Effect of Sterilization Method and Other Modifications on the Wear Resistance of Acetabular Cups Made of Ultra-High Molecular Weight Polyethylene A Hip-Simulator Study*
Harry McKellop, Ph.D.†; Fu-wen Shen, Ph.D.†; Bin Lu, M.S.†; Patricia Campbell, Ph.D.†; Ronald Salovey, Ph.D.‡
View Disclosures and Other Information
Investigation performed at the J. Vernon Luck Orthopaedic Research Center and the Joint Replacement Institute, Los Angeles Orthopaedic Hospital, and the Departments of Orthopaedics, Biomedical Engineering, and Materials Science, University of Southern California, Los Angeles, California
*No benefits in any form have been received or will be received from a commercial party related directly or indirectly to the subject of this article. Funds were received in total or partial support of the research or clinical study presented in this article. The funding sources were the National Institutes of Health Grant 40996 and the Los Angeles Orthopaedic Hospital Foundation. Materials were donated by DePuy-DuPont Orthopaedics; Howmedica, Incorporated; Intermedics Orthopedics, Incorporated; Poly Hi Solidur, Incorporated; Spire Corporation; and Zimmer, Incorporated.
†The J. Vernon Luck Orthopaedic Research Center (H. McK., F.-W. S., and B. L.) and the Joint Replacement Institute (P. C.), Orthopaedic Hospital, 2400 South Flower Street, Los Angeles, California 90007. E-mail address for H. McKellop: hmckellop@laoh.ucla.edu.
‡Department of Materials Science, University of Southern California, University Park Campus, Los Angeles, California 90089.

The Journal of Bone & Joint Surgery.  2000; 82:1708-1708 
5 Recommendations (Recommend) | 3 Comments | Saved by 3 Users Save Case

Abstract

Background: Wear of ultra-high molecular weight polyethylene acetabular cups in hip prostheses produces billions of submicrometer wear particles annually that can cause osteolysis and loosening of the components. Thus, substantial improvement of the wear resistance of ultra-high molecular weight polyethylene could extend the clinical life span of total hip prostheses. It has become apparent that the conditions under which ultra-high molecular weight polyethylene cups have been sterilized can markedly affect their long-term wear properties, and new sterilization methods and other modifications have been developed to minimize the negative effects.

Methods: In the present study, a hip-joint simulator was used to assess whether it is preferable to sterilize ultra-high molecular weight polyethylene cups without gamma irraSdiation, to avoid radiation-induced oxidative degradation, or to sterilize with gamma irradiation while the cups are packaged in a suitable low-oxygen atmosphere to minimize oxidation while retaining the increased wear resistance conferred by the radiation-induced cross-linking. Ion-implanted cups and cups made of a highly crystalline polyethylene (Hylamer) also were investigated. Cups made of each material were subjected to wear-testing prior to and after artificial thermal aging to accelerate oxidative degradation.

Results: The results of the present study demonstrated that the cross-linking induced by gamma irradiation improves the wear resistance of ultra-high molecular weight polyethylene, while oxidation reduces it. Without thermal aging, the two types of cups that were sterilized with gamma irradiation while in low-oxygen packaging exhibited about a 50 percent lower rate of wear than did either the nonsterilized cups or the nonirradiated cups sterilized with gas plasma. There was a comparable advantage in the rate of wear after fourteen days of thermal aging. However, after thirty days of aging, the cups sterilized with gamma irradiation in low-oxygen packaging wore several times faster than did the nonirradiated cups. Ion-implanting improved the wear resistance without thermal aging, but after extensive thermal aging the oxidation and wear were greater than those of the controls. Hylamer cups (that is, those that were sterilized with gas plasma) exhibited wear properties very close to those of the nonsterilized ultra-high molecular weight polyethylene cups (the controls) with or without aging.

Conclusions: Sterilizing an ultra-high molecular weight polyethylene acetabular cup without radiation (for example, with ethylene oxide or gas plasma) avoids immediate and long-term oxidative degradation of the implant but does not improve the inherent wear resistance of the polyethylene. Sterilizing with use of gamma irradiation with the implant packaged in a low-oxygen atmosphere avoids immediate oxidation and cross-links the polyethylene, thereby increasing its wear resistance, but long-term oxidation of the residual free radicals may markedly reduce the wear resistance. Ideally, cross-linking with gamma irradiation to reduce wear should be done in a manner that avoids both immediate and long-term oxidation.

Clinical Relevance: The present study demonstrated how the fabrication and sterilization processes influence the resistance to oxidation and wear of the various types of ultra-high molecular weight polyethylene that are currently available. As an exact quantitative relationship between days of thermal aging and years of real-time aging (on the shelf and/or in vivo) has not yet been established, it is not possible to predict precisely when, if ever, the in vivo wear rate of cups sterilized with gamma irradiation while in low-oxygen packaging would exceed that of nonirradiated cups. Nevertheless, the results of these wear tests with use of a hip simulator suggest that, for at least ten years of clinical use, the in vivo wear rate of cups sterilized with gamma irradiation while in low-oxygen packaging will be substantially lower than that of cups sterilized without irradiation. The fundamental interactions among radiation, cross-linking, and oxidation exhibited by the specific materials included in the present study may also apply to acetabular cups of other types of polyethylene. Understanding these fundamental interactions will assist the surgeon in making an informed choice among the materials examined in the present study and among other types of modified polyethylene already in clinical use, including those sterilized with ethylene oxide, those sterilized with gamma irradiation in other forms of low-oxygen packaging, and the various new cross-linked and thermally stabilized polyethylenes.

Figures in this Article
    Wear of the polyethylene bearing surfaces of prosthetic hip joints produces billions of submicrometer wear particles annually7,26,36, often causing a foreign-body response that may lead to bone resorption (osteolysis) and loosening of the components1,18,20,24,53,71. This is of particular concern for young and/or active patients as they may face one or more revisions with cumulative bone loss in their lifetime. Thus, a substantial improvement in the wear resistance of the polyethylene could substantially extend the clinical life span of total hip prostheses.
    The majority of ultra-high molecular weight polyethylene components implanted in the past two decades were sterilized by exposure to gamma radiation at doses between 2.5 and 4.0 megarad (twenty-five and forty kilogray), with the components packaged in air. It is now known that the oxygen that is present in the polyethylene when it is irradiated, or that diffuses into the material during shelf storage and/or during in vivo use, reacts with residual free radicals that are generated by the radiation13,29,47,48,72. The resultant chain scission reduces the molecular weight, leading to increased density, stiffness and brittleness, and reduced fracture strength and elongation to failure4,8,13,17,49,63. These changes, in turn, reduce the resistance to wear50,57.
    On the other hand, particularly if little oxygen is present, the free radicals generated during irradiation can form carbon-carbon cross-links between adjacent polyethylene molecules, and a number of laboratory tests have shown that, in an appropriate amount, such cross-linking improves the wear resistance of the polymer12,22,44,51. Consequently, several manufacturers currently seal their polyethylene components in various types of low-oxygen packaging in order to minimize the detrimental effects of oxidation during sterilization with irradiation and during subsequent shelf storage2,60,21. Some manufacturers take the additional step of heating acetabular polyethylene cups after irradiation to reduce the level of residual free radicals, thereby reducing subsequent oxidation69. Other manufacturers advocate sterilization without irradiation - for example, sterilization with ethylene oxide66 or gas plasma15 - since, even with irradiation in a low-oxygen package, free radicals that can lead to oxidation during prolonged shelf storage and/or in vivo use remain in the polymer. However, while sterilization without irradiation avoids the problem of oxidation of the free radicals entirely, it also eliminates any improvement of the wear resistance that might be conferred by radiation-induced cross-linking.
    Since surgeons performing hip arthroplasty presently must choose between these two fundamentally different approaches to sterilization, it is important to assess the relative resistance to wear, including the potential effects of long-term oxidative degradation, of acetabular cups sterilized by each method. In the present study, a hip-joint simulator was used to compare the resistance to wear of acetabular cups prepared by five manufacturers with use of methods that included sterilization without irradiation and sterilization with irradiation in a low-oxygen atmosphere, along with other modifications. The cups were tested prior to and after a thermal aging treatment designed to accelerate oxidative degradation.
     
    Anchor for JumpAnchor for Jump
    +Fig. 1:Flowchart showing the production of the ultra-high molecular weight polyethylene (UHMWPE) as modified by the orthopaedic implant manufacturers. Beginning with a single lot of extruded bar stock produced from a single batch of GUR 4150 resin, the polyethylene was routed through the five participating manufacturers. The cups that were used as nonsterilized controls and those sterilized with irradiation in air were machined by Zimmer as a service to the study, but they do not represent Zimmer's current product. All of the gamma sterilization was done with the cups in a single container to provide an equal, simultaneous radiation dose.
     
    Anchor for JumpAnchor for Jump
    +Fig. 2:Graph showing the oxidation levels from the inner bearing surface to the center of the cup wall (left to right) obtained from an ultra-high molecular weight polyethylene cup after seventeen years of shelf-aging and from the nonworn zone of an ultra-high molecular weight polyethylene cup retrieved after thirteen years of use in vivo. Both cups were originally sterilized with gamma irradiation in air. These oxidation levels are among the highest that have been measured in our laboratory for cups sterilized with gamma irradiation in air. (The microtomed sections from the retrieved cup were soaked in hexane for sixteen hours to reduce contaminants prior to Fourier transform infrared analysis.)
     
    Anchor for JumpAnchor for Jump
    +Fig. 3:Graph showing the oxidation levels from the inner surface to the center of the cup wall (left to right) in the nonworn zones of the cups tested without artificial aging. (The corresponding wear volumes and rates are shown in Figure 4-A and Figure 4-B.) Two cups of each material were analyzed. Since the oxidation profiles were comparable for the two cups in a pair, only the mean oxidation levels are shown for clarity.
     
    Anchor for JumpAnchor for Jump
    +Fig. 4-A:Graph showing the volumetric wear of the two cups of each material tested without artificial aging. (The corresponding oxidation profiles are shown in Figure 3.) The scale at right indicates the approximate maximum depth of wear - that is, the depth at the center of the contact zone.
     
    Anchor for JumpAnchor for Jump
    +Fig. 4-B:Graph showing the mean wear rates and the standard deviations for the six different types of polyethylene and for the two cycle intervals without artificial aging.
     
    Anchor for JumpAnchor for Jump
    +Fig. 5:Graph showing the oxidation levels from the inner surface to the center of the cup wall (left to right) in the nonworn zones of the cups tested after fourteen days of artificial aging at 80 degrees Celsius. (The corresponding wear volumes and rates are shown in Figure 6-A and Figure 6-B.) Since the levels of oxidation were comparable for the two cups in a pair, only the mean oxidation levels are shown for each material for clarity.
     
    Anchor for JumpAnchor for Jump
    +Fig. 6-A:Graph showing the volumetric wear of the two cups of each material tested after fourteen days of artificial aging at 80 degrees Celsius. (The corresponding oxidation profiles are shown in Figure 5.) The scale at right indicates the approximate maximum depth of wear - that is, the depth at the center of the contact zone.
     
    Anchor for JumpAnchor for Jump
    +Fig. 6-B:Graph showing the mean wear rates and the standard deviations for the six different types of polyethylene and for the two cycle intervals after fourteen days of artificial aging at 80 degrees Celsius.
     
    Anchor for JumpAnchor for Jump
    +Fig. 7:Graph showing the oxidation levels from the inner surface to the center of the cup wall (left to right) in the nonworn zones of the cups tested after thirty days of artificial aging at 80 degrees Celsius. (The corresponding wear volumes and rates are shown in Figure 8-A and Figure 8-B.) When the oxidation profiles were comparable for the two cups in a pair, only the mean oxidation levels are shown for clarity. The profiles are shown for both cups made of ion-implanted polyethylene and both made of polyethylene irradiated with an oxygen scavenger since the two cups in each of these pairs exhibited substantially different oxidation levels.
     
    Anchor for JumpAnchor for Jump
    +Fig. 8-A:Graph showing the volumetric wear of the two cups of each material tested after thirty days of artificial aging at 80 degrees Celsius. (The corresponding oxidation profiles are shown in Figure 7.) The scale at right indicates the approximate maximum depth of wear - that is, the depth at the center of the contact zone.
     
    Anchor for JumpAnchor for Jump
    +Fig. 8-B:Graph showing the mean wear rates and the standard deviations for the six different types of polyethylene and for the three cycle intervals after thirty days of artificial aging at 80 degrees Celsius.
     
    Anchor for JumpAnchor for JumpTABLE I:  Modified Polyethylene as Fabricated by Orthopaedic Manufacturers
    *All of the cups sterilized by gamma irradiation were placed in a single container and irradiated to 2.72 megarad (27.2 kilogray) by a commercial sterilizer.†The Abtox Plazlyte sterilization system is manufactured by Abtox, Mundelein, Illinois.
    Type of PolyethyleneGroup SymbolFabrication Method and SourceSterilization Method*
    Nonsterilized controlsCMachined from extruded bar stock at Zimmer None
    Hylamer, gas plasmaHExtruded bar stock converted to Hylamer at DePuy-DuPont Orthopaedics; cups machined from bar stockGas-plasma sterilized by DePuy-DuPont with use of an Abtox Plazlyte unit†
    Irradiated in airAMachined from extruded bar stock at ZimmerGamma-irradiated in air to 2.72 megarad (27.2 kilogray)
    Ion-implanted (IONGUARD III and irradiated in air IMachined from extruded bar stock at Zimmer; implanted with nitrogen ions at Spire Gamma-irradiated in air to 2.72 megarad (27.2 kilogray)
    Irradiated in nitrogen and thermally stabilized (Duration)NMachined from extruded bar stock at Howmedica Packaged in nitrogen and gamma- irradiated to 2.72 megarad (27.2 kilogray), then returned to Howmedica for thermal stabilization by heating to 50 degrees Celsius for 6 days while still packaged in nitrogen
    Irradiated with oxygen scavenger (Oxygenless Packaging)OMachined from extruded bar stock at Intermedics OrthopedicsPackaged in foil with an oxygen scavenger (nontoxic, activated iron particles) in a Tyvek (high-density polyethylene) pouch, and gamma- irradiated to 2.72 megarad (27.2 kilogray)
     
    Anchor for JumpAnchor for JumpTABLE II:  Wear Rates of Acetabular Cups without Artificial Aging
    *The values are given in cubic millimeters per million cycles.
    Type of CupCup No.Between 0.5 and 3.0 Million Cycles*Between 3.0 and 5.0 Million Cycles*
    Wear RateMean and Standard DeviationWear RateMean and Standard Deviation
    Nonsterilized controlsC1C234.936.935.9 ± 1.440.642.141.4 ± 1.0
    Hylamer, gas plasmaH1H235.034.034.5 ± 0.736.233.334.8 ± 2.1
    Irradiated in airA1A237.636.036.8 ± 1.126.127.326.7 ± 0.8
    Ion-implanted (IONGUARD III) and irradiated in air I1I225.025.325.2 ± 0.224.224.124.2 ± 0.1
    Irradiated in nitrogen and thermally stabilized (Duration)N1N218.615.617.1 ± 2.123.019.921.5 ± 2.2
    Irradiated with oxygen scavenger (Oxygenless Packaging)O1O218.616.917.8 ± 1.218.517.117.8 ± 1.0
     
    Anchor for JumpAnchor for JumpTABLE III:  Comparisons of the Wear Rates of Various Materials without Artificial Aging
    *C = nonsterilized controls; H = Hylamer, gas plasma; A = irradiated in air; I = ion-implanted and irradiated in air; N = irradiated in nitrogen and thermally stabilized; and O = irradiated with oxygen scavenger.†Power is shown in parentheses.
    Compared Groups*P Value for Difference Between Groups†
    At 0.5 to 3.0 million cycles
      C > H0.35 (99.9%)
      C < A0.56
      C > I0.009
      C > N0.009
      C > O0.005
      H < A0.14
      H > I0.003
      H > N0.008
      H > O0.004
      A > I0.005
      A > N0.007
      A > O0.004
      I > N0.03
      I > O0.014
      N < O0.75 (84%)
    At 3.0 to 5.0 million cycles
      C > H 0.06 (74%)
      C > A 0.004
      C > I 0.002
      C > N0.007
      C > O 0.002
      H > A 0.04
      H > I0.02
      H > N0.02
      H > O0.009
      A > I 0.05
      A > N 0.09
      A > O 0.01
      I > N 0.22
      I > O0.01
      N > O 0.16 (61%)
     
    Anchor for JumpAnchor for JumpTABLE IV:  Wear Rates of Acetabular Cups After Artificial Aging at 80 Degrees Celsius in Air for Fourteen Days
    *The values are given in cubic millimeters per million cycles.
    Type of CupCup No.Between 0.5 and 3.5 Million Cycles*Between 3.5 and 5.0 Million Cycles*
    Wear Rate  Mean and Standard DeviationWear RateMean and Standard Deviation
    Nonsterilized controlsC3C426.124.325.2 ± 1.229.625.927.8 ± 2.6
    Hylamer, gas plasmaH3H427.529.528.5 ± 1.431.632.632.1 ± 0.7
    Irradiated in airA3A481.857.9  69.9 ± 16.982.342.362.3 ± 28.3
    Ion-implanted (IONGUARD III) and irradiated in airI3I428.229.829.0 ± 1.160.052.956.5 ± 5.0
    Irradiated in nitrogen and thermally stabilized (Duration)N3N413.413.213.3 ± 0.219.215.117.2 ± 2.9
    Irradiated with oxygen scavenger (Oxygenless Packaging)O3O414.314.014.2 ± 0.213.614.113.9 ± 0.4
     
    Anchor for JumpAnchor for JumpTABLE V:  Comparisons of the Wear Rates of the Various Materials After Fourteen Days of Artificial Aging
    *C = nonsterilized controls; H = Hylamer, gas plasma; A = irradiated in air; I = ion-implanted and irradiated in air; N = irradiated in nitrogen and thermally stabilized; and O = irradiated with oxygen scavenger.†Power is shown in parentheses.
    Compared Groups*P Value for Difference Between Groups†
    At 0.5 to 3.5 million cycles
      C < H0.13 (88%)
      C < A0.07
      C < I0.08
      C > N0.005
      C > O0.006
      H < A0.08
      H < I0.75
      H > N0.004
      H > O0.005
      A > I0.08
      A > N0.04
      A > O0.04
      I > N0.002
      I > O0.003
      N < O0.05 (99.9%)
    At 3.5 to 5.0 million cycles
      C < H0.15 (77%)
      C < A0.23
      C < I0.02
      C > N 0.06
      C > O 0.02
      H < A 0.27
      H < I0.02
      H > N 0.02
      H > O 0.001
      A > I 0.8
      A > N 0.15
      A > O 0.14
      I > N 0.01
      I > O 0.007
      N > O 0.25 (46%)
     
    Anchor for JumpAnchor for JumpTABLE VI:  Wear Rates of Acetabular Cups After Artificial Aging at 80 Degrees Celsius in Air for Thirty Days
    *The values are given in cubic millimeters per million cycles.
    Type of Cup  Cup No.Between 0.5 and 3.0 Million Cycles*Between 3.0 and 5.0 Million Cycles*Between 5.0 and 7.5 Million Cycles*
      Wear RateMean and Standard Deviation  WearRateMean and Standard Deviation  Wear RateMean and Standard Deviation
    Nonsterilized controlsC5C633.333.033.2 ± 0.2  34.9 38.3  36.6 ± 2.433.936.935.4 ± 2.1
    Hylamer, gas plasmaH5H645.732.539.1 ± 9.4  41.8 37.3  39.6 ± 3.236.936.236.6 ± 0.5
    Irradiated in airA5A682.165.373.7 ± 11.8  82.7 68.4  75.6 ± 10.161.072.866.9 ± 8.3
    Ion-implanted (IONGUARD III) and irradiated in air I5I646.351.448.9 ± 3.6105.5112.2108.9 ± 4.783.968.876.4 ± 10.7
    Irradiated in nitrogen and thermally stabilized (Duration)N5N661.843.652.7 ± 12.9  69.0 58.7  63.9 ± 7.348.951.650.3 ± 1.9
    Irradiated with oxygen scavenger (Oxygenless Packaging)O5O649.320.835.1 ± 20.2  90.9 31.7  61.3 ± 41.990.071.280.6 ± 13.3
     
    Anchor for JumpAnchor for JumpTABLE VII:  Comparisons of Wear Rates of the Various Materials After Thirty Days of Artificial Aging
    *C = nonsterilized controls; H = Hylamer, gas plasma; A = irradiated in air; I = ion-implanted and irradiated in air; N = irradiated in nitrogen and thermally stabilized; and O = irradiated with oxygen scavenger.†Power is shown in parentheses.
      Compared Groups*P Value for Difference Between Groups†
    At 0.5 to 3.0 million cycles
      C < H0.46 (40%)
      C < A 0.04
      C < I0.03
      C < N 0.17
      C < O 0.9
      H < A 0.08
      H < I0.30
      H < N 0.35
      H > O 0.82
      A > I0.10
      A > N 0.23
      A > O 0.14
      I < N 0.72
      I > O0.44
      N > O 0.41
    At 3.0 to 5.0 million cycles
      C < H 0.41 (91%)
      C < A 0.03
      C < I0.003
      C < N 0.04
      C < O 0.49
      H < A 0.04
      H < I0.003
      H < N0.05
      H < O 0.54
      A < I 0.05
      A > N 0.32
      A > O 0.69
      I > N 0.02
      I > O 0.25
      N > O0.94
    At 5.0 to 7.5 million cycles
      C < H 0.53 (99.9%)
      C < A 0.04
      C < I0.03
      C < N 0.02
      C < O 0.04
      H < A 0.04
      H < I 0.03
      H < N 0.01
      H < O 0.04
      A < I0.43
      A > N 0.11
      A < O 0.34
      I > N 0.08
      I < O 0.76
      N < O 0.09
     
    Anchor for JumpAnchor for JumpTABLE VIII:  Results of Wear Debris Analysis
    *The values are given as the mean and the standard deviation, with the composition of the debris in parentheses.
    Type of CupSize and Composition of Debris* (m)
    Without Aging14 Days of Aging30 Days of Aging
    GranulesFibrilsFlakesGranulesFibrilsFlakesGranulesFibrilsFlakes
    Nonsterilized controls0.21 ± 0.04 (66%)1.9 ± 1.1 (26%)0.34 ± 0.1 (7%)0.23 ± 0.05 (85%)2.1 ± 1.1 (14%)0.36 ± 0.03 (1%)0.22 ± 0.04 (81%)2.4 ± 1.2 (6%)0.34 ± 0.09 (14%)
    Hylamer, gas plasma 0.23 ± 0.06 (60%)1.9 ± 1.1 (30%)0.41 ± 0.1 (9%)0.23 ± 0.04 (74%)2.3 ± 1.2 (23%)0.38 ± 0.13 (2%)0.23 ± 0.04 (69%)2.0 ± 1.1 (15%)0.39 ± 0.10 (14%)
    Irradiated in air0.20 ± 0.04 (83%)1.5 ± 0.8 (14%)0.36 ± 0.1 (3%)0.22 ± 0.05 (90%)1.9 ± 1.1 (7%)0.29 ± 0.12 (5%)0.21 ± 0.04 (79%)1.7 ± 0.8 (11%)0.35 ± 0.08 (9%)
    Ion-implanted (IONGUARD III) and irradiated in air0.21 ± 0.04 (81%)1.5 ± 1.0 (14%)0.41 ± 0.1 (4%)0.23 ± 0.07 (77%)2.1 ± 1.2 (18%)0.44 ± 0.09 (5%)0.23 ± 0.04 (73%)1.9 ± 0.9 (10%)0.38 ± 0.11 (18%)
    Irradiated in nitrogen and thermally stabilized (Duration)0.21 ± 0.07 (61%)2.0 ± 1.5 (30%)0.56 ± 0.6 (9%)0.23 ± 0.06 (77%)1.8 ± 0.8 (18%)0.43 ± 0.09 (5%)0.19 ± 0.03 (83%)1.6 ± 0.6 (5%)0.29 ± 0.06 (11%)
    Irradiated with oxygen scavenger (Oxygenless Packaging)0.20 ± 0.04 (75%)1.6 ± 1.0 (17%)0.44 ± 0.2 (8%)0.24 ± 0.06 (76%)2.0 ± 0.8 (20%)0.39 ± 0.11 (4%)0.19 ± 0.04 (90%)1.6 ± 0.8 (4%)0.34 ± 0.12 (7%)

    Types of Polyethylene Tested

    In order to eliminate variation in the wear rates due to batch-to-batch differences in the original polymer, all of the acetabular cups (Table I) were prepared from a single batch of GUR 4150 ultra-high molecular weight polyethylene (Hoechst Celanese, Houston, Texas) that had been extruded into three-inch (7.6-centimeter) diameter bars in a single run (Poly Hi Solidur, Fort Wayne, Indiana) and had been made available to numerous laboratories as a reference material (S. Li, The Hospital for Special Surgery, New York, N.Y.). Sections of bar stock were forwarded to the five participating manufacturers, who prepared them according to their respective processes (Table I, Fig. 1). The cups were machined to a specified generic shape (a 28.2-millimeter inside diameter and a ten-millimeter wall thickness) suitable for use on the hip simulator. These cups were tested with balls that had a nominal diameter of twenty-eight millimeters. The inside diameter of the cups, which were measured with a dial caliper after all processing steps were complete, ranged from 28.11 to 28.21 millimeters, so the ball-cup clearances averaged 0.2 0.01 millimeter. The materials (Table I) included nonsterilized ultra-high molecular weight polyethylene cups (the controls), Hylamer cups (DePuy-DuPont Orthopaedics, Wilmington, Delaware) sterilized with gas plasma, conventional polyethylene cups sterilized with gamma irradiation in air (as has been done historically), cups implanted with nitrogen ions prior to irradiation in air, and cups irradiated while in one of two types of low-oxygen packages. The nitrogen ions (IONGUARD III) were implanted at Spire Corporation (Bedford, Massachusetts) with use of an ion implanter (model 10-80; Varian, Palo Alto, California) at 160 kiloelectron volts, to a total dose of about 3 l015 ions per square centimeter. With use of theoretical modeling software TRIM (Transport of Ions in Matter)73, the depth of penetration was calculated as about 0.5 micrometer.
    Since the amount of cross-linking induced in ultra-high molecular weight polyethylene by gamma irradiation markedly affects its wear resistance, the cups that were sterilized with gamma irradiation were placed together in a single container (SteriGenics, Charlotte, North Carolina) to ensure equal and simultaneous doses. Five dosimeters (Howmedica, Rutherford, New Jersey) that were spaced evenly in the container were subsequently analyzed, and they indicated a mean dose of 2.72 0.04 megarad (27.2 0.4 kilogray). After irradiation, the nitrogen-packaged cups (Table I) were returned to the manufacturer (Howmedica) for thermal stabilization to reduce the residual free radicals. This process included heating the cups, while they were still packaged in nitrogen, to 50 degrees Celsius for about six days. The cups were then returned to our laboratory for inclusion in the wear study.

    Artificial Aging of the Cups

    Three wear tests were carried out: the first was performed without artificial aging, the second was run after thirty days of thermal aging, and the third was conducted after fourteen days of aging. Although all of the cups in a given test were treated identically, because of the availability of the wear simulator the storage times varied among the three tests. For the first test (without thermal aging), the cups that were irradiated in the two types of low-oxygen packaging (Howmedica and Intermedics Orthopedics, Austin, Texas) were removed from their packets about one month after the irradiation and stored in air along with the remaining cups for about three months prior to the start of the wear test. For the second test the cups were stored under the same conditions for about six months prior to the thirty days of thermal aging, and for the third test the cups were similarly stored for about twelve months prior to the fourteen days of thermal aging. For aging, the cups were placed in an oven (inside dimensions, thirty-three by thirty-three by forty-eight centimeters) in ambient air and the temperature was gradually increased by approximately 0.2 degree Celsius per minute to 80 degrees Celsius, maintained for fourteen or thirty days, and then slow-cooled (that is, the heating unit was shut off with the cups still in the oven) to room temperature61. There was no forced circulation of the air in the oven during aging.
    Oxygen that diffuses into the surface layer of polyethylene cups during storage in air tends to increase the amount of oxidation during the subsequent thermal aging59. Thus, the prestorage of the cups in air served to reduce the duration of the subsequent oven heating that was required to artificially age the cups. In contrast, most manufacturers currently recommend storage of the cups in their low-oxygen packages, and they recommend against long-term storage on a shelf prior to implantation. Thus, the extended period of storage in air prior to thermal aging used in the present study should be considered an inherent part of the artificial aging process rather than a model of current clinical practice.
    The first wear test, performed without artificial aging, was intended to compare the wear characteristics of the various materials that would likely occur if they were implanted shortly after fabrication and sterilization. For the second wear test, the thirty days of thermal aging were sufficient to oxidize the cups that had been irradiated in air to a degree comparable with the greatest degree of oxidation that we have measured in any clinically retrieved polyethylene cups that had been irradiated in air. For example, Figure 2 shows the oxidation profiles of polyethylene cups after extended periods of shelf-aging and/or in vivo implantation. Since the exact composition, processing, and storage history of these cups were unknown, their oxidation profiles are presented only as examples of the highest levels of oxidation that have been observed in cups irradiated in air. The question addressed was: if nonirradiated cups or cups irradiated in low-oxygen packaging were to be subjected to a combination of shelf-aging and in vivo aging that, in the past, has led to high levels of oxidation in cups irradiated in air (Fig. 2), how much lower would the oxidation levels in the cups irradiated in low-oxygen packaging be and how would this amount of oxidation affect their wear resistance?
    Since the thirty days of artificial aging used in the second wear test markedly affected the oxidation levels and the wear resistance of the irradiated materials, and this amount of oxidation represented the most severe that we had observed on retrieved implants, the third set of cups was thermally aged for only fourteen days to evaluate the effects of an intermediate amount of oxidative aging.

    Method of Wear-Testing

    In each of the three wear tests, two cups made of each material, bearing against forged cobalt-chromium-molybdenum balls (Howmedica), were tested on a twelve-station hip-joint simulator (Shore Western Manufacturing, Monrovia, California) under a double-peaked, Paul-type physiological hip load45, with a 2000-newton maximum, in bovine blood-serum lubricant (HyClone, Logan, Utah). EDTA was added to a concentration of twenty millimoles to minimize precipitation of calcium phosphate onto the bearing surfaces30, and sodium azide was added to 0.2 weight percent to retard bacterial degradation. The cups were mounted below the balls and were oscillated through a 46-degree biaxial arc at one cycle per second32,33.
    To determine wear, the cups were removed from the simulator, cleaned ultrasonically, vacuum-dried, and then weighed on a microbalance (Mettler AT 261; Mettler-Toledo AG, Greifensee, Switzerland). The absorption of fluid by a worn cup was corrected for by increasing its measured weight loss by the mean weight gain of two cyclically loaded but nonoscillated soak-control cups of the same material. The corrected weight loss was converted to volumetric wear by dividing it by the original density of each cup (that is, 0.927, 0.935, and 0.960 gram per cubic centimeter for the nonsterilized cups, the cups sterilized with gamma irradiation, and the Hylamer polyethylene cups, respectively). While the use of a single value for the density ignored any change in density caused by subsequent oxidation of the polyethylene or other effects, the resultant error in the calculated volumetric wear rates was estimated to be less than 3 percent. The wear rates (Table II, Table III, Table IV, Table V, Table VI, and Table VII) were calculated by linear regression applied to the corrected weight-loss data. Linear regression was applied to separate subintervals during which, from inspection of the wear graphs (Fig. 4-A, Fig. 4-B, Fig. 6-A, Fig. 6-B, Fig. 8-A, and Fig. 8-B), it was apparent that one or more of the materials had exhibited a wear rate substantially different from that in the other intervals in the same test. These intervals are indicated in the tables listing the wear rates.
    The p values indicating the levels of significance of the differences between the mean wear rates were calculated with use of a double-sided t test. When the wear behavior of two materials was similar, the power to detect a difference of at least 50 percent (delta) of the mean wear rate at a significance (p value) of 0.05 (alpha) or less was also calculated. This choice for delta was based on the assumption that, in clinical use, a difference of at least 50 percent between the mean wear rates of two types of polyethylene would be necessary for the prevalence of osteolysis to be substantially different37.
    The wear of the cups subjected to fourteen days of artificial aging also was assessed by comparing the dimensions of the cups before and after the wear test, with use of a coordinate-measuring machine (BRT 504; Mitutoyo, Aurora, Illinois) fitted with a ruby-tipped touch probe (TP-200; Renishaw, Hoffman Estates, Illinois). (Because the coordinate-measuring machine was not obtained until after the first two wear tests were performed, wear could not be determined by the coordinate-measuring machine for the cups from the other two tests.) Calibration tests indicated that the accuracy was within approximately two micrometers. The coordinate-measuring-machine readings were made over a grid of 301 points evenly distributed over the convex interior of the cup. These data were imported into a computer, and a commercial program (Qualstar 2.0H; ICAMP, Los Alamos, New Mexico) was used to fit a spherical surface to the data with use of a least-squares method. The cups were measured again after the wear test, and the displacement of each measured point from the original reference sphere was calculated, providing a map of the total penetration over the inner surface of the worn cup. When the maximum depth for each was plotted against its corresponding wear volume (that is, as determined from the corrected weight measurement), an approximate linear fit was obtained, with a slope of about 1.5 micrometers per cubic millimeter. On the basis of the assumption that a comparable relationship held for the other two wear tests, this scale factor was used to indicate the approximate corresponding depth of penetration on the plots of wear volume versus cycles (with an estimated accuracy of within approximately twenty-five micrometers).

    Measurement of the Levels of Oxidation

    Oxidation was measured in each wear-tested cup (that is, two cups for each material and aging condition) with use of transmission Fourier transform infrared spectroscopy. Segments were cut from the cups, embedded in polymethylmethacrylate, and microtomed into 200-micrometer-thick sections. The sections were soaked in hexane for sixteen hours prior to Fourier transform infrared analysis since, in preliminary studies, hexane extraction had reduced the contaminants absorbed from the serum lubricant that can cause erroneously high oxidation readings25,58,62. Measurements were made with use of a Fourier transform infrared spectroscopy system (Polaris IR 10410; Mattson Instruments, Madison, Wisconsin) with an attached microscope (Spectra-Tech, Shelton, Connecticut) with use of a 0.1 by 0.1-millimeter window, starting at the inner (bearing) surface of the cup and moving across the thickness in 100-micrometer steps, with use of sixty-four scan summations at a resolution of sixteen centimeters-1 with a mercury cadmium telluride detector. The oxidation ratio58,72 was calculated as the height of the carbonyl absorption peak at 1717 centimeters-1 divided by the height of the reference peak at 2022 centimeters-1. Because the wear process removed much of the oxidized surface layer in the contact zone, the oxidation profiles in the present study were made across the entire thickness in nonworn regions of the cups to provide the complete profiles. The interactive effects of oxidation and cross-linking on the wear resistance for a particular cup then could be assessed by comparing the oxidation ratio at any particular depth to the slope of the wear curve at the same depth. In most cases, the oxidation profiles were similar for the two cups in a pair, and the mean oxidation ratios of the two cups were plotted. For the exceptions (that is, the ion-implanted cups and the cups irradiated with an oxygen scavenger and aged for thirty days), the profiles were plotted for both cups in the pair.

    Characterization of Wear Debris

    At the end of each wear test, the polyethylene wear debris was isolated from the serum with use of a density-gradient technique5. The particles were examined with a scanning electron microscope (DSM 960; Carl Zeiss, Oberkochen, Germany), and micrographs of representative particles were transferred to a digital image-analysis system (Image One, West Chester, Pennsylvania). For each cup tested, 150 to 400 randomly selected particles were measured and categorized as smooth, round, submicrometer-sized granules; as elongated fibrils; or as irregularly shaped flakes. The level of significance of the differences in the particle morphologies, as indicated by the associated p value, was calculated with use of a double-sided t test.

    Oxidation and Wear without Artificial Aging

    There was little or no oxidation in the nonsterilized control cups or in the gas-plasma-sterilized Hylamer cups (DePuy-DuPont Orthopaedics) (Fig. 3), and the two materials wore at nearly constant and comparable rates throughout five million cycles (Fig. 4-A and Fig. 4-B, Table II and Table III). In contrast, all of the irradiated cups showed some oxidation, with the differences among the materials occurring primarily from the surface to about three millimeters deep. The conventional polyethylene cups irradiated in air showed maximum oxidation at the surface, with a secondary peak about 800 micrometers deep. The corresponding wear rate was initially slightly higher than that for the nonsterilized cups (Fig. 4-A and Fig. 4-B, Table II and Table III), but it decreased with increasing depth from the surface, becoming about 65 percent of that of the nonsterilized cups between three and five million cycles.
    Initially, the ion-implanted cups irradiated in air had a markedly yellow-brown coloration, but this was worn off the contact area of the bearing surface during the first 0.25 million cycles. These cups exhibited a peak of oxidation that was comparable with that of the conventional polyethylene cups irradiated in air at the bearing surface, but they showed less oxidation at a depth of greater than approximately 400 micrometers (Fig. 3). Initially, wear was less than that of the conventional polyethylene cups irradiated in air; it decreased further during the middle section of the test, and then it increased to a comparable rate during the final two million cycles (Fig. 4-A and Fig. 4-B, Table II and Table III).
    The two sets of cups that were irradiated in a low-oxygen environment had substantially less oxidation in the surface layer than did the cups irradiated in air (Fig. 3). The wear rates also were lower, and they were nearly constant for the duration of the five million cycles (Fig. 4-A and Fig. 4-B), reflecting the relatively constant levels of oxidation (and, therefore, cross-linking) across the thickness. Although the oxidation levels in the two types of cups irradiated in low-oxygen packaging were higher than those in the nonsterilized controls and in the gas-plasma-sterilized Hylamer cups, the wear rates were about 50 percent lower (Table II and Table III), presumably because of the radiation-induced cross-linking.

    Oxidation and Wear After Fourteen Days of Artificial Aging

    Aging of the nonsterilized cups and the gas-plasma-sterilized Hylamer cups (DePuy-DuPont Orthopaedics) for fourteen days at 80 degrees Celsius produced little or no oxidation through the thickness of the polyethylene (Fig. 5), and the wear rates were again nearly constant over the duration of the test (Fig. 6-A and Fig. 6-B, Table IV and Table V). In contrast, the cups irradiated in air exhibited a substantial oxidation peak at the surface, and there was correspondingly rapid wear during the first 1.5 million cycles. Although the wear rate decreased with increasing depth, the wear rate during the final 1.5 million cycles was still about two and one-quarter times greater than that of the aged nonsterilized controls (Table IV and Table V); this finding reflected the effect of elevated oxidation caused by the aging - that is, the degradation of the cross-linked network.
    After fourteen days of aging, the oxidation level at the surface of the ion-implanted cups was substantially lower than that of the conventional polyethylene cups irradiated in air, but it increased to a comparable level at a depth of about 0.4 millimeter (Fig. 5). Consistent with this, the wear rate was initially much lower than that of the cups irradiated in air, but it increased to a comparable level after 3.5 million cycles (Fig. 6-A and Fig. 6-B, Table IV and Table V).
    The two sets of cups that had been irradiated in a low-oxygen atmosphere again showed comparable oxidation profiles (Fig. 5) and wear behavior (Fig. 6-A and Fig. 6-B) throughout the five million cycles. Although the oxidation levels were greater than those of the same materials prior to artificial aging (Fig. 3 compared with Fig. 5), they were still substantially lower than those of the cups irradiated in air, and the wear rates were again lower than those of the nonirradiated (that is, the non-cross-linked) cups.

    Oxidation and Wear After Thirty Days of Artificial Aging

    Thirty days of aging at 80 degrees Celsius produced no measurable oxidation of the nonsterilized cups and only slight oxidation of the gas-plasma-sterilized Hylamer cups (Fig. 7). The corresponding wear rates were nearly constant and were comparable for the two materials throughout the 7.5 million cycles (Fig. 8-A and Fig. 8-B, Table VI and Table VII), again reflecting the lack of oxidation or cross-linking gradients within these cups. In contrast, there was severe oxidation of the surface layer of the cups irradiated in air (Fig. 7), which wore off during the initial 0.5 million cycles. Between three and five million cycles, the mean wear rate of the cups irradiated in air and subjected to thirty days of aging was about 21 percent greater than that of the cups with fourteen days of aging and 183 percent greater than that of the cups without aging (Fig. 8-A and Fig. 8-B, Table II, Table IV, and Table VI).
    The oxidation at the surface of the ion-implanted cups after thirty days of aging was again much lower than that of the cups irradiated in air without ion implantation, but both ion-implanted cups exhibited substantial subsurface oxidation peaks (Fig. 7). Consistent with these oxidation profiles, the wear rates of the ion-implanted cups were initially substantially lower than those of the cups irradiated in air - that is, they were comparable with those of the materials irradiated in a low-oxygen environment (Fig. 8-A) - but they increased during the test, becoming about 44 percent greater than those of the cups irradiated in air between three and five million cycles and 14 percent greater between five and 7.5 million cycles (Fig. 8-B, Table VI and Table VII).
    Thirty days of aging of the cups that were irradiated in nitrogen and thermally stabilized produced substantially greater surface and subsurface oxidation than was present in these materials after fourteen days of aging (Fig. 7 compared with Fig. 5). Nevertheless, the oxidation level within one millimeter of the surface was again much lower than that of the cups irradiated in air, demonstrating the effect of sterilization in the low-oxygen atmosphere. Throughout the 7.5 million cycles, the mean wear rate of the nitrogen-packaged and thermally stabilized cups was lower than that of the cups irradiated in air but was greater than that of the nonsterilized controls and the Hylamer gas-plasma-sterilized cups (Fig. 8-A and Fig. 8-B, Table VI and Table VII).
    In contrast, although both of the cups that were irradiated while packaged with an oxygen scavenger showed low levels of oxidation at the surface (Fig. 7), one of these cups (O5) showed a severe oxidation peak at a depth of about 0.5 millimeter (comparable with the maximum for one of the ion-implanted cups) and the wear rate was several times greater than that of the nonirradiated cups (Fig. 8-A and Fig. 8-B, Table VI and Table VII). The other oxygen-scavenger cup (O6) exhibited much less oxidation (Fig. 7) and underwent the least wear of any cup for the first six million cycles of the test (Fig. 8-A and Fig. 8-B). Of the eighteen pairs of cups in the three wear tests, only this pair exhibited such marked differences between the wear rates of the two cups. Although it was not determined why the oxidation level was much higher in cup O5 than it was in cup O6, additional cups that were subsequently prepared by the manufacturer with use of the same process and then were irradiated and aged exhibited oxidation levels comparable with those of the less oxidized wear-test cup (O6), suggesting the possibility of a packaging defect with cup O5.
    The wear debris recovered from the serum lubricant consisted primarily of granules (60 to 90 percent of the particles were granules), with fibrils and flakes making up the remainder (Table VIII), as is typically found in periarticular tissues from hips with ultra-high molecular weight polyethylene cups irradiated in air36. In addition, the distribution of the particle sizes was comparable among the different types of polyethylene tested. However, for most of the materials, the wear debris from the cups that had been artificially aged for thirty days contained a lower percentage of fibrils (p = 0.006) and a higher percentage of granules (p = 0.08) and flakes (p = 0.02) than did those without aging or those after fourteen days of aging (Table VIII).
    With the exceptions noted above, the oxidation profiles and wear patterns were very similar for two cups of a particular material and for different materials that had comparable radiation-cross-linking histories. This repeatability was due in part to the minimum variation in material properties obtained by the use of a single parent batch of ultra-high molecular weight polyethylene and the fact that all of the cups that were sterilized with radiation were exposed together, in the same container, to a single dose. When large systematic differences were found between the mean wear rates of two materials that had been prepared under substantially different conditions (for example, nonsterilized cups compared with those irradiated in a low-oxygen atmosphere [Fig. 4-A and Fig. 4-B]), testing of only two samples of each material was sufficient to provide high levels of statistical reliability, indicating that these differences did not occur by chance selection (Table III, Table V, and Table VII).
    In contrast, the power to detect substantial differences between pairs of materials that exhibited similar wear rates within a particular test interval was sometimes low. For example, the power to detect a 50 percent difference between the mean wear rates, during the interval from 3.5 to 5.0 million cycles, of the two sets of cups irradiated in a low-oxygen atmosphere and then subjected to fourteen days of artificial aging (materials N and O) (Fig. 6-A and Fig. 6-B) was 46 percent. However, it should be emphasized that this power calculation applies only to the likelihood of detecting a difference between the wear rates within that specific interval, and it does not take into account the fact that these two materials also showed comparable behavior throughout that entire wear test and throughout the test run without aging (Fig. 4-A and Fig. 4-B). This consistent similarity of the wear behavior increases the reliability of a conclusion that the wear rates are not different in general - that is, the conclusion has greater reliability than that indicated by the power calculation for that interval alone. In addition, the similarity of the wear behavior was consistent with the fact that the cups of the two materials had similar distributions of oxidation and, presumably, cross-linking, at least until they were extensively oxidized (Fig. 8-A and Fig. 8-B). This reasoning applies to the other pairs of materials that showed comparable wear rates (for example, the two types of polyethylene that were not irradiated or cross-linked [materials C and H]) (Fig. 4-A, Fig. 4-B, Fig. 6-A, Fig. 6-B, Fig. 8-A, and Fig. 8-B).
    In general, when the cups were heated in air61 the maximum oxidation was near the surface, whereas the peak is typically below the surface in shelf-aged and/or in vivo aged cups (Fig. 2) and in cups aged under pressurized oxygen15. While this difference must be taken into account when one predicts the long-term clinical wear behavior of a given material, placement of the oxidized zone nearer the surface of the cup had the practical advantage of shortening the duration of the wear-testing required for the wear to penetrate into the most oxidized (and, therefore, least cross-linked) layer. Thus, the differences among the wear rates of the artificially aged materials in the present study probably represent the maximums that might occur over decades of clinical use.
    The shift in the composition of the wear debris away from drawn-out fibrils and toward granules and flakes for each of the materials after thirty days of thermal aging probably reflected the embrittlement of the polyethylene due to oxidative degradation, as has been measured in tensile tests of sections cut from retrieved cups63. Nevertheless, the similar morphology and size distributions of the wear debris from the six types of polyethylene (Table VIII) suggested that the osteolytic reactions to comparable amounts of wear debris would also be similar.
    For the cups sterilized with gamma irradiation in air, the higher wear rates after artificial aging were consistent with the marked oxidation (and, therefore, reduced cross-linking) of their surface layers. Because of this effect, most manufacturers no longer sterilize polyethylene components while packaged in air. In contrast, the nonsterilized and gas-plasma-sterilized cups had no free radicals to induce oxidation, and these cups showed comparable wear rates that were relatively immune to artificial aging. Although we did not include cups that had been sterilized with ethylene oxide in the present study, other tests performed in this laboratory41 and elsewhere69 have shown that the wear resistance of such cups is closely comparable with that of nonsterilized controls, as used in the present study. Clinically, the present results indicate that one can expect the wear rates of cups sterilized without irradiation to be lower than that of cups sterilized with gamma irradiation in air, particularly if the latter cups are stored for a substantial period prior to implantation.
    The cups prepared by the Hylamer process (DePuy-DuPont Orthopaedics) followed by gas-plasma sterilization exhibited little oxidation, and they had wear rates and particle morphologies comparable with those of the nonsterilized conventional polyethylene, even after extensive artificial aging. Thus, the two materials that were neither cross-linked nor oxidized exhibited nearly identical wear properties. Similarly, in a previous hip-simulator study in our laboratory34,35, the wear resistance of Hylamer cups that had been sterilized with gamma irradiation in air (but without artificial aging) was very comparable with that of cups made of conventional ultra-high molecular weight polyethylene, produced from the same parent batch of material, and sterilized with gamma irradiation in air. In contrast, with use of a hip simulator comparable with that used in the present study, Wang et al.69 found that the wear of Hylamercups sterilized with gamma irradiation in air was about 2.3 times greater than that of cups made of conventional polyethylene and sterilized in the same manner. However, in their study, the cups were not all fabricated from the same parent batch of polyethylene, they were irradiated separately, and they were shelf-stored for different intervals prior to the wear-testing; each of these factors might have affected the relative levels of oxidation and/or cross-linking and the resultant wear rates.
    Clinical studies also have shown contradictory results with Hylamer cups sterilized with gamma irradiation in air. For example, in a five-year follow-up study, Sychterz et al.65 recently reported that the mean wear rate of Hylamer cups sterilized with gamma irradiation in air was about 25 percent less than that of conventional cups during the first three years and about 15 percent greater during the fifth year. In contrast, other clinical series have shown that the wear rates, and the resultant prevalence and severity of osteolysis, were apparently greater in patients with Hylamercups sterilized with gamma irradiation in air6,31.
    Unfortunately, it is difficult to isolate quantitatively the relative influences of variables, other than the type of polymer, on the relative rate of wear in such clinical studies. Patient age and activity, the mixing of components from different manufacturers, and the likelihood of contamination by third-body abrasive particles are some of the variables that may influence wear23,36,55. Nevertheless, the results of recent laboratory analyses of retrieved Hylamer implants have suggested that the presence of radiation-induced free radicals, combined with differences in the duration of storage of the Hylamer cups sterilized with gamma irradiation in air prior to wear-testing or clinical implantation, could have been a key factor leading to the differences in the results among the above studies. For example, in an earlier study, we found that the level of oxidation in Hylamer cups sterilized with gamma irradiation in air was comparable with that in conventional ultra-high molecular weight polyethylene cups after 6.5 years of shelf storage, regardless of whether this storage was in air or in water38. However, in tensile tests of sections cut from retrieved implants, even at comparable levels of oxidation, Hylamer sterilized with gamma irradiation in air exhibited substantially greater reductions in ultimate tensile strength and elongation than did conventional polyethylene9. If there is a corresponding time-dependent decrease in the wear resistance of Hylamer sterilized with gamma irradiation in air, this could explain the elevated wear rates exhibited by Hylamer both in the laboratory study by Wang et al.69 and in some of the clinical series6,31.
    The fact that the yellow-brown discoloration of the ion-implanted cups in our study was worn off early in the wear test was consistent with the very shallow depth of penetration of the nitrogen ions (that is, 0.5 micrometer). For the balance of the wear test, the variation of wear with depth in the ion-implanted cups was consistent with their oxidation-cross-linking profiles. Although the nitrogen ions penetrated to a depth of only about 0.5 micrometer, the differences in the oxidation levels and the wear rates suggested that ion-implanting somehow served to retard oxidation28 at a substantially greater depth in the polymer during subsequent thermal aging (Fig. 5), such that there was less reduction of the wear resistance. However, the mechanism by which the superficial implanting of ions affected the oxidation properties deeper in the polyethylene was unexplained.
    Similarly, the cause of the very high subsurface oxidation that occurred after thirty days of aging in one of the ion-implanted cups, but not in the other (Fig. 7), was unexplained. Nevertheless, the variation in wear with depth for these cups relative to conventional cups sterilized with gamma irradiation in air (Fig. 8-A) again reflected the principle that higher oxidation and/or reduced cross-linking resulted in a higher rate of wear.
    Packaging the cups in a low-oxygen atmosphere resulted in a lower concentration of oxygen in the surface layer during irradiation, as reflected by the lower oxidation levels in the surface layers (Fig. 3). In a separate study41, we found that this also results in greater cross-linking of the surface layers, as was reflected in the lower wear rates relative to those of the cups irradiated in air and those of the nonirradiated cups. Nevertheless, the increase in oxidation levels of the cups irradiated in a low-oxygen atmosphere and subjected to thirty days of artificial aging (Fig. 7) indicated that a substantial number of free radicals remained in these cups, even in those that were thermally stabilized after irradiation. The elevated wear rates exhibited on wear-testing after thirty days of artificial aging by the cups irradiated in a low-oxygen atmosphere might have been a direct effect of the oxidative degradation of the material, which may have been due to a lowering of the molecular weight by breaking of the intramolecular carbon-carbon bonds and/or to an indirect effect of a reduction in cross-linking caused by the oxidation14,41. Regardless of the precise mechanism, the results of the present hip-simulator wear-testing study suggested that, if a comparable amount of oxidation were to occur clinically in cups that had been irradiated in a low-oxygen atmosphere, then the clinical wear rates might eventually exceed those of nonirradiated cups.
    Thus, the reliability of using the results of the present hip-simulator study to predict the relative clinical performance of the materials depends on how closely the oxidation levels and other changes produced by artificial aging compare with those occurring during shelf storage and subsequent use in vivo and on how many days of artificial aging are the equivalent of a given number of years of real-time aging. While it is possible to find examples of cups sterilized with irradiation in air, shelf-aged, and clinically retrieved with oxidation levels as high as those generated in the cups irradiated in air and artificially aged for thirty days in the present study (Fig. 2 compared with Fig. 7), much of the oxidation seen in the clinically retrieved cups might have occurred during shelf storage prior to implantation rather than during in vivo use16. Furthermore, since cups sterilized with gamma irradiation are now almost exclusively irradiated and stored in some type of low-oxygen packaging, and most manufacturers recommend against prolonged shelf storage, such cups are unlikely to reach oxidation levels as high as those generated by the thirty days of artificial aging in the present study.
    The range of linear penetration rates exhibited by the acetabular cups in the present study fell within the range of linear penetration rates measured radiographically for cups of comparable materials in vivo. One million cycles on the hip simulator is approximately the equivalent of one year of average clinical use of a hip prosthesis54,67. On the basis of the coordinate-measuring-machine measurements, the equivalent clinical linear penetration rates for the ultra-high molecular weight polyethylene cups sterilized with gamma irradiation in air and tested in the simulator were about 0.04 millimeter per year (between three and five million cycles [Table II]) without artificial aging, about 0.09 millimeter per year (between 3.5 and five million cycles [Table IV]) after fourteen days of aging, and about 0.10 millimeter per year (between five and 7.5 million cycles [Table VI]) after thirty days of aging. The highest wear rate measured after testing in the simulator, which occurred during the first one million cycles for the cups sterilized with gamma irradiation in air and then subjected to thirty days of aging (Fig. 8-A and Fig. 8-B), was the equivalent of a linear penetration rate of about 0.27 millimeter per year clinically. The latter linear penetration rate is somewhat above the range of 0.1 to 0.2 millimeter per year that has been reported for the mean clinical linear penetration rate of polyethylene sterilized with gamma irradiation in air10,56,64, but it is well below the maximum clinical linear penetration rates, which can be as high as 0.5 millimeter per year. This finding is consistent with the fact that, in clinical use, factors in addition to oxidative degradation, such as poor initial quality of the polyethylene, a patient with an unusually high body weight and/or activity level, and the entrapment of third-body abrasive particles3,11,23,36,46,55, can combine to accelerate the wear rate, and these latter factors were not included in the simulator model of the present study.
    The equivalent clinical wear rates of the other materials in the present study also were somewhat below the mean values reported for polyethylene irradiated in air. For example, after thirty days of aging, the mean wear rates between 5.0 and 7.5 million cycles (Table VI) were the equivalent of about 0.053 millimeter per year for the nonsterilized cups of conventional polyethylene, 0.054 millimeter per year for the gas-plasma-sterilized Hylamer cups, 0.11 millimeter per year for the ion-implanted polyethylene cups irradiated in air, 0.075 millimeter per year for the polyethylene cups irradiated in nitrogen and thermally stabilized, and 0.12 millimeter per year for the polyethylene cups irradiated with an oxygen scavenger.
    Another factor that might have contributed to the somewhat lower wear rates of cups tested in the simulator compared with wear rates associated with in vivo use was the use of 90 percent bovine serum as a lubricant. The protein concentration of 90 percent serum is at the high end of the range that has been measured in fluids obtained from hips with artificial joints43, and it has been shown that the wear of a polyethylene acetabular cup in a hip simulator decreases as the concentration of proteins in the test lubricant increases above this range19,69. The optimum concentration of serum proteins necessary for the simulator wear rate to best approximate that of an acetabular cup in vivo has not yet been established30,52,70. Fortunately, the type of wear generated in 90 percent serum is closely comparable with that of polyethylene cups in vivo36,68 so the relative wear rates of different types of polyethylene, as measured in the present study, should also be comparable with those in vivo.
    If testing of the cups in the hip simulator in the present study is a reasonably accurate model of in vivo conditions, then the wear rate of an ultra-high molecular weight polyethylene acetabular cup that has been sterilized with gamma radiation while packaged in a suitable low-oxygen environment can be expected to be substantially lower than that of a nonirradiated cup during the first decade of clinical use and possibly longer; however, this advantage eventually might be substantially reduced or eliminated by oxidative degradation of residual free radicals. In view of the present uncertainty of long-term predictions based on accelerated-aging techniques, some surgeons might consider use of an acetabular cup that has been sterilized without radiation in their youngest patients because of the potential for long-term oxidative degradation of the cups irradiated in a low-oxygen environment to eventually offset the benefits of the cross-linking. On the other hand, if a cup without cross-linking fails during the first decade as a result of wear-induced osteolysis, there will be no long-term benefit from its greater resistance to oxidation.
    Another approach to minimizing both oxidation and wear is to use radiation to cross-link the polyethylene while it is still in the form of an extruded rod or molded block. The cross-linked material can then be heated above the melt temperature (about 150 degrees Celsius) to extinguish the residual free radicals before it is machined into the final components. In laboratory tests, acetabular cups made of polyethylene that had been cross-linked with radiation and then remelted have demonstrated resistance to oxidation comparable with that of nonirradiated polyethylene and, when cross-linking was sufficient, wear resistance that was superior to that of acetabular cups sterilized with the conventional gamma dose in low-oxygen packaging, as used in the present study27,39,40,42.
    Finally, it should be emphasized that the specific results of the present study might apply only to the type of wear generated in a hip prosthesis. Since oxidation also tends to degrade the fatigue resistance of ultra-high molecular weight polyethylene, the modifications investigated in the present study could have very different relative effects on the wear resistance of some designs of total knee replacements, in which the cyclic contact stresses are substantially greater.
    Amstutz, H. C.; Campbell, P.; Kossovsky, N.; and Clarke, I. C.: Mechanisms and clinical significance of wear debris-induced osteolysis. Clin. Orthop.,276: 7-18, 1992.2767  1992  [PubMed]
     
    Bapst, J. M.; Valentine, R. H.; and Vasquez, R.: Wear simulation testing of direct compression molded UHMWPE irradiated in oxygenless packaging. Trans. Soc. Biomater.,20: 72, 1997.2072  1997 
     
    Bauer, T. W.; Taylor, S. K.; Jiang, M.; and Medendorp, S. V.: An indirect comparison of third-body wear in retrieved hydroxyapatite-coated, porous, and cemented femoral components. Clin. Orthop.,298: 11-18, 1994.29811  1994  [PubMed]
     
    Bhateja, S. K.; Duerst, R. W.; Martens, J. A.; and Andrews, E. H.: Radiation-induced enhancement of crystallinity in polymers. J. Macromol. Sci.-Revs. Chem. Phys.,C35: 581-659, 1995.C35581  1995 
     
    Campbell, P.; Ma, S.; Yeom, B.; McKellop, H.; Schmalzried, T. P.; and Amstutz, H. C.: Isolation of predominantly submicron-sized UHMWPE wear particles from periprosthetic tissues. J. Biomed. Mater. Res.,29: 127-131, 1995.29127  1995  [PubMed]
     
    Chmell, M. J.; Poss, R.; Thomas, W. H.; and Sledge, C. B.: Early failure of Hylamer acetabular inserts due to eccentric wear. J. Arthroplasty,11: 351-353, 1996.11351  1996  [PubMed]
     
    Clarke, I. C.; Campbell, P.; and Kossovsky, N.: Debris-mediated osteolysis - A cascade phenomenon involving motion, wear, particulates, macrophage induction and bone lysis. In Particulate Debris from Medical Implants, pp. 7-26. Edited by K. St. John. ASTM Special Technical Publications 1144. Philadelphia, American Society for Testing and Materials, 1992 
     
    Collier, J. P.; Sperling, D. K.; Currier, J. H.; Sutula, L. C.; Saum, K. A.; and Mayor, M. B.: Impact of gamma sterilization on clinical performance of polyethylene in the knee. J. Arthroplasty,11: 377-389, 1996.11377  1996  [PubMed]
     
    Collier, J. P.; Bargmann, L. S.; Currier, B. H.; Mayor, M. B.; Currier, J. H.; and Bargmann, B. C.: An analysis of Hylamer and polyethylene bearings from retrieved acetabular components. Orthopedics,,21: 865-871, 1998.21865  1998 
     
    Devane, P. A.; Robinson, E. J.; Bourne, R. B.; Rorabeck, C. H.; Nayak, N. N.; and Horne, J. G.: Measurement of polyethylene wear in acetabular components inserted with and without cement. A randomized trial. J. Bone and Joint Surg.,79-A: 682-689, May 1997.79-A682  1997 
     
    Diwan, A., and Drummond, R.: Unusual case of third-body wear in total hip arthroplasty. J. Arthroplasty,12: 586-588, 1997.12586  1997  [PubMed]
     
    Dumbleton, J. H.; Shen, C.; and Miller, E. H.: Study of the wear of some materials in connection with total hip replacement. Wear, ,29: 163-171, 1974.29163  1974 
     
    Eyerer, P., and Ke, Y. C.: Property changes of UHMW polyethylene hip cup endoprostheses during implantation. J. Biomed. Mater. Res.,18: 1137-1151, 1984.181137  1984  [PubMed]
     
    Eyerer, P.; Kurth, M.; McKellop, H. A.; and Mittlmeier, T.: Characterization of UHMWPE hip cups run on joint simulators. J. Biomed. Mater. Res.,21: 275-291, 1987.21275  1987  [PubMed]
     
    Fisher, J.; Reeves, E. A.; Isaac, G. H.; Saum, K. A.; and Sanford, W. M.: Comparison of the wear of aged and non-aged UHMWPE sterilized by gamma irradiation and by gas plasma. J. Mater. Sci. Mater. Med.,8: 375-378, 1997.8375  1997  [PubMed]
     
    Furman, B. D.; Kasprzak, D.; and Li, S.: Differences in oxidation between shelf life aged and retrieved ultra high molecular weight polyethylene components. Trans. Orthop. Res. Soc.,21: 484, 1996.21484  1996 
     
    Gillis, A. M.; Furman, B. D.; and Li, S.: Influence of ultra high molecular weight polyethylene resin type and manufacturing method on real time oxidation. Trans. Orthop. Res. Soc.,23: 360, 1998.23360  1998 
     
    Goldring, S. R.; Jasty, M.; Roelke, M. S.; Rourke, C. M.; Bringhurst, F. R.; and Harris, W. H.: Formation of a synovial-like membrane at the bone-cement interface. Its role in bone resorption and implant loosening after total hip replacement. Arthrit. and Rheumat.,29: 836-842, 1986.29836  1986 
     
    Good, V.; Clarke, I. C.; Gustafson, A.; Braham, A.; Downs, B.; Sorensen, K.; and Anissian, L.: Dose response of protein concentration to wear in PTFE and UHMWPE. Trans. Orthop. Res. Soc.,24: 839, 1999.24839  1999 
     
    Goodman, S.: The role of polymer particles in the process of aseptic loosening of joint arthroplasties. In Surface Modification Technologies V, pp. 36-48. Edited by T. Sudarshon and J. Braza. London, Institute of Materials, 1992.  
     
    Greer, K. W.; Schmidt, M. B.;, and Hamilton, J. V.:: The hip simulator wear of gamma-vacuum, gamma-air, and ethylene oxide sterilized UHMWPE following a severe oxidative challenge. Trans. Orthop. Res. Soc., ,23: 52, 1998.2352  1998 
     
    Grobbelaar, C. J.; du Plessis, T. A.; and Marais, F.: The radiation improvement of polyethylene prostheses. A preliminary study. J. Bone and Joint Surg.,60-B(3): 370-374, 1978.60-B(3)370  1978 
     
    Hop, J. D.; Callaghan, J. J.; Olejniczak, J. P.; Pedersen, D. R.; Brown, T. D.; and Johnston, R. C.: Contribution of cable debris generation to accelerated polyethylene wear. Clin. Orthop.,344: 20-32, 1997.34420  1997  [PubMed]
     
    Howie, D. W.; Vernon-Roberts, B.; Oakeshott, R.; and Manthey, B.: A rat model of resorption of bone at the cement-bone interface in the presence of polyethylene wear particles. J. Bone and Joint Surg.,70-A: 257-263, Feb 1988.70-A257  1988 
     
    James, S. P.; Blazka, S.; Merrill, E. W.; Jasty, M.; Lee, K. R.; Bragdon, C. R.; and Harris, W. H.: Challenge to the concept that UHMWPE acetabular components oxidize in vivo. Biomaterials,14: 643-647, 1993.14643  1993  [PubMed]
     
    Kabo, J. M.; Gebhard, J. S.; Loren, G.; and Amstutz, H. C.: In vivo wear of polyethylene acetabular components. J. Bone and Joint Surg.,75-B(2): 254-258, 1993.75-B(2)254  1993 
     
    Kurtz, S. M.; Muratoglu, O. K.; Evans, M.; and Edidin, A. A.: Advances in the processing, sterilization, and crosslinking of ultra-high molecular weight polyethylene for total joint arthroplasty. Biomaterials,20: 1659-1688, 1999.201659  1999  [PubMed]
     
    Lee, E. H.; Rao, G. R.; Lewis, M. B.; and Mansur, L. K.: Ion beam application for improved polymer surface properties. Nucl. Instrum. Meth.,B74: 326-330, 1993.B74326  1993 
     
    Li, S.,, and Burstein, A. H.: Ultra-high molecular weight polyethylene. The material and its use in total joint implants. J. Bone and Joint Surg. ,76-A: 1080-1090, July 1994.76-A1080  1994 
     
    Liao, Y. S.; Benya, P. D.; and McKellop, H. A.: Effect of protein lubrication on the wear properties of materials for prosthetic joints. J. Biomed. Mater. Res.,48: 465-473, 1999.48465  1999  [PubMed]
     
    Livingston, B. J.; Chmell, M. J.; Spector, M.; and Poss, R.: Complications of total hip arthroplasty associated with the use of an acetabular component with a Hylamer liner. J. Bone and Joint Surg.,,79-A: 1529-1538, Oct 1997.79-A1529  1997 
     
    McKellop, H.; Clarke, I.; Markolf, K.; and Amstutz, H.: Friction and wear properties of polymer, metal, and ceramic prosthetic joint materials evaluated on a multichannel screening device. J. Biomed. Mater. Res.,15: 619-653, 1981.15619  1981  [PubMed]
     
    McKellop, H., and Clarke, I.: Evolution and evaluation of materials screening machines and joint simulators in predicting in vivo wear phenomena. In Functional Behavior of Orthopaedic Biomaterials, pp. 51-85. Edited by P. Ducheyne and G. Hastings. Boca Raton, CRC Press, 1984 
     
    McKellop, H. A.; Lu, B.; and Li, S.: Wear of acetabular cups of conventional and modified UHMW polyethylenes compared on a hip joint simulator. Trans. Orthop. Res. Soc.,17: 356, 1992.17356  1992 
     
    McKellop, H. A.: Wear modes, mechanisms, damage, and debris. Separating cause from effect in the wear of total hip replacements. In Total Hip Revision Surgery, pp. 21-39. Edited by J. O. Galante, A. G. Rosenberg, and J. J. Callaghan. New York, Raven Press, 1995.  
     
    McKellop, H. A.; Campbell, P.; Park, S.-H.; Schmalzried, T. P.; Grigoris, P.; Amstutz, H. C.; and Sarmiento, A.: The origin of submicron polyethylene wear debris in total hip arthroplasty. Clin. Orthop.,311: 3-20, 1995.3113  1995  [PubMed]
     
    McKellop, H. A.: Wear assessment. In The Adult Hip, edited by J. J. Callaghan, A. G. Rosenberg, and H. E. Rubash. Vol. 1, pp. 23l-246. Philadelphia, Lippincott-Raven, 1998 
     
    McKellop, H.; Shen, F.-W.; Yu, Y.-J.; and Lu, B.: Comparison of long-term oxidation of conventional and Hylamer UHMWPE. Trans. Soc. Biomater.,,21: 127, 1998.21127  1998 
     
    McKellop, H.; Shen, F.-W.; DiMaio, W.; and Lancaster, J. G.: Wear of gamma-crosslinked polyethylene acetabular cups against roughened femoral balls. Clin. Orthop.,369: 73-82, 1999.36973  1999  [PubMed]
     
    McKellop, H.; Shen, F.-W.; Lu, B.; Campbell, P.; and Salovey, R.: Development of an extremely wear-resistant ultra high molecular weight polyethylene for total joint replacements. J. Orthop. Res.,17: 157-167, 1999.17157  1999  [PubMed]
     
    McKellop, H. A.; Shen, F.-W.; Campbell, P.; and Ota, T.: Effect of molecular weight, calcium stearate, and sterilization methods on the wear of ultra high molecular weight polyethylene acetabular cups in a hip joint simulator. J. Orthop. Res.,,17: 329-339, 1999.17329  1999 
     
    Muratoglu, O. K.; Bragdon, C. R.; O'Connor, D. O.; Jasty, M.; Harris, W. H.; Gul, R.; and McGarry, F.: Unified wear model for highly crosslinked ultra-high molecular weight polyethylenes (UHMWPE). Biomaterials,,20: 1463-1470, 1999.201463  1999 
     
    Noordin, S.; Schmalzried, T. P.; Campbell, P.; and Amstutz, H. C.: Synovial fluid from patients with prosthetic joint arthroplasty: protein concentration and in vivo wear of polyethylene. Trans. Orthop. Res. Soc.,22: 769, 1997.22769  1997 
     
    Oonishi, H.; Takayama, Y.; and Tsuji, E.: Improvement of polyethylene by irradiation in artificial joints. Radiat. Phys. Chem.,39: 495-504, 1992. .39495  1992.  
     
    Paul, J. P.:: Forces transmitted by joints in the human body. Proc. Inst. Mech. Eng.,,181(3J): 8-15, 1967.181(3J)8  1967 
     
    Poggie, R. A.; Mishra, A. K.; and Davidson, J. A.: 3-body abrasive wear behavior of orthopedic implant bearing surfaces from titanium debris. J. Mater. Sci. Mater. Med.,,5: 387-392, 1994.5387  1994 
     
    Premnath, V.; Harris, W. H.; Jasty, M.; and Merrill, E. W.: Gamma sterilization of UHMWPE articular implants: an analysis of the oxidation problem. Biomaterials,,17: 1741-1753, 1996.171741  1996 
     
    Rimnac, C. M.; Klein, R. W.; Betts, F.; and Wright, T. M.: Post-irradiation aging of ultra-high molecular weight polyethylene. J. Bone and Joint Surg., ,76-A: 1052-1056, July 1994.76-A1052  1994 
     
    Roe, R. J.; Grood, E. S.; Shastri, R.; Gosselin, C. A.; and Noyes, F. R.: Effect of radiation sterilization and aging on ultrahigh molecular weight polyethylene. J. Biomed. Mater. Res.,15: 209-230, 1981.15209  1981  [PubMed]
     
    Rose, R. M.; Nusbaum, H. J.; Schneider, H.; Ries, M.; Paul, I.; Crugnola, A.; Simon, S. R.; and Radin, E. L.: On the true wear rate of ultra high-molecular-weight polyethylene in the total hip prosthesis. J. Bone and Joint Surg.,62-A: 537-549, June 1980.62-A537  1980 
     
    Rose, R. M.; Cimino, W. R.; Ellis, E.; and Crugnola, A. N.: Exploratory investigations on the structure dependence of the wear resistance of polyethylene. Wear, ,77: 89-104, 1982.7789  1982 
     
    Sauer, W., and Salehi, A.: Effects of varying lubricant, frequency, and motion in hip simulator wear testing of UHMWPE. Trans. Soc. Biomater.,21: 416, 1998.21416  1998 
     
    Schmalzried, T. P.; Jasty, M.; and Harris, W. H.: Periprosthetic bone loss in total hip arthroplasty. Polyethylene wear debris and the concept of the effective joint space. J. Bone and Joint Surg.,74-A: 849-863, July 1992.74-A849  1992 
     
    Schmalzried, T. P.; Szuszczewicz, E. S.; Northfield, M. R.; Akizuki, K. H.; Frankel, R. E.; Belcher, G.; and Amstutz, H. C.: Quantitative assessment of walking activity after total hip or knee replacement. J. Bone and Joint Surg., 80-A: 54-59, Jan 1998. 80-A54  1998 
     
    Schmalzried, T. P.; Dorey, F. J.; and McKellop, H.: The multifactorial nature of polyethylene wear in vivo. J. Bone and Joint Surg. ,80-A: 1234-1243, Aug 1998.80-A1234  1998 
     
    Shaver, S. M.; Brown, T. D.; Hillis, S. L.; and Callaghan, J. J.: Digital edge-detection measurement of polyethylene wear after total hip arthroplasty. J. Bone and Joint Surg.,79-A: 690-700, May 1997.79-A690  1997 
     
    Shen, C., and Dumbleton, J. H.: The friction and wear behavior of irradiated very high molecular weight polyethylene. Wear,30: 349-364, 1974.30349  1974 
     
    Shen, F.-W.; Yu, Y.-J.; and McKellop, H.: Potential errors in FTIR measurement of oxidation in ultrahigh molecular weight polyethylene implants. J. Biomed. Mater. Res.,48: 203-210, 1999.48203  1999  [PubMed]
     
    Sperling, D. K.; Collier, J. P.; Mayor, M. B.; Sutula, L. C.; and Bartholf, H. A.: Accelerated aging of UHMWPE. Trans. Orthop. Res. Soc.,22: 771, 1997.22771  1997 
     
    Streicher, R. M.: Influence of ionizing irradiation in air and nitrogen for sterilization of surgical grade polyethylene for implants. Radiat. Phys. Chem.,31: 693-698, 1988.31693  1988 
     
    Sun, D. C.; Stark, C.; and Dumbleton, J. H.: Development of an accelerated aging method for evaluation of long-term irradiation effects on UHMWPE implants. Polymer Preprints, Am. Chem. Soc.,35: 969-970, 1994.35969  1994 
     
    Sun, D. C.; Halleck, A.; Schmidig, G.; Wang, A.; Stark, C.; and Dumbleton, J. H.: FTIR oxidation analysis of UHMWPE implants: possible contamination from synovial fluid and serum. In Characterization and Properties of Ultra-High Molecular Weight Polyethylene, pp. 39-45. ASTM Special Technical Publication 1307. Edited by R. A. Gsell, H. L. Stein, and J. J. Ploskonka. Philadelphia, American Society for Testing and Materials, 1998.  
     
    Sutula, L. C.; Collier, J. P.; Saum, K. A.; Currier, B. H.; Currier, J. H.; Sanford, W. M.; Mayor, M. B.; Wooding, R. E.; Sperling, D. K.; Williams, I. R.; Kasprzak, D. J.; and Surprenant, V. A.: Impact of gamma sterilization on clinical performance of polyethylene in the hip. Clin. Orthop.,319: 28-40, 1995.31928  1995  [PubMed]
     
    Sychterz, C. J.; Engh, C. A., Jr.; Shah, N.; and Engh, C. A., Sr.: Radiographic evaluation of penetration by the femoral head into the polyethylene liner over time. J. Bone and Joint Surg.,79-A: 1040-1046, July 1997.79-A1040  1997 
     
    Sychterz, C. J.; Shah, N.; and Engh, C. A.: Examination of wear in Duraloc acetabular components: two- to five-year evaluation of Hylamer and Enduron liners. J. Arthroplasty,13: 508-514, 1998.13508  1998  [PubMed]
     
    Trieu, H. H.; Haggard, W. O.; Parr, J. E.; Needham, D. J.; Buchanon, D. J.; Rouleau, J. P.; and Morris, L. H.: Accelerated fatigue wear of UHMWPE tibial components caused by radiation-induced oxidation. Trans. Soc. Biomater.,20: 44, 1997.2044  1997 
     
    Wallbridge, N., and Dowson, D.: The walking activity of patients with artificial hip joints. Eng. Med.,11: 95-96, 1982.1195  1982  [PubMed]
     
    Wang, A.; Stark, C.; and Dumbleton, J. H.: Role of cyclic plastic deformation in the wear of UHMWPE acetabular cups. J. Biomed. Mater. Res.,29: 619-626, 1995.29619  1995  [PubMed]
     
    Wang, A.; Essner, A.; Polineni, V. K.; Stark, C.; and Dumbleton, J. H.: Lubrication and wear of UHMWPE in total joint replacements. Tribol. Internat.,31: 17-33, 1998.3117  1998 
     
    Wang, A.; Polineni, V. K.; Essner, A.; Stark, C.; and Dumbleton, J. H.: Quantitative analysis of serum degradation and its effect on the outcome of hip joint simulator wear testing of UHMWPE. Trans. Orthop. Res. Soc.,24: 73, 1999.2473  1999 
     
    Willert, H. G.: Tissue reactions around joint implants and bone cement. In Arthroplasty of the Hip, pp. 11-21. Edited by G. Chapchal. Stuttgart, Thieme, 1973.  
     
    Yeom, B.; Yu, Y.-J.; McKellop, H. A.; and Salovey, R.: Profile of oxidation in irradiated polyethylene. J. Polymer Sci., Part A, Polymer Chem.,36: 329-339, 1998.36329  1998 
     
    Ziegler, J. F.; Biersack, J. P.; and Littmark, U.: The Stopping and Range of Ions in Solids. Vol. 1. New York,Pergamon Press, 1985. 
     

    Submit a comment

    Topics

    Anchor for JumpAnchor for Jump
    +Fig. 1:Flowchart showing the production of the ultra-high molecular weight polyethylene (UHMWPE) as modified by the orthopaedic implant manufacturers. Beginning with a single lot of extruded bar stock produced from a single batch of GUR 4150 resin, the polyethylene was routed through the five participating manufacturers. The cups that were used as nonsterilized controls and those sterilized with irradiation in air were machined by Zimmer as a service to the study, but they do not represent Zimmer's current product. All of the gamma sterilization was done with the cups in a single container to provide an equal, simultaneous radiation dose.
    Anchor for JumpAnchor for Jump
    +Fig. 2:Graph showing the oxidation levels from the inner bearing surface to the center of the cup wall (left to right) obtained from an ultra-high molecular weight polyethylene cup after seventeen years of shelf-aging and from the nonworn zone of an ultra-high molecular weight polyethylene cup retrieved after thirteen years of use in vivo. Both cups were originally sterilized with gamma irradiation in air. These oxidation levels are among the highest that have been measured in our laboratory for cups sterilized with gamma irradiation in air. (The microtomed sections from the retrieved cup were soaked in hexane for sixteen hours to reduce contaminants prior to Fourier transform infrared analysis.)
    Anchor for JumpAnchor for Jump
    +Fig. 3:Graph showing the oxidation levels from the inner surface to the center of the cup wall (left to right) in the nonworn zones of the cups tested without artificial aging. (The corresponding wear volumes and rates are shown in Figure 4-A and Figure 4-B.) Two cups of each material were analyzed. Since the oxidation profiles were comparable for the two cups in a pair, only the mean oxidation levels are shown for clarity.
    Anchor for JumpAnchor for Jump
    +Fig. 4-A:Graph showing the volumetric wear of the two cups of each material tested without artificial aging. (The corresponding oxidation profiles are shown in Figure 3.) The scale at right indicates the approximate maximum depth of wear - that is, the depth at the center of the contact zone.
    Anchor for JumpAnchor for Jump
    +Fig. 4-B:Graph showing the mean wear rates and the standard deviations for the six different types of polyethylene and for the two cycle intervals without artificial aging.
    Anchor for JumpAnchor for Jump
    +Fig. 5:Graph showing the oxidation levels from the inner surface to the center of the cup wall (left to right) in the nonworn zones of the cups tested after fourteen days of artificial aging at 80 degrees Celsius. (The corresponding wear volumes and rates are shown in Figure 6-A and Figure 6-B.) Since the levels of oxidation were comparable for the two cups in a pair, only the mean oxidation levels are shown for each material for clarity.
    Anchor for JumpAnchor for Jump
    +Fig. 6-A:Graph showing the volumetric wear of the two cups of each material tested after fourteen days of artificial aging at 80 degrees Celsius. (The corresponding oxidation profiles are shown in Figure 5.) The scale at right indicates the approximate maximum depth of wear - that is, the depth at the center of the contact zone.
    Anchor for JumpAnchor for Jump
    +Fig. 6-B:Graph showing the mean wear rates and the standard deviations for the six different types of polyethylene and for the two cycle intervals after fourteen days of artificial aging at 80 degrees Celsius.
    Anchor for JumpAnchor for Jump
    +Fig. 7:Graph showing the oxidation levels from the inner surface to the center of the cup wall (left to right) in the nonworn zones of the cups tested after thirty days of artificial aging at 80 degrees Celsius. (The corresponding wear volumes and rates are shown in Figure 8-A and Figure 8-B.) When the oxidation profiles were comparable for the two cups in a pair, only the mean oxidation levels are shown for clarity. The profiles are shown for both cups made of ion-implanted polyethylene and both made of polyethylene irradiated with an oxygen scavenger since the two cups in each of these pairs exhibited substantially different oxidation levels.
    Anchor for JumpAnchor for Jump
    +Fig. 8-A:Graph showing the volumetric wear of the two cups of each material tested after thirty days of artificial aging at 80 degrees Celsius. (The corresponding oxidation profiles are shown in Figure 7.) The scale at right indicates the approximate maximum depth of wear - that is, the depth at the center of the contact zone.
    Anchor for JumpAnchor for Jump
    +Fig. 8-B:Graph showing the mean wear rates and the standard deviations for the six different types of polyethylene and for the three cycle intervals after thirty days of artificial aging at 80 degrees Celsius.
    Anchor for JumpAnchor for JumpTABLE I:  Modified Polyethylene as Fabricated by Orthopaedic Manufacturers
    *All of the cups sterilized by gamma irradiation were placed in a single container and irradiated to 2.72 megarad (27.2 kilogray) by a commercial sterilizer.†The Abtox Plazlyte sterilization system is manufactured by Abtox, Mundelein, Illinois.
    Type of PolyethyleneGroup SymbolFabrication Method and SourceSterilization Method*
    Nonsterilized controlsCMachined from extruded bar stock at Zimmer None
    Hylamer, gas plasmaHExtruded bar stock converted to Hylamer at DePuy-DuPont Orthopaedics; cups machined from bar stockGas-plasma sterilized by DePuy-DuPont with use of an Abtox Plazlyte unit†
    Irradiated in airAMachined from extruded bar stock at ZimmerGamma-irradiated in air to 2.72 megarad (27.2 kilogray)
    Ion-implanted (IONGUARD III and irradiated in air IMachined from extruded bar stock at Zimmer; implanted with nitrogen ions at Spire Gamma-irradiated in air to 2.72 megarad (27.2 kilogray)
    Irradiated in nitrogen and thermally stabilized (Duration)NMachined from extruded bar stock at Howmedica Packaged in nitrogen and gamma- irradiated to 2.72 megarad (27.2 kilogray), then returned to Howmedica for thermal stabilization by heating to 50 degrees Celsius for 6 days while still packaged in nitrogen
    Irradiated with oxygen scavenger (Oxygenless Packaging)OMachined from extruded bar stock at Intermedics OrthopedicsPackaged in foil with an oxygen scavenger (nontoxic, activated iron particles) in a Tyvek (high-density polyethylene) pouch, and gamma- irradiated to 2.72 megarad (27.2 kilogray)
    Anchor for JumpAnchor for JumpTABLE II:  Wear Rates of Acetabular Cups without Artificial Aging
    *The values are given in cubic millimeters per million cycles.
    Type of CupCup No.Between 0.5 and 3.0 Million Cycles*Between 3.0 and 5.0 Million Cycles*
    Wear RateMean and Standard DeviationWear RateMean and Standard Deviation
    Nonsterilized controlsC1C234.936.935.9 ± 1.440.642.141.4 ± 1.0
    Hylamer, gas plasmaH1H235.034.034.5 ± 0.736.233.334.8 ± 2.1
    Irradiated in airA1A237.636.036.8 ± 1.126.127.326.7 ± 0.8
    Ion-implanted (IONGUARD III) and irradiated in air I1I225.025.325.2 ± 0.224.224.124.2 ± 0.1
    Irradiated in nitrogen and thermally stabilized (Duration)N1N218.615.617.1 ± 2.123.019.921.5 ± 2.2
    Irradiated with oxygen scavenger (Oxygenless Packaging)O1O218.616.917.8 ± 1.218.517.117.8 ± 1.0
    Anchor for JumpAnchor for JumpTABLE III:  Comparisons of the Wear Rates of Various Materials without Artificial Aging
    *C = nonsterilized controls; H = Hylamer, gas plasma; A = irradiated in air; I = ion-implanted and irradiated in air; N = irradiated in nitrogen and thermally stabilized; and O = irradiated with oxygen scavenger.†Power is shown in parentheses.
    Compared Groups*P Value for Difference Between Groups†
    At 0.5 to 3.0 million cycles
      C > H0.35 (99.9%)
      C < A0.56
      C > I0.009
      C > N0.009
      C > O0.005
      H < A0.14
      H > I0.003
      H > N0.008
      H > O0.004
      A > I0.005
      A > N0.007
      A > O0.004
      I > N0.03
      I > O0.014
      N < O0.75 (84%)
    At 3.0 to 5.0 million cycles
      C > H 0.06 (74%)
      C > A 0.004
      C > I 0.002
      C > N0.007
      C > O 0.002
      H > A 0.04
      H > I0.02
      H > N0.02
      H > O0.009
      A > I 0.05
      A > N 0.09
      A > O 0.01
      I > N 0.22
      I > O0.01
      N > O 0.16 (61%)
    Anchor for JumpAnchor for JumpTABLE IV:  Wear Rates of Acetabular Cups After Artificial Aging at 80 Degrees Celsius in Air for Fourteen Days
    *The values are given in cubic millimeters per million cycles.
    Type of CupCup No.Between 0.5 and 3.5 Million Cycles*Between 3.5 and 5.0 Million Cycles*
    Wear Rate  Mean and Standard DeviationWear RateMean and Standard Deviation
    Nonsterilized controlsC3C426.124.325.2 ± 1.229.625.927.8 ± 2.6
    Hylamer, gas plasmaH3H427.529.528.5 ± 1.431.632.632.1 ± 0.7
    Irradiated in airA3A481.857.9  69.9 ± 16.982.342.362.3 ± 28.3
    Ion-implanted (IONGUARD III) and irradiated in airI3I428.229.829.0 ± 1.160.052.956.5 ± 5.0
    Irradiated in nitrogen and thermally stabilized (Duration)N3N413.413.213.3 ± 0.219.215.117.2 ± 2.9
    Irradiated with oxygen scavenger (Oxygenless Packaging)O3O414.314.014.2 ± 0.213.614.113.9 ± 0.4
    Anchor for JumpAnchor for JumpTABLE V:  Comparisons of the Wear Rates of the Various Materials After Fourteen Days of Artificial Aging
    *C = nonsterilized controls; H = Hylamer, gas plasma; A = irradiated in air; I = ion-implanted and irradiated in air; N = irradiated in nitrogen and thermally stabilized; and O = irradiated with oxygen scavenger.†Power is shown in parentheses.
    Compared Groups*P Value for Difference Between Groups†
    At 0.5 to 3.5 million cycles
      C < H0.13 (88%)
      C < A0.07
      C < I0.08
      C > N0.005
      C > O0.006
      H < A0.08
      H < I0.75
      H > N0.004
      H > O0.005
      A > I0.08
      A > N0.04
      A > O0.04
      I > N0.002
      I > O0.003
      N < O0.05 (99.9%)
    At 3.5 to 5.0 million cycles
      C < H0.15 (77%)
      C < A0.23
      C < I0.02
      C > N 0.06
      C > O 0.02
      H < A 0.27
      H < I0.02
      H > N 0.02
      H > O 0.001
      A > I 0.8
      A > N 0.15
      A > O 0.14
      I > N 0.01
      I > O 0.007
      N > O 0.25 (46%)
    Anchor for JumpAnchor for JumpTABLE VI:  Wear Rates of Acetabular Cups After Artificial Aging at 80 Degrees Celsius in Air for Thirty Days
    *The values are given in cubic millimeters per million cycles.
    Type of Cup  Cup No.Between 0.5 and 3.0 Million Cycles*Between 3.0 and 5.0 Million Cycles*Between 5.0 and 7.5 Million Cycles*
      Wear RateMean and Standard Deviation  WearRateMean and Standard Deviation  Wear RateMean and Standard Deviation
    Nonsterilized controlsC5C633.333.033.2 ± 0.2  34.9 38.3  36.6 ± 2.433.936.935.4 ± 2.1
    Hylamer, gas plasmaH5H645.732.539.1 ± 9.4  41.8 37.3  39.6 ± 3.236.936.236.6 ± 0.5
    Irradiated in airA5A682.165.373.7 ± 11.8  82.7 68.4  75.6 ± 10.161.072.866.9 ± 8.3
    Ion-implanted (IONGUARD III) and irradiated in air I5I646.351.448.9 ± 3.6105.5112.2108.9 ± 4.783.968.876.4 ± 10.7
    Irradiated in nitrogen and thermally stabilized (Duration)N5N661.843.652.7 ± 12.9  69.0 58.7  63.9 ± 7.348.951.650.3 ± 1.9
    Irradiated with oxygen scavenger (Oxygenless Packaging)O5O649.320.835.1 ± 20.2  90.9 31.7  61.3 ± 41.990.071.280.6 ± 13.3
    Anchor for JumpAnchor for JumpTABLE VII:  Comparisons of Wear Rates of the Various Materials After Thirty Days of Artificial Aging
    *C = nonsterilized controls; H = Hylamer, gas plasma; A = irradiated in air; I = ion-implanted and irradiated in air; N = irradiated in nitrogen and thermally stabilized; and O = irradiated with oxygen scavenger.†Power is shown in parentheses.
      Compared Groups*P Value for Difference Between Groups†
    At 0.5 to 3.0 million cycles
      C < H0.46 (40%)
      C < A 0.04
      C < I0.03
      C < N 0.17
      C < O 0.9
      H < A 0.08
      H < I0.30
      H < N 0.35
      H > O 0.82
      A > I0.10
      A > N 0.23
      A > O 0.14
      I < N 0.72
      I > O0.44
      N > O 0.41
    At 3.0 to 5.0 million cycles
      C < H 0.41 (91%)
      C < A 0.03
      C < I0.003
      C < N 0.04
      C < O 0.49
      H < A 0.04
      H < I0.003
      H < N0.05
      H < O 0.54
      A < I 0.05
      A > N 0.32
      A > O 0.69
      I > N 0.02
      I > O 0.25
      N > O0.94
    At 5.0 to 7.5 million cycles
      C < H 0.53 (99.9%)
      C < A 0.04
      C < I0.03
      C < N 0.02
      C < O 0.04
      H < A 0.04
      H < I 0.03
      H < N 0.01
      H < O 0.04
      A < I0.43
      A > N 0.11
      A < O 0.34
      I > N 0.08
      I < O 0.76
      N < O 0.09
    Anchor for JumpAnchor for JumpTABLE VIII:  Results of Wear Debris Analysis
    *The values are given as the mean and the standard deviation, with the composition of the debris in parentheses.
    Type of CupSize and Composition of Debris* (m)
    Without Aging14 Days of Aging30 Days of Aging
    GranulesFibrilsFlakesGranulesFibrilsFlakesGranulesFibrilsFlakes
    Nonsterilized controls0.21 ± 0.04 (66%)1.9 ± 1.1 (26%)0.34 ± 0.1 (7%)0.23 ± 0.05 (85%)2.1 ± 1.1 (14%)0.36 ± 0.03 (1%)0.22 ± 0.04 (81%)2.4 ± 1.2 (6%)0.34 ± 0.09 (14%)
    Hylamer, gas plasma 0.23 ± 0.06 (60%)1.9 ± 1.1 (30%)0.41 ± 0.1 (9%)0.23 ± 0.04 (74%)2.3 ± 1.2 (23%)0.38 ± 0.13 (2%)0.23 ± 0.04 (69%)2.0 ± 1.1 (15%)0.39 ± 0.10 (14%)
    Irradiated in air0.20 ± 0.04 (83%)1.5 ± 0.8 (14%)0.36 ± 0.1 (3%)0.22 ± 0.05 (90%)1.9 ± 1.1 (7%)0.29 ± 0.12 (5%)0.21 ± 0.04 (79%)1.7 ± 0.8 (11%)0.35 ± 0.08 (9%)
    Ion-implanted (IONGUARD III) and irradiated in air0.21 ± 0.04 (81%)1.5 ± 1.0 (14%)0.41 ± 0.1 (4%)0.23 ± 0.07 (77%)2.1 ± 1.2 (18%)0.44 ± 0.09 (5%)0.23 ± 0.04 (73%)1.9 ± 0.9 (10%)0.38 ± 0.11 (18%)
    Irradiated in nitrogen and thermally stabilized (Duration)0.21 ± 0.07 (61%)2.0 ± 1.5 (30%)0.56 ± 0.6 (9%)0.23 ± 0.06 (77%)1.8 ± 0.8 (18%)0.43 ± 0.09 (5%)0.19 ± 0.03 (83%)1.6 ± 0.6 (5%)0.29 ± 0.06 (11%)
    Irradiated with oxygen scavenger (Oxygenless Packaging)0.20 ± 0.04 (75%)1.6 ± 1.0 (17%)0.44 ± 0.2 (8%)0.24 ± 0.06 (76%)2.0 ± 0.8 (20%)0.39 ± 0.11 (4%)0.19 ± 0.04 (90%)1.6 ± 0.8 (4%)0.34 ± 0.12 (7%)
    Amstutz, H. C.; Campbell, P.; Kossovsky, N.; and Clarke, I. C.: Mechanisms and clinical significance of wear debris-induced osteolysis. Clin. Orthop.,276: 7-18, 1992.2767  1992  [PubMed]
     
    Bapst, J. M.; Valentine, R. H.; and Vasquez, R.: Wear simulation testing of direct compression molded UHMWPE irradiated in oxygenless packaging. Trans. Soc. Biomater.,20: 72, 1997.2072  1997 
     
    Bauer, T. W.; Taylor, S. K.; Jiang, M.; and Medendorp, S. V.: An indirect comparison of third-body wear in retrieved hydroxyapatite-coated, porous, and cemented femoral components. Clin. Orthop.,298: 11-18, 1994.29811  1994  [PubMed]
     
    Bhateja, S. K.; Duerst, R. W.; Martens, J. A.; and Andrews, E. H.: Radiation-induced enhancement of crystallinity in polymers. J. Macromol. Sci.-Revs. Chem. Phys.,C35: 581-659, 1995.C35581  1995 
     
    Campbell, P.; Ma, S.; Yeom, B.; McKellop, H.; Schmalzried, T. P.; and Amstutz, H. C.: Isolation of predominantly submicron-sized UHMWPE wear particles from periprosthetic tissues. J. Biomed. Mater. Res.,29: 127-131, 1995.29127  1995  [PubMed]
     
    Chmell, M. J.; Poss, R.; Thomas, W. H.; and Sledge, C. B.: Early failure of Hylamer acetabular inserts due to eccentric wear. J. Arthroplasty,11: 351-353, 1996.11351  1996  [PubMed]
     
    Clarke, I. C.; Campbell, P.; and Kossovsky, N.: Debris-mediated osteolysis - A cascade phenomenon involving motion, wear, particulates, macrophage induction and bone lysis. In Particulate Debris from Medical Implants, pp. 7-26. Edited by K. St. John. ASTM Special Technical Publications 1144. Philadelphia, American Society for Testing and Materials, 1992 
     
    Collier, J. P.; Sperling, D. K.; Currier, J. H.; Sutula, L. C.; Saum, K. A.; and Mayor, M. B.: Impact of gamma sterilization on clinical performance of polyethylene in the knee. J. Arthroplasty,11: 377-389, 1996.11377  1996  [PubMed]
     
    Collier, J. P.; Bargmann, L. S.; Currier, B. H.; Mayor, M. B.; Currier, J. H.; and Bargmann, B. C.: An analysis of Hylamer and polyethylene bearings from retrieved acetabular components. Orthopedics,,21: 865-871, 1998.21865  1998 
     
    Devane, P. A.; Robinson, E. J.; Bourne, R. B.; Rorabeck, C. H.; Nayak, N. N.; and Horne, J. G.: Measurement of polyethylene wear in acetabular components inserted with and without cement. A randomized trial. J. Bone and Joint Surg.,79-A: 682-689, May 1997.79-A682  1997 
     
    Diwan, A., and Drummond, R.: Unusual case of third-body wear in total hip arthroplasty. J. Arthroplasty,12: 586-588, 1997.12586  1997  [PubMed]
     
    Dumbleton, J. H.; Shen, C.; and Miller, E. H.: Study of the wear of some materials in connection with total hip replacement. Wear, ,29: 163-171, 1974.29163  1974 
     
    Eyerer, P., and Ke, Y. C.: Property changes of UHMW polyethylene hip cup endoprostheses during implantation. J. Biomed. Mater. Res.,18: 1137-1151, 1984.181137  1984  [PubMed]
     
    Eyerer, P.; Kurth, M.; McKellop, H. A.; and Mittlmeier, T.: Characterization of UHMWPE hip cups run on joint simulators. J. Biomed. Mater. Res.,21: 275-291, 1987.21275  1987  [PubMed]
     
    Fisher, J.; Reeves, E. A.; Isaac, G. H.; Saum, K. A.; and Sanford, W. M.: Comparison of the wear of aged and non-aged UHMWPE sterilized by gamma irradiation and by gas plasma. J. Mater. Sci. Mater. Med.,8: 375-378, 1997.8375  1997  [PubMed]
     
    Furman, B. D.; Kasprzak, D.; and Li, S.: Differences in oxidation between shelf life aged and retrieved ultra high molecular weight polyethylene components. Trans. Orthop. Res. Soc.,21: 484, 1996.21484  1996 
     
    Gillis, A. M.; Furman, B. D.; and Li, S.: Influence of ultra high molecular weight polyethylene resin type and manufacturing method on real time oxidation. Trans. Orthop. Res. Soc.,23: 360, 1998.23360  1998 
     
    Goldring, S. R.; Jasty, M.; Roelke, M. S.; Rourke, C. M.; Bringhurst, F. R.; and Harris, W. H.: Formation of a synovial-like membrane at the bone-cement interface. Its role in bone resorption and implant loosening after total hip replacement. Arthrit. and Rheumat.,29: 836-842, 1986.29836  1986 
     
    Good, V.; Clarke, I. C.; Gustafson, A.; Braham, A.; Downs, B.; Sorensen, K.; and Anissian, L.: Dose response of protein concentration to wear in PTFE and UHMWPE. Trans. Orthop. Res. Soc.,24: 839, 1999.24839  1999 
     
    Goodman, S.: The role of polymer particles in the process of aseptic loosening of joint arthroplasties. In Surface Modification Technologies V, pp. 36-48. Edited by T. Sudarshon and J. Braza. London, Institute of Materials, 1992.  
     
    Greer, K. W.; Schmidt, M. B.;, and Hamilton, J. V.:: The hip simulator wear of gamma-vacuum, gamma-air, and ethylene oxide sterilized UHMWPE following a severe oxidative challenge. Trans. Orthop. Res. Soc., ,23: 52, 1998.2352  1998 
     
    Grobbelaar, C. J.; du Plessis, T. A.; and Marais, F.: The radiation improvement of polyethylene prostheses. A preliminary study. J. Bone and Joint Surg.,60-B(3): 370-374, 1978.60-B(3)370  1978 
     
    Hop, J. D.; Callaghan, J. J.; Olejniczak, J. P.; Pedersen, D. R.; Brown, T. D.; and Johnston, R. C.: Contribution of cable debris generation to accelerated polyethylene wear. Clin. Orthop.,344: 20-32, 1997.34420  1997  [PubMed]
     
    Howie, D. W.; Vernon-Roberts, B.; Oakeshott, R.; and Manthey, B.: A rat model of resorption of bone at the cement-bone interface in the presence of polyethylene wear particles. J. Bone and Joint Surg.,70-A: 257-263, Feb 1988.70-A257  1988 
     
    James, S. P.; Blazka, S.; Merrill, E. W.; Jasty, M.; Lee, K. R.; Bragdon, C. R.; and Harris, W. H.: Challenge to the concept that UHMWPE acetabular components oxidize in vivo. Biomaterials,14: 643-647, 1993.14643  1993  [PubMed]
     
    Kabo, J. M.; Gebhard, J. S.; Loren, G.; and Amstutz, H. C.: In vivo wear of polyethylene acetabular components. J. Bone and Joint Surg.,75-B(2): 254-258, 1993.75-B(2)254  1993 
     
    Kurtz, S. M.; Muratoglu, O. K.; Evans, M.; and Edidin, A. A.: Advances in the processing, sterilization, and crosslinking of ultra-high molecular weight polyethylene for total joint arthroplasty. Biomaterials,20: 1659-1688, 1999.201659  1999  [PubMed]
     
    Lee, E. H.; Rao, G. R.; Lewis, M. B.; and Mansur, L. K.: Ion beam application for improved polymer surface properties. Nucl. Instrum. Meth.,B74: 326-330, 1993.B74326  1993 
     
    Li, S.,, and Burstein, A. H.: Ultra-high molecular weight polyethylene. The material and its use in total joint implants. J. Bone and Joint Surg. ,76-A: 1080-1090, July 1994.76-A1080  1994 
     
    Liao, Y. S.; Benya, P. D.; and McKellop, H. A.: Effect of protein lubrication on the wear properties of materials for prosthetic joints. J. Biomed. Mater. Res.,48: 465-473, 1999.48465  1999  [PubMed]
     
    Livingston, B. J.; Chmell, M. J.; Spector, M.; and Poss, R.: Complications of total hip arthroplasty associated with the use of an acetabular component with a Hylamer liner. J. Bone and Joint Surg.,,79-A: 1529-1538, Oct 1997.79-A1529  1997 
     
    McKellop, H.; Clarke, I.; Markolf, K.; and Amstutz, H.: Friction and wear properties of polymer, metal, and ceramic prosthetic joint materials evaluated on a multichannel screening device. J. Biomed. Mater. Res.,15: 619-653, 1981.15619  1981  [PubMed]
     
    McKellop, H., and Clarke, I.: Evolution and evaluation of materials screening machines and joint simulators in predicting in vivo wear phenomena. In Functional Behavior of Orthopaedic Biomaterials, pp. 51-85. Edited by P. Ducheyne and G. Hastings. Boca Raton, CRC Press, 1984 
     
    McKellop, H. A.; Lu, B.; and Li, S.: Wear of acetabular cups of conventional and modified UHMW polyethylenes compared on a hip joint simulator. Trans. Orthop. Res. Soc.,17: 356, 1992.17356  1992 
     
    McKellop, H. A.: Wear modes, mechanisms, damage, and debris. Separating cause from effect in the wear of total hip replacements. In Total Hip Revision Surgery, pp. 21-39. Edited by J. O. Galante, A. G. Rosenberg, and J. J. Callaghan. New York, Raven Press, 1995.  
     
    McKellop, H. A.; Campbell, P.; Park, S.-H.; Schmalzried, T. P.; Grigoris, P.; Amstutz, H. C.; and Sarmiento, A.: The origin of submicron polyethylene wear debris in total hip arthroplasty. Clin. Orthop.,311: 3-20, 1995.3113  1995  [PubMed]
     
    McKellop, H. A.: Wear assessment. In The Adult Hip, edited by J. J. Callaghan, A. G. Rosenberg, and H. E. Rubash. Vol. 1, pp. 23l-246. Philadelphia, Lippincott-Raven, 1998 
     
    McKellop, H.; Shen, F.-W.; Yu, Y.-J.; and Lu, B.: Comparison of long-term oxidation of conventional and Hylamer UHMWPE. Trans. Soc. Biomater.,,21: 127, 1998.21127  1998 
     
    McKellop, H.; Shen, F.-W.; DiMaio, W.; and Lancaster, J. G.: Wear of gamma-crosslinked polyethylene acetabular cups against roughened femoral balls. Clin. Orthop.,369: 73-82, 1999.36973  1999  [PubMed]
     
    McKellop, H.; Shen, F.-W.; Lu, B.; Campbell, P.; and Salovey, R.: Development of an extremely wear-resistant ultra high molecular weight polyethylene for total joint replacements. J. Orthop. Res.,17: 157-167, 1999.17157  1999  [PubMed]
     
    McKellop, H. A.; Shen, F.-W.; Campbell, P.; and Ota, T.: Effect of molecular weight, calcium stearate, and sterilization methods on the wear of ultra high molecular weight polyethylene acetabular cups in a hip joint simulator. J. Orthop. Res.,,17: 329-339, 1999.17329  1999 
     
    Muratoglu, O. K.; Bragdon, C. R.; O'Connor, D. O.; Jasty, M.; Harris, W. H.; Gul, R.; and McGarry, F.: Unified wear model for highly crosslinked ultra-high molecular weight polyethylenes (UHMWPE). Biomaterials,,20: 1463-1470, 1999.201463  1999 
     
    Noordin, S.; Schmalzried, T. P.; Campbell, P.; and Amstutz, H. C.: Synovial fluid from patients with prosthetic joint arthroplasty: protein concentration and in vivo wear of polyethylene. Trans. Orthop. Res. Soc.,22: 769, 1997.22769  1997 
     
    Oonishi, H.; Takayama, Y.; and Tsuji, E.: Improvement of polyethylene by irradiation in artificial joints. Radiat. Phys. Chem.,39: 495-504, 1992. .39495  1992.  
     
    Paul, J. P.:: Forces transmitted by joints in the human body. Proc. Inst. Mech. Eng.,,181(3J): 8-15, 1967.181(3J)8  1967 
     
    Poggie, R. A.; Mishra, A. K.; and Davidson, J. A.: 3-body abrasive wear behavior of orthopedic implant bearing surfaces from titanium debris. J. Mater. Sci. Mater. Med.,,5: 387-392, 1994.5387  1994 
     
    Premnath, V.; Harris, W. H.; Jasty, M.; and Merrill, E. W.: Gamma sterilization of UHMWPE articular implants: an analysis of the oxidation problem. Biomaterials,,17: 1741-1753, 1996.171741  1996 
     
    Rimnac, C. M.; Klein, R. W.; Betts, F.; and Wright, T. M.: Post-irradiation aging of ultra-high molecular weight polyethylene. J. Bone and Joint Surg., ,76-A: 1052-1056, July 1994.76-A1052  1994 
     
    Roe, R. J.; Grood, E. S.; Shastri, R.; Gosselin, C. A.; and Noyes, F. R.: Effect of radiation sterilization and aging on ultrahigh molecular weight polyethylene. J. Biomed. Mater. Res.,15: 209-230, 1981.15209  1981  [PubMed]
     
    Rose, R. M.; Nusbaum, H. J.; Schneider, H.; Ries, M.; Paul, I.; Crugnola, A.; Simon, S. R.; and Radin, E. L.: On the true wear rate of ultra high-molecular-weight polyethylene in the total hip prosthesis. J. Bone and Joint Surg.,62-A: 537-549, June 1980.62-A537  1980 
     
    Rose, R. M.; Cimino, W. R.; Ellis, E.; and Crugnola, A. N.: Exploratory investigations on the structure dependence of the wear resistance of polyethylene. Wear, ,77: 89-104, 1982.7789  1982 
     
    Sauer, W., and Salehi, A.: Effects of varying lubricant, frequency, and motion in hip simulator wear testing of UHMWPE. Trans. Soc. Biomater.,21: 416, 1998.21416  1998 
     
    Schmalzried, T. P.; Jasty, M.; and Harris, W. H.: Periprosthetic bone loss in total hip arthroplasty. Polyethylene wear debris and the concept of the effective joint space. J. Bone and Joint Surg.,74-A: 849-863, July 1992.74-A849  1992 
     
    Schmalzried, T. P.; Szuszczewicz, E. S.; Northfield, M. R.; Akizuki, K. H.; Frankel, R. E.; Belcher, G.; and Amstutz, H. C.: Quantitative assessment of walking activity after total hip or knee replacement. J. Bone and Joint Surg., 80-A: 54-59, Jan 1998. 80-A54  1998 
     
    Schmalzried, T. P.; Dorey, F. J.; and McKellop, H.: The multifactorial nature of polyethylene wear in vivo. J. Bone and Joint Surg. ,80-A: 1234-1243, Aug 1998.80-A1234  1998 
     
    Shaver, S. M.; Brown, T. D.; Hillis, S. L.; and Callaghan, J. J.: Digital edge-detection measurement of polyethylene wear after total hip arthroplasty. J. Bone and Joint Surg.,79-A: 690-700, May 1997.79-A690  1997 
     
    Shen, C., and Dumbleton, J. H.: The friction and wear behavior of irradiated very high molecular weight polyethylene. Wear,30: 349-364, 1974.30349  1974 
     
    Shen, F.-W.; Yu, Y.-J.; and McKellop, H.: Potential errors in FTIR measurement of oxidation in ultrahigh molecular weight polyethylene implants. J. Biomed. Mater. Res.,48: 203-210, 1999.48203  1999  [PubMed]
     
    Sperling, D. K.; Collier, J. P.; Mayor, M. B.; Sutula, L. C.; and Bartholf, H. A.: Accelerated aging of UHMWPE. Trans. Orthop. Res. Soc.,22: 771, 1997.22771  1997 
     
    Streicher, R. M.: Influence of ionizing irradiation in air and nitrogen for sterilization of surgical grade polyethylene for implants. Radiat. Phys. Chem.,31: 693-698, 1988.31693  1988 
     
    Sun, D. C.; Stark, C.; and Dumbleton, J. H.: Development of an accelerated aging method for evaluation of long-term irradiation effects on UHMWPE implants. Polymer Preprints, Am. Chem. Soc.,35: 969-970, 1994.35969  1994 
     
    Sun, D. C.; Halleck, A.; Schmidig, G.; Wang, A.; Stark, C.; and Dumbleton, J. H.: FTIR oxidation analysis of UHMWPE implants: possible contamination from synovial fluid and serum. In Characterization and Properties of Ultra-High Molecular Weight Polyethylene, pp. 39-45. ASTM Special Technical Publication 1307. Edited by R. A. Gsell, H. L. Stein, and J. J. Ploskonka. Philadelphia, American Society for Testing and Materials, 1998.  
     
    Sutula, L. C.; Collier, J. P.; Saum, K. A.; Currier, B. H.; Currier, J. H.; Sanford, W. M.; Mayor, M. B.; Wooding, R. E.; Sperling, D. K.; Williams, I. R.; Kasprzak, D. J.; and Surprenant, V. A.: Impact of gamma sterilization on clinical performance of polyethylene in the hip. Clin. Orthop.,319: 28-40, 1995.31928  1995  [PubMed]
     
    Sychterz, C. J.; Engh, C. A., Jr.; Shah, N.; and Engh, C. A., Sr.: Radiographic evaluation of penetration by the femoral head into the polyethylene liner over time. J. Bone and Joint Surg.,79-A: 1040-1046, July 1997.79-A1040  1997 
     
    Sychterz, C. J.; Shah, N.; and Engh, C. A.: Examination of wear in Duraloc acetabular components: two- to five-year evaluation of Hylamer and Enduron liners. J. Arthroplasty,13: 508-514, 1998.13508  1998  [PubMed]
     
    Trieu, H. H.; Haggard, W. O.; Parr, J. E.; Needham, D. J.; Buchanon, D. J.; Rouleau, J. P.; and Morris, L. H.: Accelerated fatigue wear of UHMWPE tibial components caused by radiation-induced oxidation. Trans. Soc. Biomater.,20: 44, 1997.2044  1997 
     
    Wallbridge, N., and Dowson, D.: The walking activity of patients with artificial hip joints. Eng. Med.,11: 95-96, 1982.1195  1982  [PubMed]
     
    Wang, A.; Stark, C.; and Dumbleton, J. H.: Role of cyclic plastic deformation in the wear of UHMWPE acetabular cups. J. Biomed. Mater. Res.,29: 619-626, 1995.29619  1995  [PubMed]
     
    Wang, A.; Essner, A.; Polineni, V. K.; Stark, C.; and Dumbleton, J. H.: Lubrication and wear of UHMWPE in total joint replacements. Tribol. Internat.,31: 17-33, 1998.3117  1998 
     
    Wang, A.; Polineni, V. K.; Essner, A.; Stark, C.; and Dumbleton, J. H.: Quantitative analysis of serum degradation and its effect on the outcome of hip joint simulator wear testing of UHMWPE. Trans. Orthop. Res. Soc.,24: 73, 1999.2473  1999 
     
    Willert, H. G.: Tissue reactions around joint implants and bone cement. In Arthroplasty of the Hip, pp. 11-21. Edited by G. Chapchal. Stuttgart, Thieme, 1973.  
     
    Yeom, B.; Yu, Y.-J.; McKellop, H. A.; and Salovey, R.: Profile of oxidation in irradiated polyethylene. J. Polymer Sci., Part A, Polymer Chem.,36: 329-339, 1998.36329  1998 
     
    Ziegler, J. F.; Biersack, J. P.; and Littmark, U.: The Stopping and Range of Ions in Solids. Vol. 1. New York,Pergamon Press, 1985. 
     
    Accreditation Statement
    These activities have been planned and implemented in accordance with the Essential Areas and policies of the Accreditation Council for Continuing Medical Education (ACCME) through the joint sponsorship of the American Academy of Orthopaedic Surgeons and The Journal of Bone and Joint Surgery, Inc. The American Academy of Orthopaedic Surgeons is accredited by the ACCME to provide continuing medical education for physicians.
    CME Activities Associated with This Article
    Submit a Comment
    Please read the other comments before you post yours. Contributors must reveal any conflict of interest.
    Comments are moderated and will appear on the site at the discretion of JBJS editorial staff.

    * = Required Field
    (if multiple authors, separate names by comma)
    Example: John Doe




    Related Articles
    Related Cases
    Related Content
    Topic Collections
    Hip
    Related Audio and Videos
    PubMed Articles
    Franklin H. Epstein Lecture: Sirtuins, aging, and medicine.
    The New England journal of medicine: Issue date- 2011 Jun 9
    The role of mitochondria in osteoarthritis.
    Nature reviews. Rheumatology: Issue date- 2011 Mar
    Clinical Trials
    Readers of This Also Read...
    jbjs jobs
    12/22/2011
    ME - Central Maine Medical Center
    12/22/2011
    VA - Charleston Area Medical Center