Background: Following the reintroduction of metal-on-metal
articulating surfaces for total hip arthroplasty in Europe in 1988, we
developed a surface arthroplasty prosthetic system using a metal-on-metal
articulation. The present study describes the clinical and radiographic
results of the first 400 hips treated with metal-on-metal hybrid surface
arthroplasties at an average follow-up of three and a half years.
Methods: Between November 1996 and November 2000, 400 metal-on-metal
hybrid surface arthroplasties were performed in 355 patients. All femoral head
components were cemented, but only fifty-nine of the short metaphyseal stems
were cemented. The patients had an average age of forty-eight years, 73% were
men, and 66% had a diagnosis of osteoarthritis. Clinical and radiographic
follow-up were performed at three months postoperatively and yearly
thereafter.
Results: The majority of the patients returned to a high level of
activity, including sports, and 54% had activity scores of >7 on the
University of California at Los Angeles activity assessment system.
Kaplan-Meier survivorship curves demonstrated that the rate of survival of the
components at four years was 94.4%. For patients with a surface arthroplasty
risk index score of >3, the rate of survival of the components at four
years was 89% compared with a rate of 97% for those with a score of =3. The
patients with a higher risk index were 4.2 times more likely to undergo
revision to a total hip replacement at four years. Twelve hips (3%) had a
revision to a total hip replacement. Seven of the twelve hips were revised
because of loosening of the femoral component, and three were revised because
of a femoral neck fracture. Substantial radiolucencies were seen around
sixteen uncemented metaphyseal femoral stems. No femoral radiolucencies were
observed among the hips in which the metaphyseal stem was cemented. The most
important risk factors for femoral component loosening and substantial stem
radiolucencies were large femoral head cysts (p = 0.029), patient height (p =
0.032), female gender (p = 0.005), and smaller component size in male patients
(p = 0.005).
Conclusions: The preliminary experience with this hybrid
metal-on-metal bearing is encouraging. Optimal femoral bone preparation and
component fixation are critical to improving durability. The metal-on-metal
hybrid surface arthroplasty is easily revised to a standard femoral component
if necessary.
Level of Evidence: Therapeutic study, Level IV (case
series [no, or historical, control group]). See Instructions to Authors for a
complete description of levels of evidence.