Comparison of Two and Three-Dimensional Computerized Polyethylene Wear Analysis After Total Hip Arthroplasty
John M. Martell, MD; Eric Berkson, MD, MA; Richard Berger, MD; Joshua Jacobs, MD
The Journal of Bone & Joint Surgery.  2003; 85:1111-1117 
5 Recommendations (Recommend) | 3 Comments | Saved by 3 Users Save Case

Abstract

Background: The accurate determination of acetabular polyethylene wear in vivo is necessary to assess the clinical performance of the bearing surfaces of total hip replacements. Our objective in this study was to determine the clinical performance of two and three-dimensional computerized wear analysis and to assess the implications of this performance on requirements for patient enrollment in studies designed to detect wear of total hip prostheses.

Methods: Two and three-dimensional digital computerized analyses of acetabular polyethylene wear were performed on 153 hips in 140 patients. The acetabular components consisted of a polyethylene insert in a titanium shell, articulating with a 28-mm cobalt-chromium femoral component. The average duration of radiographic follow-up was 8.4 years. The correlation coefficient for two-dimensional versus three-dimensional analysis was calculated, as was the difference between the wear detected by the two techniques. The same observer analyzed each image twice, allowing an assessment of the repeatability of the two-dimensional and three-dimensional analyses. The impact of the clinical performance of each technique on the sample size needed for adequate power in prospective studies was evaluated.

Results: There was a high correlation between two-dimensional and three-dimensional wear analysis (r 2 = 0.933). In thirty-one (5.2%) of 595 observations, the wear values derived with the two-dimensional and three-dimensional techniques were not consistent. Logistic regression demonstrated that acetabular anteversion had a significant effect on the likelihood of such inconsistency occurring. The two-dimensional technique detected 90.1% of the total linear wear subsequently detected by the three-dimensional analysis. The average wear value was 1.09 mm as detected by two-dimensional analysis and 1.21 mm as detected by three-dimensional analysis. The two-dimensional technique was four times more repeatable than the three-dimensional technique. Power analysis indicated that up to 1.4 times more patients need to be enrolled if the three-dimensional technique is used for wear analysis.

Conclusions: While three-dimensional analysis detected 10% more wear, its repeatability was four times worse than that of the two-dimensional technique and, as a consequence, patient enrollment requirements for wear detection were higher. The poor quality of the lateral radiographs contributed to the decrease in the repeatability of the three-dimensional analysis. Three-dimensional analysis may be useful for highly anteverted cups, but the limited improvement in wear detection achieved with that technique, coupled with its inferior repeatability, limits its clinical value.

Level of Evidence: Diagnostic study, Level IV-2 (poor reference standard). See Instructions to Authors for a complete description of levels of evidence.

Figures in this Article
    Sign In to Your Personal ProfileSign In To Access Full Content
    Not a Subscriber?
    Get online access for 30 days for $30
    New to JBJS?
    Sign up for a full subscription to both the print and online editions
    Register for a FREE limited account to get full access to all CME activities, to comment on public articles, or to sign up for alerts.
    Register for a FREE limited account to get full access to all CME activities
    Have a subscription to the print edition?
    Current subscribers to The Journal of Bone & Joint Surgery in either the print or quarterly DVD formats receive free online access to JBJS.org.
    Forgot your password?
    Enter your username and email address. We'll send you a reminder to the email address on record.

     
    Forgot your username or need assistance? Please contact customer service at subs@jbjs.org. If your access is provided
    by your institution, please contact you librarian or administrator for username and password information. Institutional
    administrators, to reset your institution's master username or password, please contact subs@jbjs.org
    Accreditation Statement
    These activities have been planned and implemented in accordance with the Essential Areas and policies of the Accreditation Council for Continuing Medical Education (ACCME) through the joint sponsorship of the American Academy of Orthopaedic Surgeons and The Journal of Bone and Joint Surgery, Inc. The American Academy of Orthopaedic Surgeons is accredited by the ACCME to provide continuing medical education for physicians.
    CME Activities Associated with This Article
    Submit a Comment
    Please read the other comments before you post yours. Contributors must reveal any conflict of interest.
    Comments are moderated and will appear on the site at the discretion of JBJS editorial staff.

    * = Required Field
    (if multiple authors, separate names by comma)
    Example: John Doe




    Related Articles
    Related Cases
    Related Content
    Topic Collections
    Hip
    Related Audio and Videos
    PubMed Articles
    Adverse reaction to metal release from a modular metal-on-polyethylene hip prosthesis.
    The Journal of bone and joint surgery. British volume: Issue date- 2011 Oct
    Edge-loading severity as a function of cup lip radius in metal-on-metal total hips--a finite element analysis.
    Journal of orthopaedic research : official publication of the Orthopaedic Research Society: Issue date- 2012 Feb
    Clinical Trials
    Readers of This Also Read...
    jbjs jobs
    12/22/2011
    ME - Central Maine Medical Center
    12/22/2011
    VA - Charleston Area Medical Center
    12/22/2011
    Virginia - Charleston Area Medical Center