0
Scientific Article   |    
Effect of Terminal Sterilization with Gas Plasma or Gamma Radiation on Wear of Polyethylene Liners
Robert H. HopperJr., PhD; Anthony M. Young, MS; Karl F. Orishimo, MS; C. Anderson EnghJr., PhD
The Journal of Bone & Joint Surgery.  2003; 85:464-468 
5 Recommendations (Recommend) | 3 Comments | Saved by 3 Users Save Case

Abstract

Background: Although terminal sterilization with gamma radiation in air promotes cross-linking, which improves wear resistance, it also results in free radicals, which can oxidize and degrade the mechanical properties of polyethylene liners used for total hip arthroplasty. For this reason, non-cross-linked polyethylene components have also been sterilized with chemical surface treatments, such as gas plasma. In this study, we tested the hypothesis that conventional polyethylene liners cross-linked by sterilization with gamma radiation in air had better in vivo wear performance than non-cross-linked liners sterilized with gas plasma.

Methods: We retrospectively reviewed the wear rates in a series of hips treated with a Duraloc 100 cup, a 28-mm femoral head, and an Enduron liner that had been sterilized with either gamma radiation (sixty-one hips followed for a mean of 5.2 years) or gas plasma (sixty-three hips followed for a mean of 3.9 years). The irradiated liners had been stored with access to ambient oxygen for an average of 1.0 year (range, 0.05 to 5.72 years) prior to implantation. Multiple linear regression was used to assess the effect of the sterilization method on the wear rate while accounting for the possible influence of other factors, including liner geometry, femoral head material, patient gender, cup abduction angle, and age at surgery.

Results: The polyethylene liners that had been sterilized with gamma radiation in air had a significantly lower wear rate than did the gas-plasma-sterilized liners (0.097 compared with 0.19 mm/yr, p < 0.001). The sterilization method (p < 0.001) and age at surgery (p = 0.001) were the only factors that we analyzed that correlated with the wear rate.

Conclusions: The in vivo wear of conventional polyethylene liners that had been sterilized with gamma radiation in air was, on the average, 50% less than that of non-cross-linked liners sterilized with gas plasma.

Level of Evidence: Therapeutic study, Level III-2 (retrospective cohort study). See Intructions to Authors for a complete description of levels of evidence.

Figures in this Article
    Sign In to Your Personal ProfileSign In To Access Full Content
    Not a Subscriber?
    Get online access for 30 days for $30
    New to JBJS?
    Sign up for a full subscription to both the print and online editions
    Register for a FREE limited account to get full access to all CME activities, to comment on public articles, or to sign up for alerts.
    Register for a FREE limited account to get full access to all CME activities
    Have a subscription to the print edition?
    Current subscribers to The Journal of Bone & Joint Surgery in either the print or quarterly DVD formats receive free online access to JBJS.org.
    Forgot your password?
    Enter your username and email address. We'll send you a reminder to the email address on record.

     
    Forgot your username or need assistance? Please contact customer service at subs@jbjs.org. If your access is provided
    by your institution, please contact you librarian or administrator for username and password information. Institutional
    administrators, to reset your institution's master username or password, please contact subs@jbjs.org
    Accreditation Statement
    These activities have been planned and implemented in accordance with the Essential Areas and policies of the Accreditation Council for Continuing Medical Education (ACCME) through the joint sponsorship of the American Academy of Orthopaedic Surgeons and The Journal of Bone and Joint Surgery, Inc. The American Academy of Orthopaedic Surgeons is accredited by the ACCME to provide continuing medical education for physicians.
    CME Activities Associated with This Article
    Submit a Comment
    Please read the other comments before you post yours. Contributors must reveal any conflict of interest.
    Comments are moderated and will appear on the site at the discretion of JBJS editorial staff.

    * = Required Field
    (if multiple authors, separate names by comma)
    Example: John Doe




    Related Articles
    Related Cases
    Related Content
    Topic Collections
    Hip
    Related Audio and Videos
    PubMed Articles
    Clinical Trials
    Readers of This Also Read...
    jbjs jobs
    12/22/2011
    VA - Charleston Area Medical Center
    12/22/2011
    ME - Central Maine Medical Center
    12/22/2011
    Virginia - Charleston Area Medical Center