0
Scientific Articles   |    
Accuracy in the Measurement of Compartment Pressures: A Comparison of Three Commonly Used Devices
Antony R. Boody, MD1; Montri D. Wongworawat, MD1
1 Department of Orthopaedic Surgery, Loma Linda University, 11406 Loma Linda Drive, Suite 218, Loma Linda, CA 92354. E-mail address for M.D. Wongworawat: wongwora@usc.edu
The Journal of Bone & Joint Surgery.  2005; 87:2415-2422  doi:10.2106/JBJS.D.02826
5 Recommendations (Recommend) | 3 Comments | Saved by 3 Users Save Case

Abstract

Background: In situations in which accurate physical diagnosis is inconclusive, an objective method for measuring compartment pressure can aid in the diagnosis of compartment syndrome. Previous studies have compared measurement devices with each other but not with an accurately determined gold standard. The purpose of the present study was to devise a reproducible in vitro model of compartment pressure and to compare commonly used measurement devices in order to determine their accuracy.

Methods: With a graduated cylinder being used to generate a known pressure, freshly harvested ovine muscle was placed into a chamber for testing. The cylinder was incrementally filled with saline solution (in fifty-five steps), and measurements of tissue pressure were obtained with use of the Stryker Intracompartmental Pressure Monitor System, an arterial line manometer, and the Whitesides apparatus. Each device was tested with a straight needle, a side-port needle, and a slit catheter, for a total of nine setups in all. Five trials were done with each setup. Control pressures were calculated on the basis of the height of the saline solution column (test range, 0.13 to 10.80 kPa). Multiple regression analysis was used to compare measured tissue pressures with calculated control pressures.

Results: Most methods demonstrated excellent correlation (R2 > 0.95) between calculated and measured pressures. The arterial line manometer with the slit catheter showed the best correlation (R2 = 0.9978), and the Whitesides apparatus with the side-port needle showed the worst (R2 = 0.9115). Furthermore, the Stryker system with the side-port needle demonstrated the least constant bias (+0.06 kPa). Straight needles tended to overestimate pressure. Two of the three needle configurations involving the Whitesides apparatus overestimated pressure. The data for the Whitesides methods had the highest standard errors, showing clinically unacceptable scatter.

Conclusion: Side-port needles and slit catheters are more accurate than straight needles are. The arterial line manometer is the most accurate device. The Stryker device is also very accurate. The Whitesides manometer apparatus lacks the precision needed for clinical use.

Clinical Relevance: When physical examination findings are inconclusive, accurate measurement of compartment pressures can aid in timely management and can minimize patient morbidity. Measurement should be done with use of the most accurate technique available.

Figures in this Article
    Sign In to Your Personal ProfileSign In To Access Full Content
    Not a Subscriber?
    Get online access for 30 days for $30
    New to JBJS?
    Sign up for a full subscription to both the print and online editions
    Register for a FREE limited account to get full access to all CME activities, to comment on public articles, or to sign up for alerts.
    Register for a FREE limited account to get full access to all CME activities
    Have a subscription to the print edition?
    Current subscribers to The Journal of Bone & Joint Surgery in either the print or quarterly DVD formats receive free online access to JBJS.org.
    Forgot your password?
    Enter your username and email address. We'll send you a reminder to the email address on record.

     
    Forgot your username or need assistance? Please contact customer service at subs@jbjs.org. If your access is provided
    by your institution, please contact you librarian or administrator for username and password information. Institutional
    administrators, to reset your institution's master username or password, please contact subs@jbjs.org
    Accreditation Statement
    These activities have been planned and implemented in accordance with the Essential Areas and policies of the Accreditation Council for Continuing Medical Education (ACCME) through the joint sponsorship of the American Academy of Orthopaedic Surgeons and The Journal of Bone and Joint Surgery, Inc. The American Academy of Orthopaedic Surgeons is accredited by the ACCME to provide continuing medical education for physicians.
    CME Activities Associated with This Article
    Submit a Comment
    Please read the other comments before you post yours. Contributors must reveal any conflict of interest.
    Comments are moderated and will appear on the site at the discretion of JBJS editorial staff.

    * = Required Field
    (if multiple authors, separate names by comma)
    Example: John Doe




    Montri D. Wongworawat
    Posted on December 11, 2005
    Drs. Wongworawat and Boody respond to Dr. Barnes
    Loma Linda University, Loma Linda, CA 92354

    We thank Dr. Barnes for his letter and respond to his comments below.

    The arterial line manometer is an electronic pressure transducer used in our operating suites and telemetry monitored units. The tubing and manometer are single-use items packaged and manufactured by Maxxim Medical (Athens, TX). The electronic portion of the device and monitor screen are made by HP (Houston, TX). The mechanical pressure is transformed to a visual waveform and a number (mmHg) which is read from the monitor. There is no diameter listed for the tubing itself, but we “zeroed” (tared) the device at the level of measurement prior to data collection, so differential diameter of needle and tubing should not be an issue. We did not change tubing in the arterial line manometer system for any of the measurements.

    The slit catheter is 18-gauge. It is inserted into tissue using a 16 -gauge needle which is removed prior to use.

    Regarding the question pertaining to temperature, it is true that temperature affects the pressure of any system. However, the effect of varying temperatures on fluid density is nearly negligible from a clinical standpoint (0.998 g/cm³ at room temperature and 0.993 g/cm³ at body temperature, which represents a 0.5% change)—see reference 20 in our manuscript. The primary reason for using warmed saline was to keep the muscle tissue in a physiologic environment so as to avoid the development of muscle stiffness associated with excessive cooling. While the temperature in the column was not measured with each trial, warm saline was added in an incremental fashion throughout the experiment; we believe that this sufficed in keeping the system at near physiologic conditions.

    Dr. Barnes is absolutely correct regarding the effect ( or lack thereof) of atmospheric pressure on our measurements.

    Michael R Barnes
    Posted on November 30, 2005
    Measurement of Compartment Pressures
    Leicester General Hospital, Leicestershire, LE5 4PW, UK

    To The Editor:

    This very worthwhile, comprehensive and long overdue paper addresses important aspects about the accuracy of compartment pressure measurements. While there are some interesting explanations about the importance of a good tissue model and the effect of different needles and catheters, the authors leave unanswered a number of important methodologic questions.

    Comprehensive details and references are given about where to find further information about the Stryker system and the Whitesides method, but only the phrase “arterial line manometer” is used to describe the third device. What does this term mean? Is it a pressure transducer and, if so, we need more details about its manufacturer and its characteristics. How is it connected to the needles?

    This is important information to know because all the associated external plumbing with these systems can have a significant effect on the pressures recorded (possibly greater than the differences between the three methods that were compared).

    It is clearly stated that the needles used were 18g. but the diameter of the slit catheter is not stated. The diameter may have a significant effect on the pressure measured.

    The authors are obviously aware of the effect that temperature changes can have on the pressure recorded, but it is not clear how these were taken into account. They state that normal saline was heated to 37°C to mimic physiological conditions. Was the whole system then maintained at 37°C throughout the measurements, or was the temperature of the saline measured each time a pressure measurement was recorded? Did the entire experimental set up gradually cool to room temperature as the pressure was increased? The authors were measuring small differences between the different devices, but the effects of changes in temperature could far outweigh these measured differences.

    Immediate atmospheric pressure changes could not have affected the results. Since all the pressures were measured with respect to atmospheric pressure, they are not absolute values. If there was a change in atmospheric pressure then it would affect both the measurement device and the source in exactly the same way. I look forward to a clarification of these issues from the authors.

    Related Articles
    Related Cases
    Related Content
    Topic Collections
    Related Audio and Videos
    PubMed Articles
    Clinical Trials
    Readers of This Also Read...
    jbjs jobs
    12/22/2011
    ME - Central Maine Medical Center
    12/22/2011
    VA - Charleston Area Medical Center
    12/22/2011
    Virginia - Charleston Area Medical Center