Background: Intramedullary fixation with use of a telescopic rod
with a T-piece is one of the standard methods for long-bone stabilization in
growing children with osteogenesis imperfecta. However, installation and
removal of this device can cause substantial damage to the distal joint, which
limits its use, especially in the tibia. We devised a modification of the
telescopic rod system—the interlocking telescopic rod—in which the
obturator is a simple rod with a hole, instead of a T-piece, at its distal
end.
Methods: The clinical and radiographic outcomes were evaluated more
than two years following treatment of thirty-two limb segments (twenty-three
tibiae and nine femora) with this new rod system in fifteen patients with
osteogenesis imperfecta.
Results: All rods were inserted without an arthrotomy of the distal
joint, and all telescoped successfully. The interlocking pin used in the first
five limb segments backed out between five and thirty-three months
postoperatively. A revised fixation technique was used in the remaining
twenty-seven limb segments, and the interlocking pin had not backed out at an
average 3.1 years postoperatively. Proximal migration of the obturator was
observed in four tibiae after 2.5 years. The cumulative survival rate of the
rod at four years postoperatively was 88.7%.
Conclusions: Both insertion and removal of an interlocking
telescopic rod are much less invasive than insertion and removal of a
conventional telescopic rod with a T-piece anchor. The interlocking pin at the
distal epiphysis provides effective anchorage for telescoping. Our interim
results showed survival of the device to be comparable with, or better than,
that of the conventional telescopic rod.
Level of Evidence: Therapeutic Level IV. See Instructions
to Authors for a complete description of levels of evidence.