Background: Anterior plates provide stability following
decompression and fusion of the cervical spine. Various plate designs have
emerged, and they include static plates with fixed-angle screws, rotationally
dynamic plates that allow the screws to toggle in the plate, and
translationally dynamic plates that allow the screws to both toggle and
translate vertically. The goal of this study was to document the effects of
plate design following a single-level corpectomy and placement of a
full-length strut graft and the effects following 10% subsidence of the
graft.
Methods: A total of twenty-one cadaveric cervical spines (C2-T1)
were randomized into three treatment groups and were tested for initial range
of motion. A C5 corpectomy was performed, reconstruction was done with a
full-length interbody spacer containing a load-cell, and an anterior cervical
plate was applied. Load-sharing data were recorded with incremental axial
loads. The range of motion was measured with ±2.5 Nm of torque in
flexion-extension, lateral bending, and axial rotation. Then, the total length
of the interbody spacer was reduced by 10% to simulate subsidence, and
load-sharing and the range of motion were retested.
Results: With the full-length interbody spacer, there were no
significant differences in the abilities of the constructs to share load or
limit motion. Following shortening of the interbody spacer, the static plate
construct lost nearly 70% of its load-sharing capability, while neither of the
dynamic plate constructs lost load-sharing capabilities. Also, the static
plate construct allowed significantly more motion in flexion-extension
following simulated subsidence than did either of the dynamic plate constructs
(p < 0.05).
Conclusions: Although all of the tested anterior cervical plating
systems provide similar load-sharing and stiffness following initial placement
of the interbody spacer, the static plate system lost its ability to share
load and limit motion following simulated subsidence of the interbody spacer.
Both dynamic plate systems maintained load-sharing and stiffness despite
simulated subsidence.
Clinical Relevance: This study provides an improved understanding of
the immediate performance of anterior cervical fusion surgery with plate
fixation.