0
Scientific Article   |    
In Vivo Comparison of Hip Separation After Metal-on-Metal or Metal-on-Polyethylene Total Hip Arthroplasty
Richard D. Komistek, PhD; Douglas A. Dennis, MD; Jorge A. Ochoa, PhD; Brian D. Haas, MD; Curt Hammill, BS
The Journal of Bone & Joint Surgery.  2002; 84:1836-1841 
5 Recommendations (Recommend) | 3 Comments | Saved by 3 Users Save Case

Abstract

Background: Twenty subjects were analyzed in vivo with use of video fluoroscopy to determine if the femoral head separates from the acetabular component during normal gait and to determine if the amount of separation differs between metal-on-metal and metal-on-polyethylene total hip prostheses.

Methods: Ten subjects had been treated with a metal-on-metal total hip arthroplasty and ten, with a metal-on-polyethylene total hip arthroplasty. All of the prostheses were implanted by the same surgeon utilizing the same surgical technique, and all were judged to be clinically successful (a Harris hip score of >90 points). Each subject walked with a normal gait on a level treadmill while under fluoroscopic surveillance. The two-dimensional fluoroscopic videotapes were then converted into three-dimensional images with use of a computer-automated model-fitting technique. Each implant was analyzed at various flexion angles to assess the amount of femoral head sliding.

Results: No femoral head sliding was observed in the subjects with a metal-on-metal implant, whereas all ten subjects with a metal-on-polyethylene implant had sliding that was greater than our threshold value of 0.75 mm. The average amount of femoral head sliding in these subjects was 2.0 mm, and the sliding was observed during the swing phase of gait. The sliding was typically seen medially while the femoral head remained in contact with the acetabular component superolaterally.

Conclusions: Femoral head sliding commonly occurred following traditional metal-on-polyethylene total hip arthroplasty but not after metal-on-metal arthroplasty. These kinematic data may be of value in future hip-simulation studies to better duplicate wear patterns observed in retrieval analyses, assist in the understanding of the lubrication and wear rates of metal-on-metal designs, and facilitate designing of prosthetic components that minimize wear and optimize hip kinematics.

Figures in this Article
    Sign In to Your Personal ProfileSign In To Access Full Content
    Not a Subscriber?
    Get online access for 30 days for $30
    New to JBJS?
    Sign up for a full subscription to both the print and online editions
    Register for a FREE limited account to get full access to all CME activities, to comment on public articles, or to sign up for alerts.
    Register for a FREE limited account to get full access to all CME activities
    Have a subscription to the print edition?
    Current subscribers to The Journal of Bone & Joint Surgery in either the print or quarterly DVD formats receive free online access to JBJS.org.
    Forgot your password?
    Enter your username and email address. We'll send you a reminder to the email address on record.

     
    Forgot your username or need assistance? Please contact customer service at subs@jbjs.org. If your access is provided
    by your institution, please contact you librarian or administrator for username and password information. Institutional
    administrators, to reset your institution's master username or password, please contact subs@jbjs.org
    Accreditation Statement
    These activities have been planned and implemented in accordance with the Essential Areas and policies of the Accreditation Council for Continuing Medical Education (ACCME) through the joint sponsorship of the American Academy of Orthopaedic Surgeons and The Journal of Bone and Joint Surgery, Inc. The American Academy of Orthopaedic Surgeons is accredited by the ACCME to provide continuing medical education for physicians.
    CME Activities Associated with This Article
    Submit a Comment
    Please read the other comments before you post yours. Contributors must reveal any conflict of interest.
    Comments are moderated and will appear on the site at the discretion of JBJS editorial staff.

    * = Required Field
    (if multiple authors, separate names by comma)
    Example: John Doe




    Related Articles
    Related Cases
    Related Content
    Topic Collections
    Hip
    Related Audio and Videos
    PubMed Articles
    Clinical Trials
    Readers of This Also Read...
    jbjs jobs
    12/22/2011
    ME - Central Maine Medical Center
    12/22/2011
    VA - Charleston Area Medical Center
    12/22/2011
    Virginia - Charleston Area Medical Center