Anterior Cervical Fixation: Analysis of Load-Sharing and Stability with Use of Static and Dynamic Plates
Darrel S. Brodke, MD1; Paul KlimoJr., MD, MPH2; Kent N. Bachus, PhD1; John T. Braun, MD1; Andrew T. Dailey, MD3
1 Department of Orthopaedics (D.S.B., K.N.B., and J.T.B.) and Orthopaedic Research Laboratory (K.N.B.), University of Utah Orthopaedic Center, 590 Wakara Way, Room A0100, Salt Lake City, UT 84108. E-mail address for K.N. Bachus: kent.bachus@hsc.utah.edu
2 Department of Neurosurgery, University of Utah School of Medicine, 30N 1900E, Suite 3B409 SOM, Salt Lake City, UT 84132
3 Rocky Mountain Neurosurgical Alliance, 701 E Hampden Avenue #510, Englewood, CO 80110
The Journal of Bone & Joint Surgery.  2006; 88:1566-1573  doi:10.2106/JBJS.E.00305
5 Recommendations (Recommend) | 3 Comments | Saved by 3 Users Save Case

Abstract

Background: Anterior plates provide stability following decompression and fusion of the cervical spine. Various plate designs have emerged, and they include static plates with fixed-angle screws, rotationally dynamic plates that allow the screws to toggle in the plate, and translationally dynamic plates that allow the screws to both toggle and translate vertically. The goal of this study was to document the effects of plate design following a single-level corpectomy and placement of a full-length strut graft and the effects following 10% subsidence of the graft.

Methods: A total of twenty-one cadaveric cervical spines (C2-T1) were randomized into three treatment groups and were tested for initial range of motion. A C5 corpectomy was performed, reconstruction was done with a full-length interbody spacer containing a load-cell, and an anterior cervical plate was applied. Load-sharing data were recorded with incremental axial loads. The range of motion was measured with ±2.5 Nm of torque in flexion-extension, lateral bending, and axial rotation. Then, the total length of the interbody spacer was reduced by 10% to simulate subsidence, and load-sharing and the range of motion were retested.

Results: With the full-length interbody spacer, there were no significant differences in the abilities of the constructs to share load or limit motion. Following shortening of the interbody spacer, the static plate construct lost nearly 70% of its load-sharing capability, while neither of the dynamic plate constructs lost load-sharing capabilities. Also, the static plate construct allowed significantly more motion in flexion-extension following simulated subsidence than did either of the dynamic plate constructs (p < 0.05).

Conclusions: Although all of the tested anterior cervical plating systems provide similar load-sharing and stiffness following initial placement of the interbody spacer, the static plate system lost its ability to share load and limit motion following simulated subsidence of the interbody spacer. Both dynamic plate systems maintained load-sharing and stiffness despite simulated subsidence.

Clinical Relevance: This study provides an improved understanding of the immediate performance of anterior cervical fusion surgery with plate fixation.

Figures in this Article
    Sign In to Your Personal ProfileSign In To Access Full Content
    Not a Subscriber?
    Get online access for 30 days for $30
    New to JBJS?
    Sign up for a full subscription to both the print and online editions
    Register for a FREE limited account to get full access to all CME activities, to comment on public articles, or to sign up for alerts.
    Register for a FREE limited account to get full access to all CME activities
    Have a subscription to the print edition?
    Current subscribers to The Journal of Bone & Joint Surgery in either the print or quarterly DVD formats receive free online access to JBJS.org.
    Forgot your password?
    Enter your username and email address. We'll send you a reminder to the email address on record.

     
    Forgot your username or need assistance? Please contact customer service at subs@jbjs.org. If your access is provided
    by your institution, please contact you librarian or administrator for username and password information. Institutional
    administrators, to reset your institution's master username or password, please contact subs@jbjs.org
    Accreditation Statement
    These activities have been planned and implemented in accordance with the Essential Areas and policies of the Accreditation Council for Continuing Medical Education (ACCME) through the joint sponsorship of the American Academy of Orthopaedic Surgeons and The Journal of Bone and Joint Surgery, Inc. The American Academy of Orthopaedic Surgeons is accredited by the ACCME to provide continuing medical education for physicians.
    CME Activities Associated with This Article
    Submit a Comment
    Please read the other comments before you post yours. Contributors must reveal any conflict of interest.
    Comments are moderated and will appear on the site at the discretion of JBJS editorial staff.

    * = Required Field
    (if multiple authors, separate names by comma)
    Example: John Doe




    Related Articles
    Related Cases
    Related Content
    Topic Collections
    Related Audio and Videos
    PubMed Articles
    Galeazzi fracture.
    The Journal of the American Academy of Orthopaedic Surgeons: Issue date- 2011 Oct
    Clinical Trials
    Readers of This Also Read...
    jbjs jobs
    12/22/2011
    ME - Central Maine Medical Center
    12/22/2011
    VA - Charleston Area Medical Center
    12/22/2011
    Maine - Central Maine Medical Center