Stability of Cemented in-Line Pegged Glenoid Compared with Keeled Glenoid Components in Total Shoulder Arthroplasty
Hans Rahme, MD1; Per Mattsson, MD1; Lars Wikblad, MD1; Jan Nowak, MD1; Sune Larsson, MD1
1 Department of Orthopedics, Uppsala University Hospital, S-751 85 Uppsala, Sweden. E-mail address for H. Rahme: hans.rahme@telia.com
The Journal of Bone & Joint Surgery.  2009; 91:1965-1972  doi:10.2106/JBJS.H.00938
5 Recommendations (Recommend) | 3 Comments | Saved by 3 Users Save Case

Abstract

Background: Considerable interest has been focused on the design of the glenoid component used in total shoulder arthroplasty in order to reduce the risk of loosening. One design-related feature that has attracted attention is whether to use pegged or keeled cemented glenoid components. The main purpose of this study was to compare the fixation of cemented keeled glenoid components with that of cemented in-line pegged glenoid components.

Methods: In a prospective randomized study, we compared the stability of cemented, all-polyethylene, keeled glenoid components and cemented, all-polyethylene, in-line three-pegged glenoid components by radiostereometric analysis. Twenty-seven shoulders in twenty-five patients with osteoarthritis (twenty-two shoulders had primary and five shoulders had secondary osteoarthritis) were included. There were sixteen women and nine men, and the mean age was sixty-four years. Radiostereometric analysis and conventional radiographs were carried out at five days, at four months, and at one and two years postoperatively.

Results: The mean Constant and Murley score preoperatively and two years postoperatively was 25 and 70, respectively, for shoulders with the keeled glenoid component and 22 and 70 for the shoulders with a pegged component. No significant difference was detected between groups with regard to the average micromigration of the glenoid components at any of the time points. The average translation was <1 mm, while the median value was <0.3 mm at two years, with no significant difference between the different axes. In five shoulders (three with the keeled component and two with the pegged component), translation at two years was >1 mm. In fourteen shoulders (eight with the keeled and six with the pegged component), the rotation around one or several axes was >2°. We were not able to detect any specific pattern with regard to movement for either type of component nor were we able to detect any difference between the two types of components in the way they migrated, if migration occurred.

Conclusions: Cemented all-polyethylene keeled or in-line three-pegged glenoid components appear to have similar stability during the first two years after surgery. Studies with a longer follow-up period are needed to relate these findings to long-term clinical and radiographic outcomes.

Level of Evidence: Therapeutic Level I. See Instructions to Authors for a complete description of levels of evidence.

Figures in this Article
    Sign In to Your Personal ProfileSign In To Access Full Content
    Not a Subscriber?
    Get online access for 30 days for $30
    New to JBJS?
    Sign up for a full subscription to both the print and online editions
    Register for a FREE limited account to get full access to all CME activities, to comment on public articles, or to sign up for alerts.
    Register for a FREE limited account to get full access to all CME activities
    Have a subscription to the print edition?
    Current subscribers to The Journal of Bone & Joint Surgery in either the print or quarterly DVD formats receive free online access to JBJS.org.
    Forgot your password?
    Enter your username and email address. We'll send you a reminder to the email address on record.

     
    Forgot your username or need assistance? Please contact customer service at subs@jbjs.org. If your access is provided
    by your institution, please contact you librarian or administrator for username and password information. Institutional
    administrators, to reset your institution's master username or password, please contact subs@jbjs.org
    Accreditation Statement
    These activities have been planned and implemented in accordance with the Essential Areas and policies of the Accreditation Council for Continuing Medical Education (ACCME) through the joint sponsorship of the American Academy of Orthopaedic Surgeons and The Journal of Bone and Joint Surgery, Inc. The American Academy of Orthopaedic Surgeons is accredited by the ACCME to provide continuing medical education for physicians.
    CME Activities Associated with This Article
    Submit a Comment
    Please read the other comments before you post yours. Contributors must reveal any conflict of interest.
    Comments are moderated and will appear on the site at the discretion of JBJS editorial staff.

    * = Required Field
    (if multiple authors, separate names by comma)
    Example: John Doe




    Related Articles
    Related Cases
    Related Content
    Topic Collections
    Related Audio and Videos
    PubMed Articles
    Vacuum assist glenoid fixation: does this technique lead to a more durable glenoid component?
    Journal of shoulder and elbow surgery / American Shoulder and Elbow Surgeons ... [et al.]: Issue date- 2011 Oct
    Clinical Trials
    Readers of This Also Read...
    jbjs jobs
    12/22/2011
    ME - Central Maine Medical Center
    12/22/2011
    VA - Charleston Area Medical Center
    12/22/2011
    Virginia - Charleston Area Medical Center