Abstract

Background: While high success rates have been achieved in association with other all-polyethylene tibial components, an alarming number of failures have occurred at our institution in association with the use of an all-polyethylene version of the AGC tibial component. The purpose of the present study was to describe the survival of the AGC all-polyethylene tibial component.

Methods: Five hundred and thirty-six AGC all-polyethylene tibial components were implanted in 405 patients and were followed over a ten-year period. The average age of the patients at the time of surgery was 70.3 years, the average weight was 78 kg, and the most common diagnosis was osteoarthritis (prevalence, 92.9%). A clinical and radiographic analysis was performed, Knee Society knee and function scores were determined, and Kaplan-Meier survivorship analysis was conducted. Failure was defined as aseptic loosening as evidenced by progressive radiolucent lines and/or revision due to aseptic loosening or collapse.

Results: A high rate of failure was noted in the early postoperative period, with a survival rate of 90.04% (95% confidence interval, 87.35% to 92.72%) after three years. At ten years, the survival rate was 68.11% (95% confidence interval, 57.57% to 78.65%). Fifty-eight (73.4%) of seventy-nine failures occurred in association with loosening or collapse of the bone beneath the medial tibial plateau.

Conclusion: While some all-polyethylene tibial designs have been successful, the low success rate among knees treated with the AGC all-polyethylene tibial component suggests that the results associated with all-polyethylene tibial components are design-sensitive.

Level of Evidence: Therapeutic study, Level IV (case series [no, or historical, control group]). See Instructions to Authors for a complete description of levels of evidence.

Figures in this Article

    Topics

    tibia ; polyethylene
    Sign In to Your Personal ProfileSign In To Access Full Content
    Not a Subscriber?
    Get online access for 30 days for $30
    New to JBJS?
    Sign up for a full subscription to both the print and online editions
    Register for a FREE limited account to get full access to all CME activities, to comment on public articles, or to sign up for alerts.
    Register for a FREE limited account to get full access to all CME activities
    Have a subscription to the print edition?
    Current subscribers to The Journal of Bone & Joint Surgery in either the print or quarterly DVD formats receive free online access to JBJS.org.
    Forgot your password?
    Enter your username and email address. We'll send you a reminder to the email address on record.

     
    Forgot your username or need assistance? Please contact customer service at subs@jbjs.org. If your access is provided
    by your institution, please contact you librarian or administrator for username and password information. Institutional
    administrators, to reset your institution's master username or password, please contact subs@jbjs.org
    Accreditation Statement
    These activities have been planned and implemented in accordance with the Essential Areas and policies of the Accreditation Council for Continuing Medical Education (ACCME) through the joint sponsorship of the American Academy of Orthopaedic Surgeons and The Journal of Bone and Joint Surgery, Inc. The American Academy of Orthopaedic Surgeons is accredited by the ACCME to provide continuing medical education for physicians.
    CME Activities Associated with This Article
    Submit a Comment
    Please read the other comments before you post yours. Contributors must reveal any conflict of interest.
    Comments are moderated and will appear on the site at the discretion of JBJS editorial staff.

    * = Required Field
    (if multiple authors, separate names by comma)
    Example: John Doe




    Related Articles
    Related Cases
    Related Content
    Topic Collections
    Related Audio and Videos
    PubMed Articles
    How does TKA kinematics vary with transverse plane alignment changes in a contemporary implant?
    Clinical orthopaedics and related research: Issue date- 2012 Jan
    All-polyethylene tibial components in modern total knee arthroplasty.
    The Journal of the American Academy of Orthopaedic Surgeons: Issue date- 2011 Sep
    Clinical Trials
    Readers of This Also Read...
    jbjs jobs
    12/22/2011
    ME - Central Maine Medical Center
    12/22/2011
    VA - Charleston Area Medical Center
    12/22/2011
    Maine - Central Maine Medical Center