (1995)PolyethyleneCobalt-chromium28540.0531Compression molded cup
Cobalt-chromium28540.1168Machined bar-stock cup
Devane et al.9 (1995)PolyethyleneCobalt-chromium26, 321410.1502-dimensional computer technique
0.264793-dimensional computer technique
Woolson and Murphy29 (1995)PolyethyleneCobalt-chromium28800.140—0.3586Fixation of cup without cement
Hop et al.14 (1997)PolyethyleneStainless steel220.0832
Cobalt-chromium280.1484All polyethylene cup
Cobalt-chromium280.1059Metal-backed cup
0.0948Fixation with wire
0.1165Fixation with cable
Jasty et al.15 (1997)Polyethylene2230 (9—115)Autopsy
8460 (8—256)Fixation with cement
2290 (12—284)Metal-backed cup
32130 (10—819)Fixation without cement
Ohashi et al.22 (1989)PolyethyleneCobalt-chromium32130.0435
Stainless steel281060.0426
Alumina281870.0315
Okumura et al.23 (1989)PolyethyleneStainless steel220.1453
Alumina280.0849
Devane et al.10 (1997)PolyethyleneTitanium alloy28690.1522-dimensional computer technique, fixation with cement
0.228993-dimensional computer technique, fixation with cement
PolyethyleneTitanium alloy28700.2462-dimensional computer technique, fixation without cement
0.3581553-dimensional computer technique, fixation without cement
Shaver et al.26 (1997)PolyethyleneCobalt-chromium28430.154472-dimensional computer technique, short-term follow-up
0.087272-dimensional computer technique, steady-state wear
Sychterz et al.27 (1997)PolyethyleneCobalt-chromium32960.170.02—0.45137 (16—362)2-dimensional computer technique
Alumina3290.16
 
Anchor for JumpAnchor for Jump
+Figs. 1, 2, and 3: Plots of the linear wear rates versus the patient ages in the study by Livingston et al.18. Fig. 1: Data for all groups.
 
 
Anchor for JumpAnchor for Jump
+Fig. 3 Data for Groups 1 (with and without cement), 1a, and 3.
Bankston, A. B.; Cates, H.; Ritter, M. A.; Keating, E. M.; and Faris, P. M.: Polyethylene wear in total hip arthroplasty. Clin. Orthop.,317: 7-13, 1995.3177  1995  [PubMed]
 
Bankston, A. B.; Keating, E. M.; Ranawat, C.; Faris, P. M.; and Ritter, M. A.: Comparison of polyethylene wear in machined versus molded polyethylene. Clin. Orthop.,317: 37-43, 1995.31737  1995  [PubMed]
 
Callaghan, J. J.; Pedersen, D. R.; Olejniczak, J. P.; Goetz, D. D.; and Johnston, R. C.: Radiographic measurement of wear in 5 cohorts of patients observed for 5 to 22 years. Clin. Orthop.,317: 14-18, 1995.31714  1995  [PubMed]
 
Cates, H. E.; Faris, P. M.; Keating, E. M.; and Ritter, M. A.: Polyethylene wear in cemented metal-backed acetabular cups. J. Bone and Joint Surg.,75-B(2): 249-253, 1993.75-B(2)249  1993 
 
Charnley, J., and Cupic, Z.: The nine and ten year results of the low-friction arthroplasty of the hip. Clin. Orthop.,95: 9-25, 1973.959  1973  [PubMed]
 
Charnley, J., and Halley, D. K.: Rate of wear in total hip replacement. Clin. Orthop.,112: 170-179, 1975.112170  1975  [PubMed]
 
Charnley, J.; Kamangar, A.; and Longfield, M. D.: The optimum size of prosthetic heads in relation to the wear of plastic sockets in total replacement of the hip. Med. and Biol. Eng.,7: 31-39, 1969.731  1969 
 
Collier, J. P.; Bargmann, L. S.; Currier, B. H.; Mayor, M. B.; Currier, J. H.; and Bargmann, B. C.: An analysis of Hylamer® and polyethylene bearings from retrieved acetabular components. Orthopedics, in press, 1998. 
 
Devane, P. A.; Bourne, R. B.; Rorabeck, C. H.; MacDonald, S.; and Robinson, E. J.: Measurement of polyethylene wear in metal-backed acetabular cups. II. Clinical application. Clin. Orthop.,319: 317-326, 1995.319317  1995  [PubMed]
 
Devane, P. A.; Robinson, E. J.; Bourne, R. B.; Rorabeck, C. H.; Nayak, N. N.; and Horne, J. G.: Measurement of polyethylene wear in acetabular components inserted with and without cement. A randomized trial. J. Bone and Joint Surg.,79-A: 682-689, May 1997.79-A682  1997 
 
Ebramzadeh, E.; McKellop, H.; Dorey, F.; and Sarmiento, A.: Challenging the validity of conclusions based on p-values alone: a critique of contemporary clinical research design and methods. In Instructional Course Lectures, American Academy of Orthopaedic Surgeons. Vol. 43, pp. 587-598. Rosemont, Illinois, American Academy of Orthopaedic Surgeons, 1994. 
 
Griffith, M. J.; Seidenstein, M. K.; Williams, D.; and Charnley, J.: Socket wear in Charnley low friction arthroplasty of the hip. Clin. Orthop.,137: 37-47, 1978.13737  1978  [PubMed]
 
Hernandez, J. R.; Keating, E. M.; Faris, P. M.; Meding, J. B.; and Ritter, M. A.: Polyethylene wear in uncemented acetabular components. J. Bone and Joint Surg.,76-B(2): 263-266, 1994.76-B(2)263  1994 
 
Hop, J. D.; Callaghan, J. J.; Olejniczak, J. P.; Pedersen, D. R.; Brown, T. D.; and Johnston, R. C.: Contribution of cable debris generation to accelerated polyethylene wear. Clin. Orthop.,344: 20-32, 1997.34420  1997  [PubMed]
 
Jasty, M.; Goetz, D. D.; Bragdon, C. R.; Lee, K. R.; Hanson, A. E.; Elder, J. R.; and Harris, W. H.: Wear of polyethylene acetabular components in total hip arthroplasty. An analysis of one hundred and twenty-eight components retrieved at autopsy or revision operations. J. Bone and Joint Surg.,79-A: 349-358, March 1997.79-A349  1997 
 
Kabo, J. M.; Gebhard, J. S.; Loren, G.; and Amstutz, H. C.: In vivo wear of polyethylene acetabular components. J. Bone and Joint Surg.,75-B(2): 254-258, 1993.75-B(2)254  1993 
 
Livermore, J.; Ilstrup, D.; and Morrey, B.: Effect of femoral head size on wear of the polyethylene acetabular component. J. Bone and Joint Surg.,72-A: 518-528, April 1990.72-A518  1990 
 
Livingston, B. J.; Chmell, M. J.; Spector, M.; and Poss, R.: Complications of total hip arthroplasty associated with the use of an acetabular component with a Hylamer liner. J. Bone and Joint Surg.,79-A: 1529-1538, Oct. 1997.79-A1529  1997 
 
McKellop, H. A.; Lu, B.; and Li, S.: Wear of acetabular cups of conventional and modified UHMW polyethylenes compared on a hip joint simulator. Trans. Orthop. Res. Soc.,17: 356, 1992.17356  1992 
 
McKellop, H. A.; Campbell, P.; Park, S.-H.; Schmalzried, T. P.; Grigoris, P.; Amstutz, H. C.; and Sarmiento, A.: The origin of submicron polyethylene wear debris in total hip arthroplasty. Clin. Orthop.,311: 3-20, 1995.3113  1995  [PubMed]
 
Nashed, R. S.; Becker, D. A.; and Gustilo, R. B.: Are cementless acetabular components the cause of excess wear and osteolysis in total hip arthroplasty?. Clin. Orthop.,317: 19-28, 1995.31719  1995  [PubMed]
 
Ohashi, T.; Inoue, S.; and Kajikawa, K.: The clinical wear rate of acetabular component accompanied with alumina ceramic head. In Bioceramics, p. 278. Edited by H. Oonishi, H. Aoli, and K. Sawai. St. Louis, Ishiyaku Euro America, 1989. 
 
Okumura, H.; Yamamuro, T.; and Kumar, T.: Socket wear in total hip prosthesis with alumina ceramic head. In Biomedics: Proceedings of the First International Symposium on Bioceramics, p. 284. Edited by H. Oonishi, H. Aoli, and K. Sawai. St. Louis, Ishiyaku Euro America, 1989. 
 
Schmalzried, T. P.; Szuszczewicz, E. S.; Northfield, M. R.; Akizuki, K. H.; Frankel, R. E.; Belcher, G.; and Amstutz, H. C.: Quantitative assessment of walking activity after total hip or knee replacement. J. Bone and Joint Surg.,80-A: 54-59, Jan. 1998.80-A54  1998 
 
Schmalzried, T. P.; Kwong, L. M.; Jasty, M.; Sedlacek, R. C.; Haire, T. C.; O'Connor, D.; Bragdon, C. R.; Kabo, J. M.; Malcolm, A. J.; and Harris, W. H.: The mechanism of loosening of cemented acetabular components in total hip arthroplasty. Analysis of specimens retrieved at autopsy. Clin. Orthop.,274: 60-78, 1992.27460  1992  [PubMed]
 
Shaver, S. M.; Brown, T. D.; Hillis, S. L.; and Callaghan, J. J.: Digital edge-detection measurement of polyethylene wear after total hip arthroplasty. J. Bone and Joint Surg.,79-A: 690-700, May 1997.79-A690  1997 
 
Sychterz, C. J.; Engh, C. A., Jr.; Shah, N.; and Engh, C. A., Sr.: Radiographic evaluation of penetration by the femoral head into the polyethylene liner over time. J. Bone and Joint Surg.,79-A: 1040-1046, July 1997.79-A1040  1997 
 
Weightman, B.; Swanson, S. A. V.; Isaac, G. H.; and Wroblewski, B. M.: Polyethylene wear from retrieved acetabular cups. J. Bone and Joint Surg.,73-B(5): 806-810, 1991.73-B(5)806  1991 
 
Woolson, S. T., and Murphy, M. G.: Wear of the polyethylene of Harris-Galante acetabular components inserted without cement. J. Bone and Joint Surg.,77-A: 1311-1314, Sept. 1995.77-A1311  1995 
 
Wroblewski, B. M.: Direction and rate of socket wear in Charnley low-friction arthroplasty. J. Bone and Joint Surg.,67-B(5): 757-761, 1985.67-B(5)757  1985 
 
Wroblewski, B. M.: 15-21-year results of the Charnley low-friction arthroplasty. Clin. Orthop.,211: 30-35, 1986.21130  1986  [PubMed]
 
Wroblewski, B. M.; Taylor, G. W.; and Siney, P.: Charnley low-friction arthroplasty: 19-25-year results. Orthopedics,15: 421-424, 1992.15421  1992  [PubMed]
 
Wroblewski, B. M.; Siney, P. D.; Dowson, D.; and Collins, S. N.: Prospective clinical and joint simulator studies of a new total hip arthroplasty using alumina ceramic heads and cross-linked polyethylene cups. J. Bone and Joint Surg.,78-B(2): 280-285, 1996.78-B(2)280  1996 
 

Submit a comment

Topics

Anchor for JumpAnchor for Jump
+Figs. 1, 2, and 3: Plots of the linear wear rates versus the patient ages in the study by Livingston et al.18. Fig. 1: Data for all groups.
Anchor for JumpAnchor for Jump
+Fig. 3 Data for Groups 1 (with and without cement), 1a, and 3.
Anchor for JumpAnchor for Jump  TABLE I STUDIES OF WEAR IN VIVO
*When not provided by authors, the volumetric wear was estimated from linear wear with the equation: v = (pr2h.
StudyType of Acetabular BearingType of Femoral HeadDiameter of Femoral Head (mm)No. of HipsWear RateComments
Linear (mm/yr.)Avg. Volumetric* (mm3/yr.)
Avg.Range
Charnley et al.7 (1969)Polytetrafluoro-ethyleneStainless steel22392.26859
Charnley and Cupic5 (1973)PolyethyleneStainless steel22720.1246
Charnley and Halley6 (1975)PolyethyleneStainless steel22720.150—0.657
Griffith et al.12 (1978)PolyethyleneStainless steel224930.070—0.2427
Wroblewski30 (1985)PolyethyleneStainless steel22210.210—0.4180Radiographic measurement
210.190—0.5272Direct measurement
Wroblewski31 (1986)PolyethyleneStainless steel221030.100—0.433615—21-yr. follow-up
Wroblewski et al.32 (1992)PolyethyleneStainless steel22570.070.01—0.22719—25-yr. follow-up
Wroblewski et al.33 (1996)Cross-linked polyethyleneAlumina22190.060.024—0.3222Avg. 77-mo. follow-up
140.2387First 18 mos.
140.0413Avg. 91-mo. follow-up
90.03138-yr. follow-up
Livermore et al.17 (1990)PolyethyleneStainless steel222270.130—0.3949
Stainless steel28980.080—0.3049
Cobalt-chromium32600.100—0.3280
Schmalzried et al.25 (1992)Polyethylene120.120.04—0.3036Autopsy
Kabo et al.16 (1993)PolyethyleneStainless steel2250.1348Direct measurement
Cobalt-chromium2630.23122
28230.23144
Cobalt-chromium3290.21172
36—54200.38314
Cates et al.4 (1993)PolyethyleneTitanium alloy28990.080—0.3749All polyethylene cup
Titanium alloy281340.110—0.3168Metal-backed cup
Hernandez et al.13 (1994)PolyethyleneTitanium alloy28970.140—0.9286Hybrid fixation
Titanium alloy281340.220—1.41135Fixation without cement
Bankston et al.1 (1995)PolyethyleneStainless steel28770.0637Matched patient groups
Cobalt-chromium28770.0531Matched patient groups
Titanium alloy28770.0849Matched patient groups
Callaghan et al.3 (1995)PolyethyleneStainless steel22230.12465-yr. follow-up, machined cup
Stainless steel22610.11425-yr. follow-up, molded cup
Cobalt-chromium28200.14865-yr. follow-up, molded cup
Cobalt-chromium28430.11685-yr. follow-up; molded, metal-backed cup
Cobalt-chromium28630.07435-yr. follow-up, machined cup, hybrid fixation
Cobalt-chromium28430.11687—8-yr. follow-up; molded, metal-backed cup
Stainless steel22230.124610-yr. follow-up, machined cup
Stainless steel22610.083010-yr. follow-up, molded cup
Cobalt-chromium28200.127410-yr. follow-up, molded cup
Stainless steel22230.114215-yr. follow-up, machined cup
Stainless steel22610.093415-yr. follow-up, molded cup
Stainless steel22230.103820—22-yr. follow-up, machined cup
Nashed et al.21 (1995)PolyethyleneTitanium alloy240.10Fixation with cement, polyethylene cup
Titanium alloy620.13Fixation with cement, metal-backed cup
Titanium alloy150.25Fixation without cement
Cobalt-chromium740.17Fixation without cement
Bankston et al.2 (1995)PolyethyleneCobalt-chromium28540.0531Compression molded cup
Cobalt-chromium28540.1168Machined bar-stock cup
Devane et al.9 (1995)PolyethyleneCobalt-chromium26, 321410.1502-dimensional computer technique
0.264793-dimensional computer technique
Woolson and Murphy29 (1995)PolyethyleneCobalt-chromium28800.140—0.3586Fixation of cup without cement
Hop et al.14 (1997)PolyethyleneStainless steel220.0832
Cobalt-chromium280.1484All polyethylene cup
Cobalt-chromium280.1059Metal-backed cup
0.0948Fixation with wire
0.1165Fixation with cable
Jasty et al.15 (1997)Polyethylene2230 (9—115)Autopsy
8460 (8—256)Fixation with cement
2290 (12—284)Metal-backed cup
32130 (10—819)Fixation without cement
Ohashi et al.22 (1989)PolyethyleneCobalt-chromium32130.0435
Stainless steel281060.0426
Alumina281870.0315
Okumura et al.23 (1989)PolyethyleneStainless steel220.1453
Alumina280.0849
Devane et al.10 (1997)PolyethyleneTitanium alloy28690.1522-dimensional computer technique, fixation with cement
0.228993-dimensional computer technique, fixation with cement
PolyethyleneTitanium alloy28700.2462-dimensional computer technique, fixation without cement
0.3581553-dimensional computer technique, fixation without cement
Shaver et al.26 (1997)PolyethyleneCobalt-chromium28430.154472-dimensional computer technique, short-term follow-up
0.087272-dimensional computer technique, steady-state wear
Sychterz et al.27 (1997)PolyethyleneCobalt-chromium32960.170.02—0.45137 (16—362)2-dimensional computer technique
Alumina3290.16
Bankston, A. B.; Cates, H.; Ritter, M. A.; Keating, E. M.; and Faris, P. M.: Polyethylene wear in total hip arthroplasty. Clin. Orthop.,317: 7-13, 1995.3177  1995  [PubMed]
 
Bankston, A. B.; Keating, E. M.; Ranawat, C.; Faris, P. M.; and Ritter, M. A.: Comparison of polyethylene wear in machined versus molded polyethylene. Clin. Orthop.,317: 37-43, 1995.31737  1995  [PubMed]
 
Callaghan, J. J.; Pedersen, D. R.; Olejniczak, J. P.; Goetz, D. D.; and Johnston, R. C.: Radiographic measurement of wear in 5 cohorts of patients observed for 5 to 22 years. Clin. Orthop.,317: 14-18, 1995.31714  1995  [PubMed]
 
Cates, H. E.; Faris, P. M.; Keating, E. M.; and Ritter, M. A.: Polyethylene wear in cemented metal-backed acetabular cups. J. Bone and Joint Surg.,75-B(2): 249-253, 1993.75-B(2)249  1993 
 
Charnley, J., and Cupic, Z.: The nine and ten year results of the low-friction arthroplasty of the hip. Clin. Orthop.,95: 9-25, 1973.959  1973  [PubMed]
 
Charnley, J., and Halley, D. K.: Rate of wear in total hip replacement. Clin. Orthop.,112: 170-179, 1975.112170  1975  [PubMed]
 
Charnley, J.; Kamangar, A.; and Longfield, M. D.: The optimum size of prosthetic heads in relation to the wear of plastic sockets in total replacement of the hip. Med. and Biol. Eng.,7: 31-39, 1969.731  1969 
 
Collier, J. P.; Bargmann, L. S.; Currier, B. H.; Mayor, M. B.; Currier, J. H.; and Bargmann, B. C.: An analysis of Hylamer® and polyethylene bearings from retrieved acetabular components. Orthopedics, in press, 1998. 
 
Devane, P. A.; Bourne, R. B.; Rorabeck, C. H.; MacDonald, S.; and Robinson, E. J.: Measurement of polyethylene wear in metal-backed acetabular cups. II. Clinical application. Clin. Orthop.,319: 317-326, 1995.319317  1995  [PubMed]
 
Devane, P. A.; Robinson, E. J.; Bourne, R. B.; Rorabeck, C. H.; Nayak, N. N.; and Horne, J. G.: Measurement of polyethylene wear in acetabular components inserted with and without cement. A randomized trial. J. Bone and Joint Surg.,79-A: 682-689, May 1997.79-A682  1997 
 
Ebramzadeh, E.; McKellop, H.; Dorey, F.; and Sarmiento, A.: Challenging the validity of conclusions based on p-values alone: a critique of contemporary clinical research design and methods. In Instructional Course Lectures, American Academy of Orthopaedic Surgeons. Vol. 43, pp. 587-598. Rosemont, Illinois, American Academy of Orthopaedic Surgeons, 1994. 
 
Griffith, M. J.; Seidenstein, M. K.; Williams, D.; and Charnley, J.: Socket wear in Charnley low friction arthroplasty of the hip. Clin. Orthop.,137: 37-47, 1978.13737  1978  [PubMed]
 
Hernandez, J. R.; Keating, E. M.; Faris, P. M.; Meding, J. B.; and Ritter, M. A.: Polyethylene wear in uncemented acetabular components. J. Bone and Joint Surg.,76-B(2): 263-266, 1994.76-B(2)263  1994 
 
Hop, J. D.; Callaghan, J. J.; Olejniczak, J. P.; Pedersen, D. R.; Brown, T. D.; and Johnston, R. C.: Contribution of cable debris generation to accelerated polyethylene wear. Clin. Orthop.,344: 20-32, 1997.34420  1997  [PubMed]
 
Jasty, M.; Goetz, D. D.; Bragdon, C. R.; Lee, K. R.; Hanson, A. E.; Elder, J. R.; and Harris, W. H.: Wear of polyethylene acetabular components in total hip arthroplasty. An analysis of one hundred and twenty-eight components retrieved at autopsy or revision operations. J. Bone and Joint Surg.,79-A: 349-358, March 1997.79-A349  1997 
 
Kabo, J. M.; Gebhard, J. S.; Loren, G.; and Amstutz, H. C.: In vivo wear of polyethylene acetabular components. J. Bone and Joint Surg.,75-B(2): 254-258, 1993.75-B(2)254  1993 
 
Livermore, J.; Ilstrup, D.; and Morrey, B.: Effect of femoral head size on wear of the polyethylene acetabular component. J. Bone and Joint Surg.,72-A: 518-528, April 1990.72-A518  1990 
 
Livingston, B. J.; Chmell, M. J.; Spector, M.; and Poss, R.: Complications of total hip arthroplasty associated with the use of an acetabular component with a Hylamer liner. J. Bone and Joint Surg.,79-A: 1529-1538, Oct. 1997.79-A1529  1997 
 
McKellop, H. A.; Lu, B.; and Li, S.: Wear of acetabular cups of conventional and modified UHMW polyethylenes compared on a hip joint simulator. Trans. Orthop. Res. Soc.,17: 356, 1992.17356  1992 
 
McKellop, H. A.; Campbell, P.; Park, S.-H.; Schmalzried, T. P.; Grigoris, P.; Amstutz, H. C.; and Sarmiento, A.: The origin of submicron polyethylene wear debris in total hip arthroplasty. Clin. Orthop.,311: 3-20, 1995.3113  1995  [PubMed]
 
Nashed, R. S.; Becker, D. A.; and Gustilo, R. B.: Are cementless acetabular components the cause of excess wear and osteolysis in total hip arthroplasty?. Clin. Orthop.,317: 19-28, 1995.31719  1995  [PubMed]
 
Ohashi, T.; Inoue, S.; and Kajikawa, K.: The clinical wear rate of acetabular component accompanied with alumina ceramic head. In Bioceramics, p. 278. Edited by H. Oonishi, H. Aoli, and K. Sawai. St. Louis, Ishiyaku Euro America, 1989. 
 
Okumura, H.; Yamamuro, T.; and Kumar, T.: Socket wear in total hip prosthesis with alumina ceramic head. In Biomedics: Proceedings of the First International Symposium on Bioceramics, p. 284. Edited by H. Oonishi, H. Aoli, and K. Sawai. St. Louis, Ishiyaku Euro America, 1989. 
 
Schmalzried, T. P.; Szuszczewicz, E. S.; Northfield, M. R.; Akizuki, K. H.; Frankel, R. E.; Belcher, G.; and Amstutz, H. C.: Quantitative assessment of walking activity after total hip or knee replacement. J. Bone and Joint Surg.,80-A: 54-59, Jan. 1998.80-A54  1998 
 
Schmalzried, T. P.; Kwong, L. M.; Jasty, M.; Sedlacek, R. C.; Haire, T. C.; O'Connor, D.; Bragdon, C. R.; Kabo, J. M.; Malcolm, A. J.; and Harris, W. H.: The mechanism of loosening of cemented acetabular components in total hip arthroplasty. Analysis of specimens retrieved at autopsy. Clin. Orthop.,274: 60-78, 1992.27460  1992  [PubMed]
 
Shaver, S. M.; Brown, T. D.; Hillis, S. L.; and Callaghan, J. J.: Digital edge-detection measurement of polyethylene wear after total hip arthroplasty. J. Bone and Joint Surg.,79-A: 690-700, May 1997.79-A690  1997 
 
Sychterz, C. J.; Engh, C. A., Jr.; Shah, N.; and Engh, C. A., Sr.: Radiographic evaluation of penetration by the femoral head into the polyethylene liner over time. J. Bone and Joint Surg.,79-A: 1040-1046, July 1997.79-A1040  1997 
 
Weightman, B.; Swanson, S. A. V.; Isaac, G. H.; and Wroblewski, B. M.: Polyethylene wear from retrieved acetabular cups. J. Bone and Joint Surg.,73-B(5): 806-810, 1991.73-B(5)806  1991 
 
Woolson, S. T., and Murphy, M. G.: Wear of the polyethylene of Harris-Galante acetabular components inserted without cement. J. Bone and Joint Surg.,77-A: 1311-1314, Sept. 1995.77-A1311  1995 
 
Wroblewski, B. M.: Direction and rate of socket wear in Charnley low-friction arthroplasty. J. Bone and Joint Surg.,67-B(5): 757-761, 1985.67-B(5)757  1985 
 
Wroblewski, B. M.: 15-21-year results of the Charnley low-friction arthroplasty. Clin. Orthop.,211: 30-35, 1986.21130  1986  [PubMed]
 
Wroblewski, B. M.; Taylor, G. W.; and Siney, P.: Charnley low-friction arthroplasty: 19-25-year results. Orthopedics,15: 421-424, 1992.15421  1992  [PubMed]
 
Wroblewski, B. M.; Siney, P. D.; Dowson, D.; and Collins, S. N.: Prospective clinical and joint simulator studies of a new total hip arthroplasty using alumina ceramic heads and cross-linked polyethylene cups. J. Bone and Joint Surg.,78-B(2): 280-285, 1996.78-B(2)280  1996 
 
Accreditation Statement
These activities have been planned and implemented in accordance with the Essential Areas and policies of the Accreditation Council for Continuing Medical Education (ACCME) through the joint sponsorship of the American Academy of Orthopaedic Surgeons and The Journal of Bone and Joint Surgery, Inc. The American Academy of Orthopaedic Surgeons is accredited by the ACCME to provide continuing medical education for physicians.
CME Activities Associated with This Article
Submit a Comment
Please read the other comments before you post yours. Contributors must reveal any conflict of interest.
Comments are moderated and will appear on the site at the discretion of JBJS editorial staff.

* = Required Field
(if multiple authors, separate names by comma)
Example: John Doe




Related Articles
Related Cases
Related Content
Topic Collections
Hip
Related Audio and Videos
PubMed Articles
Groin pain after replacement of the hip: Aetiology, evaluation and treatment.
The Journal of bone and joint surgery. British volume: Issue date- 2012 Feb
Biomechanical Testing of Suture-Based Meniscal Repair Devices Containing Ultrahigh-Molecular-Weight Polyethylene Suture: Update 2011.
Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association: Issue date- 2012 Feb 6
Clinical Trials
Readers of This Also Read...
jbjs jobs
12/22/2011
VA - Charleston Area Medical Center
12/22/2011
ME - Central Maine Medical Center
12/22/2011
Maine - Central Maine Medical Center